Medical device recalls: March 2018
The FDA posted 82 medical device recall records in March 2018, from 40 recall events by 29 companies. 6 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Zimmer Biomet, Inc. · 5 events, 17 records
- Siemens Medical Solutions USA, Inc · 4 events, 10 records
- Arrow International Inc · 3 events, 11 records
- Smiths Medical ASD Inc. · 2 events, 5 records
- Getinge US Sales LLC · 2 events, 2 records
- Johnson & Johnson Consumer, Inc. · 1 event, 6 records
- Becton Dickinson & Company · 1 event, 4 records, 4 Class I
- Boston Scientific Corporation · 1 event, 2 records
FDA root cause
- Device Design · 23
- Labeling mix-ups · 12
- Under Investigation by firm · 8
- Software design · 8
- Process control · 6
- Labeling design · 6
- Process design · 4
- Process change control · 3
Class I recalls (6)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- 6F Taiga Guiding CatheterClass IZ-1207-2018 · Medtronic Inc · posted Mar 29, 2018 · Terminated
Complaints regarding 6F Taiga guide catheter tip detachment and/or tip splitting.
- Z-1139-2018 · Becton Dickinson & Company · posted Mar 22, 2018 · Terminated
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry…
- Z-1140-2018 · Becton Dickinson & Company · posted Mar 22, 2018 · Terminated
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry…
- Z-1141-2018 · Becton Dickinson & Company · posted Mar 22, 2018 · Terminated
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry…
- Z-1142-2018 · Becton Dickinson & Company · posted Mar 22, 2018 · Terminated
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry…
- Z-0194-2018 · Monteris Medical Corp · posted Mar 20, 2018 · Terminated
The firm received complaints that involved an unanticipated interaction between a NeuroBlate system and 1.5 Tesla (1.ST) MRI systems.
Class II recalls (75)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2472-2018 · Siemens Medical Solutions USA, Inc · posted Mar 30, 2018 · Terminated
A potential risk of unnecessary radiation exposure due to a software issue
- SOMATOM Emotion 16 (10165977) ↗Class IIZ-2473-2018 · Siemens Medical Solutions USA, Inc · posted Mar 30, 2018 · Terminated
A potential risk of unnecessary radiation exposure due to a software issue
- Z-2474-2018 · Siemens Medical Solutions USA, Inc · posted Mar 30, 2018 · Terminated
A potential risk of unnecessary radiation exposure due to a software issue
- Z-2475-2018 · Siemens Medical Solutions USA, Inc · posted Mar 30, 2018 · Terminated
A potential risk of unnecessary radiation exposure due to a software issue
- SOMATOM Scope ( Model 10967666)Class IIZ-2476-2018 · Siemens Medical Solutions USA, Inc · posted Mar 30, 2018 · Terminated
A potential risk of unnecessary radiation exposure due to a software issue
- Z-2477-2018 · Siemens Medical Solutions USA, Inc · posted Mar 30, 2018 · Terminated
A potential risk of unnecessary radiation exposure due to a software issue
- Z-2478-2018 · Siemens Medical Solutions USA, Inc · posted Mar 30, 2018 · Terminated
A potential risk of unnecessary radiation exposure due to a software issue
- Z-1201-2018 · Johnson & Johnson Consumer, Inc. · posted Mar 27, 2018 · Terminated
Johnson & Johnson Consumer Inc. is initiating a voluntary, Class II recall at the retail level in the U.S. of select COACH Self-Adhering Sports Wraps, BAND-AID Brand First Aid Products SECURE-FLEX Wraps and BAND-AID Brand First Aid Products HURT-FREE Wraps…
- Z-1202-2018 · Johnson & Johnson Consumer, Inc. · posted Mar 27, 2018 · Terminated
Johnson & Johnson Consumer Inc. is initiating a voluntary, Class II recall at the retail level in the U.S. of select COACH Self-Adhering Sports Wraps, BAND-AID Brand First Aid Products SECURE-FLEX Wraps and BAND-AID Brand First Aid Products HURT-FREE Wraps…
- Z-1203-2018 · Johnson & Johnson Consumer, Inc. · posted Mar 27, 2018 · Terminated
Johnson & Johnson Consumer Inc. is initiating a voluntary, Class II recall at the retail level in the U.S. of select COACH Self-Adhering Sports Wraps, BAND-AID Brand First Aid Products SECURE-FLEX Wraps and BAND-AID Brand First Aid Products HURT-FREE Wraps…
- Z-1204-2018 · Johnson & Johnson Consumer, Inc. · posted Mar 27, 2018 · Terminated
Johnson & Johnson Consumer Inc. is initiating a voluntary, Class II recall at the retail level in the U.S. of select COACH Self-Adhering Sports Wraps, BAND-AID Brand First Aid Products SECURE-FLEX Wraps and BAND-AID Brand First Aid Products HURT-FREE Wraps…
- Z-1205-2018 · Johnson & Johnson Consumer, Inc. · posted Mar 27, 2018 · Terminated
Johnson & Johnson Consumer Inc. is initiating a voluntary, Class II recall at the retail level in the U.S. of select COACH Self-Adhering Sports Wraps, BAND-AID Brand First Aid Products SECURE-FLEX Wraps and BAND-AID Brand First Aid Products HURT-FREE Wraps…
- Z-1206-2018 · Johnson & Johnson Consumer, Inc. · posted Mar 27, 2018 · Terminated
Johnson & Johnson Consumer Inc. is initiating a voluntary, Class II recall at the retail level in the U.S. of select COACH Self-Adhering Sports Wraps, BAND-AID Brand First Aid Products SECURE-FLEX Wraps and BAND-AID Brand First Aid Products HURT-FREE Wraps…
- Z-1236-2018 · Zimmer Biomet, Inc. · posted Mar 27, 2018 · Terminated
Certain lots of the Phoenix Tibia Nail 3.5mm Inserter instruments are being recalled due to reports of fracturing and remaining in the device implant.
- Z-1184-2018 · Siemens Healthcare Diagnostics, Inc. · posted Mar 26, 2018 · Terminated
An increase in the number of failed calibration events or negative bias with Quality Control (QC) and patient samples when using Dimension Vista LOCI 8 Calibrator lot 7CD076 to calibrate Estradiol (E2) Assay.
- Z-1200-2018 · Siemens Medical Solutions USA, Inc. · posted Mar 26, 2018 · Terminated
The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment.
- Z-1170-2018 · St. Jude Medical, Inc. · posted Mar 23, 2018 · Terminated
The firm received complaints of error messages that occurred during routine impedance checks on Proclaim DRG IPGs, model 3664. Some complaints were also associated with transient over stimulation which created discomfort for the patients.
- Z-1173-2018 · Biomet, Inc. · posted Mar 23, 2018 · Terminated
During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery.…
- Z-1152-2018 · Arrow International Inc · posted Mar 21, 2018 · Terminated
Certain lots may be missing information at the bottom of the lid including the lot number, expiration date, and Unique Device Identification (UDI) information.
- Z-1148-2018 · Siemens Medical Solutions USA, Inc · posted Mar 20, 2018 · Terminated
Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.
- Z-1150-2018 · Siemens Medical Solutions USA, Inc · posted Mar 20, 2018 · Terminated
Certain systems may have leaking of the plastic coolant couplings, the liquid can enter into the flat panel detector leading to a failure of the detector. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or…
- Z-1172-2018 · Getinge US Sales LLC · posted Mar 20, 2018 · Terminated
A probable mislabeling of the product: a different-sized product is referenced on the product packaging label (i.e. 2.5 cm (width) x 7.6 cm (length)) while the actual product inside the box is 2.5 cm (width) x 15 cm (length).
- 1) Sigma HP Cemented Trochlea Size 1 Narrow Left, Catalog Number: 102403100, GTIN: 10603295001614Class IIZ-1219-2018 · DePuy Orthopaedics, Inc. · posted Mar 20, 2018 · Terminated
The SIGMA HP PFJ Cemented Trochlear Implants, a standalone component of the partial knee system is being recalled based on elevated revision rates observed as part of the company s post market surveillance process.
- Z-1278-2018 · Fresenius Medical Care Renal Therapies Group, LLC · posted Mar 20, 2018 · Terminated
When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will…
- Z-1303-2018 · Siemens Medical Solutions USA, Inc · posted Mar 20, 2018 · Terminated
A software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving corrected/modified values. Outdated information…
- Z-1317-2018 · Smith & Nephew, Inc. · posted Mar 20, 2018 · Terminated
One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.
- Z-1337-2018 · Gyrus ACMI, Inc. · posted Mar 20, 2018 · Open, Classified
"Chattering" when cutting burrs are used in bone.
- ZYMUTEST HIA IgGAM.Class IIZ-1062-2018 · Hyphen BioMed · posted Mar 15, 2018 · Terminated
A voluntary recall is initiated for ZYMUTEST HIA lgGAM - Ref RK040D - Lot F1700415 and ZYMUTEST HIA MonoStrip lgGAM - Ref RK041D - Lot F1700416 because they may produce false positive results.
- ZYMUTEST HIA MonoStrip IgGAM.Class IIZ-1063-2018 · Hyphen BioMed · posted Mar 15, 2018 · Terminated
A voluntary recall is initiated for ZYMUTEST HIA lgGAM - Ref RK040D - Lot F1700415 and ZYMUTEST HIA MonoStrip lgGAM - Ref RK041D - Lot F1700416 because they may produce false positive results.
- Z-1073-2018 · Smiths Medical ASD Inc. · posted Mar 15, 2018 · Terminated
Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the…
- Z-1074-2018 · Smiths Medical ASD Inc. · posted Mar 15, 2018 · Terminated
Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the…
- Z-1191-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Certain packages labeled as size 7.5 Extended Offset Reduced Neck length hip implant, actually contained a Size 4 Standard Reduced Neck length taper hip implant.
- Z-1192-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Certain packages labeled as size 7.5 Extended Offset Reduced Neck length hip implant, actually contained a Size 4 Standard Reduced Neck length taper hip implant.
- Z-1194-2018 · Arrow International Inc · posted Mar 14, 2018 · Terminated
Certain lots of peripheral catheter systems may be missing the safety latch, intended to minimize the risk of sharps injuries.
- Z-1196-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
The AIM Tibial Nails failed a clinical analysis where the surgical revision rate exceeded the acceptance criteria in the Device Performance Review protocol for the implant system.
- Z-1197-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
The AIM Tibial Nails failed a clinical analysis where the surgical revision rate exceeded the acceptance criteria in the Device Performance Review protocol for the implant system.
- ACE Trochanteric Nail System ↗Class IIZ-1225-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1226-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1227-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1228-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1229-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1230-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1231-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1232-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1233-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1234-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- ACE Trochanteric Nail System ↗Class IIZ-1235-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- Z-1239-2018 · Zimmer Biomet, Inc. · posted Mar 14, 2018 · Terminated
One lot of humeral stem implants do not include the humeral bushing components.
- Z-1292-2018 · GE Healthcare, LLC · posted Mar 14, 2018 · Terminated
if the cables of the Monica Novii System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).
- Z-1293-2018 · GE Healthcare, LLC · posted Mar 14, 2018 · Terminated
if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).
- Z-1582-2018 · Radiometer Medical ApS · posted Mar 14, 2018 · Terminated
There is a potential problem relating to the AQURE System, versions 2.3.0 and 2.3.1, that may result in patient mix-up.
- Z-0897-2018 · Boston Scientific Corporation · posted Mar 5, 2018 · Terminated
Certain Auriga consoles have incorrectly assembled trigger wire components. The trigger wire functions in the start-up of the console. An incorrectly assembled trigger wire may result in error codes or failure to start the console.
- Z-0898-2018 · Boston Scientific Corporation · posted Mar 5, 2018 · Terminated
Certain Auriga consoles have incorrectly assembled trigger wire components. The trigger wire functions in the start-up of the console. An incorrectly assembled trigger wire may result in error codes or failure to start the console.
- Z-0936-2018 · Medtronic Minimally Invasive Therapies Group · posted Mar 5, 2018 · Terminated
Incorrect packaging. In one lot of Argyle Suction Tubing, the pouch may be incorrectly labeled as the Salem Sump Dual Lumen Stomach Tube. This product may have been distributed within the case of Argyle Suction Tubing.
- Z-1134-2018 · Smiths Medical ASD Inc. · posted Mar 5, 2018 · Terminated
Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an…
- Z-1135-2018 · Smiths Medical ASD Inc. · posted Mar 5, 2018 · Terminated
Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an…
- Z-1136-2018 · Smiths Medical ASD Inc. · posted Mar 5, 2018 · Terminated
Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an…
- Z-1158-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1159-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1160-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1161-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1162-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1163-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1164-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1165-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1166-2018 · Arrow International Inc · posted Mar 5, 2018 · Terminated
Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.
- Z-1189-2018 · Zimmer Dental Inc · posted Mar 5, 2018 · Terminated
Two lots of the Tapered Screw-Vent Implants, with different outer thread diameters, were inadvertently commingled during the manufacturing process and may be mislabeled.
- Z-1190-2018 · Zimmer Dental Inc · posted Mar 5, 2018 · Terminated
Two lots of the Tapered Screw-Vent Implants, with different outer thread diameters, were inadvertently commingled during the manufacturing process and may be mislabeled.
- Z-1241-2018 · Encore Medical, Lp · posted Mar 5, 2018 · Terminated
Broken or missing screws and retaining clips. This may be due to screw tolerance and material strength of the trial.
- Z-1927-2018 · Getinge US Sales LLC · posted Mar 5, 2018 · Terminated
Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to cracking of the spring arms at the weld seams that has the potential to develop into…
- Z-0850-2018 · Philips Electronics North America Corporation · posted Mar 3, 2018 · Terminated
Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the…
- Z-0895-2018 · Instrumentation Laboratory Co. · posted Mar 3, 2018 · Terminated
Instrumentation Laboratory Co. received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results.
- Z-0817-2018 · Circulatory Technology Inc · posted Mar 1, 2018 · Terminated
The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.
- Z-0828-2018 · Stryker Communications · posted Mar 1, 2018 · Terminated
The device could have a missing spanner nut and snap ring which could potentially result in the equipment falling.
- Recall Expanded to add these models: Bigger Better-Bladder cardiopulmonary bypass blood reservoirClass IIZ-1171-2018 · Circulatory Technology Inc · posted Mar 1, 2018 · Terminated
The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1013-2018 · Stryker Medical Division of Stryker Corporation · posted Mar 14, 2018 · Terminated
The T/Pump Operations Manual and Maintenance Manual were incorrectly revised to add the sentence, Temperature therapy can also be for heat conservation for normothermia maintenance to the Indications for Use statement.