Zimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem… recallZ-1192-2018 · Zimmer Biomet, Inc. · initiated Mar 1, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain packages labeled as size 7.5 Extended Offset Reduced Neck length hip implant, actually contained a Size 4 Standard Reduced Neck length taper hip implant.
FDA root cause category: Labeling mix-ups
What the company is doing
RISK MANAGER: 1. Review this notification and notify affected personnel of the contents. 2. Quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Certificate of Acknowledgement and send to [email protected]. DISTRIBUTOR: 1. Review this notification and notify affected team members of the contents. 2. Quarantine affected product in your inventory. 3. Return affected product from your distributorship and from affected hospitals. a. Complete Inventory Return Certification Form and send to [email protected] Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected].
Recall details
- Recall number
- Z-1192-2018
- Status
- Terminated
- Initiated by the company
- Mar 1, 2018
- Posted by the FDA
- Mar 14, 2018
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 24
- Product code
- LPH
- FDA submissions
- K060040
- Distribution
- Distributed in the following U.S. states: AZ, CA, CT, IN, MD, MI, MN, NJ, NY, OR, WA, and Puerto Rico. Distributed in China and India.
Product description
Zimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem 12/14 Neck Taper Plasma Sprayed Press-Fit Cementless Size 4 Standard Reduced Neck Length; Part Number: 00-7711-004-10
Code information (models, lots, serial numbers)
Lot Number: 63716613, UDI: (01)00889024131606(17)270731(10)63716613; Lot Number: 63716614, UDI: (01)00889024131606(17)279731(10)63716614
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA 00-7711-004-10Zimmer, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 recall4 reports
Other devices under the same FDA submission
2 devices cleared under K060040 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA 00-7711-004-40Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 recall4 reports
- NA 00-7711-005-40Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 79406).
- Z-1191-2018 · TerminatedZimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem 12/14 Neck Taper Plasma Sprayed Press-Fit Cementless Size 7.5 Extended Offset Reduced Neck Length; Part Number: 00-7711-0…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page