Stroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE… recallZ-0276-2018 · Stryker Neurovascular · initiated Nov 3, 2017
Class II Terminated All recalls posted in December 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the pack.
FDA root cause category: Labeling Change Control
What the company is doing
The firm, Stryker Neurovascular, sent an "Urgent Medical Device Voluntary Recall Immediate Action Required" letter initiating their recall on 11/01/2017. The letter described the product, problem and actions to be taken. The consignees were instructed to immediately check internal inventory; remove and discard the procedure pack carton sleeve; circulate notice; maintain awareness of notice internally until all required actions have been completed with your facility; inform Stryker of any subject devices distributed to other organization, and complete and return customer response form to your nominated Stryker Representative or to [email protected]. If you have any questions, call 510-413-2593; email: [email protected] or 510-413-2900.
Recall details
- Recall number
- Z-0276-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 3, 2017
- Posted by the FDA
- Dec 5, 2017
- Recalling firm
- Stryker Neurovascular
- Quantity
- 2 units
- Product code
- NRY
- FDA submissions
- K113260, K113778, K132641, K143077, K151667
- Distribution
- International Distribution to: Germany, Slovakia and Israel.
Product description
Stroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE RETRIEVER, 4 mm, 20 mm, ; Trevo(TM) Pro 18, TREVO PRO 18 MICROCATHETER, 0.021 in ID, 150 cm; AXS Catalyst(TM) 6, Distal Access Catheter, 0.060 in, 132 cm; UPN M0033PK42022002 Stroke intervention kit
Code information (models, lots, serial numbers)
Lot Number QPC30200039, Exp. Date 28-Aug-18; QPC30200114, Exp. Date 28-Mar18
Other devices under the same FDA submission
5 devices cleared under K113260, K113778, K132641, K143077, K151667 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Trevo XP 90185Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death40 reports
- Trevo XP 90186Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death31 reports
- Trevo 90183Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death30 reports
- Trevo 90182Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death29 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 78607).
- Z-0275-2018 · TerminatedStroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm; Excelsior(TM) XT-27(tm), MICROCATHETER,150 cm, 6 cm; AXS Catalyst(TM) 6, Distal Access Catheter, 0.0…
- Z-0277-2018 · TerminatedTrevo(TM) XP ProVue Procedure Pack, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm, .Excelsior(TM) XT-27(TM) 1 TIP STRAIGHT, 150 cm, 6 cm, Microcatheter; UPN 93068 Stroke intervention…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page