Skip to content

ZYMUTEST HIA MonoStrip IgGAM recallZ-1063-2018 · Hyphen BioMed · initiated Nov 9, 2011

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hyphen BioMed or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A voluntary recall is initiated for ZYMUTEST HIA lgGAM - Ref RK040D - Lot F1700415 and ZYMUTEST HIA MonoStrip lgGAM - Ref RK041D - Lot F1700416 because they may produce false positive results.

FDA root cause category: Other

What the company is doing

The manufacturer, HYPHEN BioMed, initiated a voluntary recall of the two lots via email. Distributors and end users are being notified and instructed to: 1) acknowledge receipt and understanding of notification, and 2) destroy affected products remaining in stock. For further questions, please call (513) 770-1993.

Recall details

Recall number
Z-1063-2018
Status
Terminated
Initiated by the company
Nov 9, 2011
Posted by the FDA
Mar 15, 2018
Recalling firm
Hyphen BioMed
Quantity
30
Product code
LCO
FDA submissions
K071255
Distribution
Worldwide Distribution - US Distribution and to the countries of : Austria and France.

Product description

ZYMUTEST HIA MonoStrip IgGAM. Model Number: RK041D. Qualitative screening assay for the detection of heparin-dependent antibodies (IgG, IgM and IgA isotypes by ELISA. 32 Tests.

Code information (models, lots, serial numbers)

UDI: 03663537046810. Lot Number: F1700416.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K071255 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 79269).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page