Skip to content

Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400 recallZ-0895-2018 · Instrumentation Laboratory Co. · initiated Aug 2, 2017

Class II Terminated All recalls posted in March 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Instrumentation Laboratory Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Instrumentation Laboratory Co. received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results.

FDA root cause category: Under Investigation by firm

What the company is doing

On August 2, 2017, Instrumentation Laboratory sent an Urgent Medical Device Removal letters dated August 2, 2017 to their customers. On October 5, 2017 Instrumentation Laboratory expanded their recall adding 3 additional lots to a second Urgent Medical Device Removal letter. On February 26, 2018, Instrumentation Laboratory expanded their recall a 3rd time to include 1 additional lot to a third Urgent Medical Device Removal letter. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete a Mandatory Tracking Response form. For questions contact your Customer Service Representative at 1-800-955-9525, option #1.

Recall details

Recall number
Z-0895-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 2, 2017
Posted by the FDA
Mar 3, 2018
Recalling firm
Instrumentation Laboratory Co.
Quantity
14,894 kits
Product code
GJS
FDA submissions
K122584
Distribution
Worldwide Distribution - US Nationwide

Product description

Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400

Code information (models, lots, serial numbers)

Lots: N1166235, N0177760, N0278358, N0478057, N0479504 & N0278177.

Other devices under the same FDA submission

1 device cleared under K122584 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page