Skip to content

Emprint recallZ-0561-2018 · Covidien LLC · initiated Aug 7, 2017

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.

FDA root cause category: Device Design

What the company is doing

All consignees were notified via Federal Express or certified mail on August 7, 2017, and the letter informs customers of the possible trocar tip disengagement and the actions they should take. Customers are requested to segregate and return affected product and acknowledge receipt and understanding of the Urgent Medical Device Recall Notice.

Recall details

Recall number
Z-0561-2018
Status
Open, Classified
Initiated by the company
Aug 7, 2017
Recalling firm
Covidien LLC
Product code
NEY
FDA submissions
K133821
Distribution
Nationwide including PR, Canada, China

Product description

Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2

Code information (models, lots, serial numbers)

UDI 10884521706606, Lot codes: S6MG007PX,S6MG010PX,S6MG016PX,S7AG002PX,S7BG005X,S7BG009X,S7DG004X,S7FG006X

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 79075).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page