HemoSphere recallZ-0496-2018 · Edwards Lifesciences, LLC · initiated Oct 2, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Pre-procedural issues related to software defects
FDA root cause category: Device Design
What the company is doing
Consignees were sent recall notification letters on approximately 10/11/2017. The letter identified the affected product, problem and actions to be taken. instructions included coordination with the respective Edwards representative to schedule the software upgrade.
Recall details
- Recall number
- Z-0496-2018
- Status
- Terminated
- Initiated by the company
- Oct 2, 2017
- Posted by the FDA
- Jan 16, 2018
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 198 devices
- Product code
- DQK
- FDA submissions
- K163381
- Distribution
- US Nationwide Distributed in the states of AZ, CA, CO, GA, IL, MO, NM, OK, OR, SC, TX
Product description
Hemosphere System with the following three components: HEM1 Advanced Monitor, HEMSGM10 (Swan-Ganz Module), and HEMOXSC100 (Oximetry Smart Cable). Product Usage: The HemoSphere Advanced Monitor (HEM1) is intended to be used in combination with a compatible Edwards hemodynamic monitoring technology module and/or cable and accompanying Edwards accessories and/or disposables. It provides Edwards hemodynamic monitoring technologies through modular connections. The HemoSphere Swan-Ganz Module (HEMSGM10) is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is a multi-use device that works in conjunction with the HemoSphere Advanced Monitor, patient CCO cable and compatible Swan-Ganz catheter to acquire and process temperature data supporting intermittent cardiac output (iCO), continuous cardiac output (CCO) and right ventricular end diastolic volume (EDV) monitoring. The HemoSphere Oximetry Smart Cable (HEMOXSC100) is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation by reflectance spectrophotometry. LEDs within the oximetry cable transmit light fiber optically to the distal end of the catheter. The amount of light absorbed, refracted, and reflected depends on the relative amounts of oxygenated and deoxygenated hemoglobin in the blood. This optical intensity data is gathered by the oximetry catheter, processed by the HemoSphere Oximetry Cable, and displayed on a compatible monitoring platform. Parameter output is mixed venous oxygen saturation (SvO2) or central venous oxygen saturation (ScvO2).
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
HEM1 UDI 00690103197006 (Serial No. 13800180, 13800185, 13800193, 13800195, 13800228, 13800162, 13800168, 13800171, 13800208, 13800210, 13800211, 13800126, 13800128, 13800129, 13800130, 13800132, 13800133, 13800135, 13800139, 13800140, 13800144, 13800146, 13800147, 13800148, 13800150, 13800187, 13800189, 13800131, 13800134, 13800136, 13800191, 13800192, 13800197, 13800172, 13800173, 13800159, 13800184, 13800188, 13800339, 13800355, 13800199, 13800200, 13800141, 13800249, 13800258, 13800259, 13800260, 13800263, 13800265, 13800268, 13800270, 13800176, 13800177, 13800179, 13800186, 13800225, 13800227, 13800230, 13800196, 13800124, 13800125, 13800137, 13800138, 13800142, 13800143, 13800145, 13800149, 13800151, 13800152); HEMSGM10 UDI 00690103196993 (Serial No. 13600223, 13600224, 13600225, 13600226, 13600257, 13600196, 13600197, 13600198, 13600227, 13600228, 13600229, 13600158, 13600159, 13600160, 13600161, 13600162, 13600163, 13600164, 13600165, 13600166, 13600167, 13600168, 13600169, 13600170, 13600171, 13600206, 13600207, 13600145, 13600146, 13600147, 13600208, 13600209, 13600211, 13600199, 13600201, 13600193, 13600194, 13600195, 13600390, 13600421, 13600212, 13600213, 13600202, 13600203, 13600204, 13600205, 13600255, 13600256, 13600258, 13600210, 13600148, 13600149, 13600150, 13600151 13600152, 13600153, 13600154, 13600155, 13600156, 13600157); HEMOXSC100 UDI 00690103196986 (Serial No. 13700241, 13700246, 13700247, 13700248, 13700303, 13700188, 13700189, 13700190, 13700280, 13700282, 13700291, 13700149, 13700150, 13700151, 13700152, 13700153, 13700154, 13700155, 13700156, 13700157, 13700158, 13700159, 13700160, 13700161, 13700162, 13700174, 13700177, 13700136, 13700137, 13700138, 13700179, 13700199, 13700182, 13700191, 13700192, 13700193, 13700194, 13700180, 13700185, 13700186, 13700183, 13700187, 13700201, 13700205, 13700215, 13700236, 13700240, 13700258, 13700265, 13700270, 13700271, 13700195, 13700196, 13700197, 13700198, 13700252, 13700301, 13700363, 13700175, 13700139, 13700140, 13700141, 13700142, 13700143, 13700144, 13700145, 13700146, 13700147, 13700148)
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- HemoSphere HEM1Edwards Lifesciences LLC · Catheter, Oximeter, Fiberoptic1 recall65 reports
- HemoSphere HEMOXSC100Edwards Lifesciences LLC · Catheter, Oximeter, Fiberoptic1 recall11 reports
- HemoSphere HEMSGM10Edwards Lifesciences LLC · Catheter, Oximeter, Fiberoptic1 recall5 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page