Medical device recalls: February 2018
The FDA posted 68 medical device recall records in February 2018, from 28 recall events by 24 companies. 21 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Aesculap Implant Systems LLC · 3 events, 3 records
- DePuy Orthopaedics, Inc. · 2 events, 3 records
- Philips Electronics North America Corporation · 2 events, 3 records
- Arrow International Inc · 1 event, 12 records
- Ventana Medical Systems Inc · 1 event, 12 records, 12 Class I
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · 1 event, 8 records, 8 Class I
- Beaver-Visitec International Inc. · 1 event, 5 records
- Philips Medical Systems (Cleveland) Inc · 1 event, 5 records
FDA root cause
- Process control · 31
- Process design · 8
- Under Investigation by firm · 7
- Device Design · 7
- Nonconforming Material/Component · 3
- Mixed-up of materials/components · 3
- No Marketing Application · 2
- Use error · 1
Class I recalls (21)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0729-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0730-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- iView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in vitro diagnosticClass IZ-0731-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0732-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0733-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0734-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0735-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- OptiView Amplification Kit, Ventana Part Number 760-099, Roche GMMI 06396518001 in vitro diagnosticClass IZ-0736-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0737-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0738-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0739-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0740-2018 · Ventana Medical Systems Inc · posted Feb 27, 2018 · Terminated
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Z-0489-2018 · Philips North America, LLC · posted Feb 8, 2018 · Terminated
Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
- Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Amplia MRI. Labeled as the following:Class IZ-0582-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 1, 2018 · Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
- Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Compia MRI. Labeled as the following:Class IZ-0583-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 1, 2018 · Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
- Z-0584-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 1, 2018 · Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
- Z-0585-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 1, 2018 · Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
- Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Claria MRI. Labeled as the following:Class IZ-0586-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 1, 2018 · Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
- Z-0587-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 1, 2018 · Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
- Z-0588-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 1, 2018 · Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
- Z-0589-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 1, 2018 · Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
Class II recalls (45)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1012-2018 · DePuy Orthopaedics, Inc. · posted Feb 27, 2018 · Terminated
The screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays.
- Z-0716-2018 · Philips Electronics North America Corporation · posted Feb 24, 2018 · Terminated
There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating…
- Z-0703-2018 · Oakworks Inc · posted Feb 23, 2018 · Terminated
The foot control used on the OAKWORKS Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table.
- Z-0705-2018 · Teleflex Medical · posted Feb 23, 2018 · Terminated
The pouch may contain the incorrect Rusch Green Lite product size. Pouches that are labeled as containing RUSCH GREEN LITE MAC 1 may instead contain RUSCH GREEN LITE MAC 2.
- Z-0712-2018 · Atrium Medical Corporation · posted Feb 23, 2018 · Terminated
An incorrect ProLite Mesh configuration was packaged as product code 1010306-06, with lot number 412298. ProLite Mesh manufactured as product code 1010306-04 was inadvertently placed inside the pouch of packages labeled with product code 1010306-06 and…
- Z-0683-2018 · Philips Electronics North America Corporation · posted Feb 22, 2018 · Terminated
Inform users of correct placement of the disposable adhesive pad on the QCPR meter and inform users of injuries associated with proper performance of CPR. As required by Consent Decree of Permanent Injunction No. 1:17-cv-11955-DJC, entered October 31,…
- Z-0684-2018 · Philips Electronics North America Corporation · posted Feb 22, 2018 · Terminated
Inform users of correct placement of the disposable adhesive pad on the QCPR meter and inform users of injuries associated with proper performance of CPR. As required by Consent Decree of Permanent Injunction No. 1:17-cv-11955-DJC, entered October 31,…
- Innova Self-Expanding Stent System, 200 mm. Labeled with Material Number and description as follows:Class IIZ-0685-2018 · Boston Scientific Corporation · posted Feb 22, 2018 · Open, Classified
Stent possibly unable to be fully released from the delivery system.
- Innova Self-Expanding Stent System, 180mm. Labeled with Material Number and description as follows:Class IIZ-0686-2018 · Boston Scientific Corporation · posted Feb 22, 2018 · Open, Classified
Stent possibly unable to be fully released from the delivery system.
- Z-0689-2018 · Philips Consumer Lifestyle · posted Feb 22, 2018 · Terminated
Device may provide inaccurate measurements.
- Z-0643-2018 · Pentax of America Inc · posted Feb 20, 2018 · Terminated
The duodenoscopes are being recalled in order to replace the forceps elevator mechanism, the O-rings, and the distal end covering to be consistent with the updated design as well as provide an updated periodic inspection as part of the Operation Manual in…
- Z-0647-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Multi Lumen Central Venous Catheterization Kit with Blue FlexTip Catheter and Sharps Safety FeaturesClass IIZ-0648-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0649-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0650-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0651-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0652-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0653-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0654-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0655-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0656-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0657-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0658-2018 · Arrow International Inc · posted Feb 20, 2018 · Terminated
These finished good kits may contain the incorrect Springwire Guide (SWG).
- Z-0829-2018 · Fresenius Medical Care Renal Therapies Group, LLC · posted Feb 20, 2018 · Terminated
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also…
- Z-0636-2018 · Covidien LLC · posted Feb 16, 2018 · Terminated
Potential for failure of the device safety interlock. The safety interlock prevents an empty single-use loading unit from being fired a second time.
- Z-0623-2018 · Aesculap Implant Systems LLC · posted Feb 15, 2018 · Terminated
The ELAN 4 Air System was released to the market place prior to an authorized FDA 510K clearance. This issue does not affected sales outside the United States (OUS).
- Z-0624-2018 · Aesculap Implant Systems LLC · posted Feb 15, 2018 · Terminated
The JS Series SterilContainer S2 System was released to the market place prior to an authorized FDA 510K clearance. No other Aesculap SterilContainer System products are impacted by this recall.
- Z-0629-2018 · Cincinnati Sub-Zero Products LLC · posted Feb 13, 2018 · Terminated
Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.
- Z-0742-2018 · Angiodynamics, Inc. · posted Feb 13, 2018 · Terminated
Product was placed into distribution prior to completion of all required post sterilization release activities.
- Z-0772-2018 · Steris Corporation · posted Feb 13, 2018 · Terminated
Correction to update sterilizer software to prohibit the use of expired sterilant prior to the start of a sterilization cycle.
- Z-0591-2018 · Aesculap Implant Systems LLC · posted Feb 12, 2018 · Terminated
The NS357R Vega PS Tibia Trial/Preparation Plateau size T4 may be incorrectly labeled. If the incorrect size trial is used the potential risk is implantation of the incorrect implant size which may lead to injury of ligaments, pain and possible revision…
- Z-0593-2018 · DePuy Orthopaedics, Inc. · posted Feb 12, 2018 · Terminated
Incorrect device in the package. A package that was labeled as the CORAIL COXA VARA HIGH OFFSET STEM Collared Size 9 (Part No. 3L93709 / Lot No. 5291990), contained the CORAIL HIGH OFFSET STEM Collarless SIZE 14 (Part No. L20314 / Lot No. 5292130).
- Z-0594-2018 · DePuy Orthopaedics, Inc. · posted Feb 12, 2018 · Terminated
Incorrect device in the package. A package that was labeled as the CORAIL COXA VARA HIGH OFFSET STEM Collared Size 9 (Part No. 3L93709 / Lot No. 5291990), contained the CORAIL HIGH OFFSET STEM Collarless SIZE 14 (Part No. L20314 / Lot No. 5292130).
- Z-0536-2018 · Bard Access Systems Inc. · posted Feb 6, 2018 · Terminated
Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction.
- BrightView X, Model #882478 ↗Class IIZ-0510-2018 · Philips Medical Systems (Cleveland) Inc · posted Feb 3, 2018 · Terminated
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning…
- BrightView XCT Model 882482 ↗Class IIZ-0511-2018 · Philips Medical Systems (Cleveland) Inc · posted Feb 3, 2018 · Terminated
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning…
- BrightView Model 882480 ↗Class IIZ-0512-2018 · Philips Medical Systems (Cleveland) Inc · posted Feb 3, 2018 · Terminated
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning…
- Precedence 6 Model # 882351 ↗Class IIZ-0513-2018 · Philips Medical Systems (Cleveland) Inc · posted Feb 3, 2018 · Terminated
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning…
- Precedence 16 Model # 882350 ↗Class IIZ-0514-2018 · Philips Medical Systems (Cleveland) Inc · posted Feb 3, 2018 · Terminated
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning…
- Z-0500-2018 · Beaver-Visitec International Inc. · posted Feb 1, 2018 · Terminated
The protective shield was retracted, and the blade was exposed.
- Z-0501-2018 · Beaver-Visitec International Inc. · posted Feb 1, 2018 · Terminated
The protective shield was retracted, and the blade was exposed.
- Z-0502-2018 · Beaver-Visitec International Inc. · posted Feb 1, 2018 · Terminated
The protective shield was retracted, and the blade was exposed.
- Z-0503-2018 · Beaver-Visitec International Inc. · posted Feb 1, 2018 · Terminated
The protective shield was retracted, and the blade was exposed.
- Z-0504-2018 · Beaver-Visitec International Inc. · posted Feb 1, 2018 · Terminated
The protective shield was retracted, and the blade was exposed.
- Z-0609-2018 · Zimmer Biomet, Inc. · posted Feb 1, 2018 · Terminated
Cut resistant surgical glove liners lack sterility assurance.
Class III recalls (2)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0899-2018 · Degania Silicone, Ltd. · posted Feb 20, 2018 · Terminated
Sterile pouches are missing labels on them. The boxes themselves are properly labeled. The issue is related to the production lot P1607442 only.
- Z-0537-2018 · Medtronic Neuromodulation · posted Feb 6, 2018 · Terminated
There is a possibility that specific lots of catheters that have recently been distributed may be at risk for an increased potential for kinking at the proximal end where the catheter connects to the drug infusion pump.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.