Tapered Screw-Vent recallZ-1190-2018 · Zimmer Dental Inc · initiated Nov 29, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Dental Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two lots of the Tapered Screw-Vent Implants, with different outer thread diameters, were inadvertently commingled during the manufacturing process and may be mislabeled.

FDA root cause category: Packaging process control

What the company is doing

On 11/29/2017, customers were notified via letter. Instructions for distributors include to ensure that affected team members are aware of the recall, immediately locate and quarantine affected devices in inventory, notify consignees if product was distributed, make arrangements to immediately return affected devices, and complete and return the Attachment 1 - Inventory Return Certification Form. Instructions for clinicians include to review the recall notice, examine inventory for affected devices, complete and return the Attachment 2 - Certificate of Acknowledgement, and return all affected devices.

Recall details

Recall number
Z-1190-2018
Status
Terminated
Initiated by the company
Nov 29, 2017
Posted by the FDA
Mar 5, 2018
Recalling firm
Zimmer Dental Inc
Quantity
995 distribution events total
Product code
DZE
FDA submissions
K072589
Distribution
Worldwide distribution US nationwide, Canada, Chile, Costa Rica, Denmark, France, Hungary, Italy, Japan, Lebanon, Portugal, Serbia, Spain, and Taiwan.

Product description

Tapered Screw-Vent MTX, Dental Implant, sterile. Model TSV4B10.

Code information (models, lots, serial numbers)

Lot No. 63619437, UDI #(01)00889024019225 (17)220331 (10)63619437.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event