ultraView SISH DNP Detection Kit, Ventana Part Number… recallZ-0739-2018 · Ventana Medical Systems Inc · initiated Dec 19, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ventana Medical Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
FDA root cause category: Process control
What the company is doing
On Dec 19th all Roche affiliates (the direct consignees) worldwide were notified via Safety Board Notification (SBN) and subsequent Field Safety Notification (EX-US) and UMDC TD00090 (US) SBN and FSN revised Dec 20th due to 3 affected part numbers missed on original letter. The instructions are as follows: "Actions taken by Roche Diagnostics All affected product has been placed on hold. Ventana has reworked all product in its inventory and is in the process of manufacturing new lots for distribution and replacement of customer affected kits. Customers will be notified when corrected product is available. Actions to be taken by the customer/user Affected kits may continue to be used by customers until corrected product is available, however Ventana is mandating that the affected IHC detection kits (iView, ultraView, OptiView) must only be used in conjunction with same-slide controls. These controls must be appropriate for each assay and capable of detecting false negative results due to a complete or partial reagent dispense failure. CINtec PLUS Cytology does not have the capacity for same slide controls, so system-level controls should be maintained. ultraView SISH Detection is used for HER2 analysis, and employs internal positive controls; no external control is required. For assays that directly relate to clinical therapy decision making (e.g. ER/PR, HER2, ALK, etc.), it is additionally important to select a same slide positive control tissue with sufficient sensitivity to detect small decreases in intensity that may cause borderline positive cases to appear as negative (e.g. HER2 2+ vs. 1+). Although the use of same slide controls is considered optimal laboratory practice and strongly recommended by Ventana, customers may revert to standard run controls once non-impacted product is received. In order to reduce the risk of this issue impacting patient care, customers not using same slide controls as a standard practice should follow their
Recall details
- Recall number
- Z-0739-2018
- Status
- Terminated
- Initiated by the company
- Dec 19, 2017
- Posted by the FDA
- Feb 27, 2018
- Recalling firm
- Ventana Medical Systems Inc
- Quantity
- 358 units
- Product code
- PPM
- Distribution
- Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
Product description
ultraView SISH DNP Detection Kit, Ventana Part Number 800-098, Roche GMMI 05907136001 in vitro diagnostic
Code information (models, lots, serial numbers)
UDI 04015630972135, Lot Number: Y17990
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Hematoxylin II 05277965001Ventana Medical Systems, Inc. · HEMATOXYLIN2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 78942).
- Z-0729-2018 · TerminatedOptiView DAB IHC Detection Kit, Ventana Part Number 760-700, Roche GMMI 06396500001 in vitro diagnostic
- Z-0730-2018 · TerminatedultraView Universal DAB Detection Kit, Ventana part Number 760-500, Roche GMMI 05269806001 in vitro diagnostic
- Z-0731-2018 · TerminatediView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in vitro diagnostic
- Z-0733-2018 · TerminatedultraView SISH DNP Detection Kit US, Ventana Part Number 760-098, Roche GMMI 05572037001 in vitro diagnostic
- Z-0736-2018 · TerminatedOptiView Amplification Kit, Ventana Part Number 760-099, Roche GMMI 06396518001 in vitro diagnostic
- Z-0737-2018 · TerminatedOptiView Amplification Kit (250 Test), Ventana Part Number 860-099, Roche GMMI 06718663001 in vitro diagnostic
- Z-0738-2018 · TerminatedHematoxylin II, Ventana Part Number 790-22087, Roche GMMI 05277965001 in vitro diagnostic
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page