6F Taiga Guiding Catheter recallZ-1207-2018 · Medtronic Inc · initiated Sep 26, 2017
Class I Terminated All recalls posted in March 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Complaints regarding 6F Taiga guide catheter tip detachment and/or tip splitting.
FDA root cause category: Nonconforming Material/Component
What the company is doing
A recall notification was sent to all costumers in September 2017. Customers actions to be done: Identify and quarantine all unused affected products as listed in your inventory. Share notification with others who need to be aware where affected product may have potentially been transferred.
Recall details
- Recall number
- Z-1207-2018
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Sep 26, 2017
- Posted by the FDA
- Mar 29, 2018
- Recalling firm
- Medtronic Inc
- Quantity
- 15,923
- Product code
- DQY
- FDA submissions
- K083422
- Distribution
- Japan
Product description
6F Taiga Guiding Catheter
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
Lot/Serial Number's: 0008104881, 0008185945, 0008448845, 0008498609, 0008591678, 0008603130, 0008128237, 0008185937, 0008359341, 0008381347, 0008520748, 0008572204, 0008584614, 00 887489, 0008132544, 0008143685, 0008153935, 0008176874, 0008181991, 0008189749, 0008195422, 0008220305, 0008399842, 0008411526, 0008479077, 0008498645, 0008510155, 0008560305, 0008588040, 0008609897, 0008614083, 0008092512, 0008117850, 0008171020, 0008173080, 0008185946, 0008195426, 0008250618, 0008374572, 0008399835, 0008411525, 0008510141, 0008584625, 0008589932, 0008609899, 0008084699, 0008092511, 0008143683, 0008156792, 0008169119, 0008191347, 0008448842, 0008479041, 0008498608, 0008527623, 0008090683, 0008121917, 0008128241, 0008148608, 0008169116, 0008191339, 0008401589, 0008503947, 0008510133, 0008580930, 0008580935, 0008580939, 0008589915, 0008121924, 0008132534, 0008150901, 0008401588, 0008418561, 0008441398, 0008448854, 0008503956, 0008588042, 0008603140, 0008614065, 0008087481, 0008090682, 0008132516, 0008498612, 0008510146, 0008572210, 0008580937, 0008588036, 0008603136, 0008123258, 0008165872, 0008490748, 0008498648, 0008143680, 0008399813, 0008090684, 0008128251, 0008169124, 0008181983, 0008185942, 0008436912, 0008448835, 0008458396, 0008471810, 0008481006, 0008492425, 0008498642, 0008503965, 0008588039, 0008150909, 0008381364, 0008440030, 0008448837, 0008492428, 0008580933, 0008580936, 0008580940, 0008603142, 0008236362, 0008506201, 0008440025, 0008158841, 0008589929, 0008117834, 0008140160, 0008140158, 0008584612, 0008140153, 0008588035, 0008087497, 0008117854, 0008171017, 0008181987, 0008185949, 0008189738, 0008195429, 0008227448, 0008240516, 0008349924, 0008374578, 0008407407, 0008428073, 0008440040, 0008492417, 0008506206, 0008589931, 0008600840, 0008603138, 0008614075, 0008090688, 0008135876, 0008169120, 0008173078, 0008180039, 0008498627, 0008510129, 0008572207, 0008579087, 0008579090, 0008591671, 0008609888, 0008117843, 0008132532, 0008150894, 0008233403, 0008381360, 0008416713, 0008479070, 0008503958, 0008525960, 0008555947, 0008586326, 0008600838, 0008606084, 0008609893, 0008614068, 0008169125, 0008284846, 0008361412, 0008436904, 0008586324, 0008609886, 0008087503, 0008098497, 0008108533, 0008123259, 0008132523, 0008132525, 0008144679, 0008144680, 0008171009, 0008177970, 0008193508, 0008218499, 0008225407, 0008231701, 0008242807, 0008329311, 0008342126, 0008349912, 0008365450, 0008374574, 0008388047, 0008407404, 0008418572, 0008428085, 0008454371, 0008456337, 0008479051, 0008479057, 0008503968, 0008518382, 0008529633, 0008546593, 0008564278, 0008588045, 0008591673, 0008603126, 0008614073, 0008623891, 0008117853, 0008165878, 0008181992, 0008185944, 0008195428, 0008312796, 0008351878, 0008418565, 0008479074, 0008510154, 0008564276, 0008584618, 0008589912, 0008603129, 0008606081, 0008611517, 0008102481, 0008128253, 0008135881, 0008150887, 0008181978, 0008181986, 0008185943, 0008189756, 0008227450, 0008242808, 0008340383, 0008349919, 0008420672, 0008458398, 0008492395, 0008506212, 0008525964, 0008551253, 0008572205, 0008589917, 0008603128, 0008609901, 0008117851, 0008132537, 0008175028, 0008185950, 0008216634, 0008403383, 0008411523, 0008490750, 0008498619, 0008510157, 0008582358, 0008582365, 0008591677, 0008121922, 0008143687, 0008171022, 0008189732, 0008189735, 0008436903, 0008448844, 0008481008, 0008498655, 0008510162, 0008588038, 0008090677, 0008102474, 0008135875, 0008143672, 0008164709, 0008169123, 0008498630, 0008520751, 0008572209, 0008586333, 0008589923, 0008606085, 0008611515, 0008572203, 0008591669, 0008609882, 0008102479, 0008185939, 0008132520, 0008169118, 0008483192, 0008498601, 0008503941, 0008087476, 0008102476, 0008483190, 0008242802, 0008471804, 0008506203, 0008603124, 0008381351, 0008164710, 0008418558, 0008510130, 0008164708, 0008346263, 0008503943, 0008546591, 0008564272, 0008175029, 0008205770, 0008407402, 0008411528, 0008498610, 0008591675, 0008227447, 0008312798, 0008448846, 0008492399, 0008202461, 0008555943, 0008087496, 0008117847, 0008456325, 0008468606, 0008479024, 0008506202, 0008535437, 0008544524, 0008572206, 0008580931, 0008586328, 0008598736, 0008609890, 0008087487, 0008106137, 0008376553, 0008586331, 0008368954, 0008488926, 0008284838, 0008084670, 0008128239, 0008169121, 0008416708, 0008428066, 0008440042, 0008479036, 0008492411, 0008498639, 0008527622, 0008586325, 0008614071, 0008092505,0 008117833, 0008128245, 0008242801, 0008370458, 0008510125, 0008102483, 0008132546, 0008148619, 0008164716, 0008176875, 0008189753, 0008191353, 0008220302, 0008329308, 0008349916, 0008409469, 0008428081, 0008441401, 0008479063, 0008503951, 0008572215, 0008589930, 0008600843, 0008606089, 0008614074, 0008535445, 0008128252, 0008140156, 0008148612, 0008153932, 0008171001, 0008191340, 0008221979, 0008436905, 0008454370, 0008468609, 0008488931, 0008492384, 0008500611, 0008506210, 0008515021, 0008525962, 0008537068, 0008579084, 0008579089, 0008591670, 0008606087, 0008092510, 0008104886, 0008128254, 0008135880, 0008150904, 0008164714, 0008175025, 0008189754, 0008191355, 0008218508, 0008242806, 0008374575, 0008401594, 0008420670, 0008436908, 0008479066, 0008500614, 0008531592, 0008572214, 0008591674, 0008603144, 0008614081, 0008132517, 0008140167, 0008143689, 0008148615, 0008150892, 0008181984, 0008189733, 0008189736, 0008242803, 0008436902, 0008448834, 0008458395, 0008468607, 0008479032, 0008490738, 0008500613, 0008503955, 0008520750, 0008560309, 0008591676, 0008405710, 0008411524, 0008440026, 0008448850, 0008481005, 0008510144, 0008546588, 0008580934, 0008586327, 0008614069, 0008087483, 0008090680, 0008132514, 0008140161, 0008148610, 0008165873, 0008242800, 0008448838, 0008468608, 0008498604, 0008537076, 0008579088, 0008589920, 0008611518, 0008090681, 0008102478, 0008128243, 0008140163, 0008164711, 0008175027, 0008117837, 0008095214, 0008121930, 0008165880, 0008175026, 0008189740, 0008191345, 0008195427, 0008374580, 0008399831, 0008424958, 0008441396, 0008448839, 0008492392, 0008498654, 0008503963, 0008535442, 0008572208, 0008586330, 0008586332, 0008589933, 0008598737, 0008603143, 0008143682, 0008171013, 0008208233, 0008448855, 0008492430, 0008506205, 0008537074, 0008562621, 0008589918, 0008603134, 0008611516, 0008084683, 0008095212, 0008121927, 0008132529, 0008156793, 0008169127, 0008189750, 0008195425, 0008216629, 0008242805, 0008351874, 0008368955, 0008374583, 0008418570, 0008441407, 0008479038, 0008506213, 0008535448, 0008551249, 0008572212, 0008589934, 0008600841, 0008611521, 0008090685, 0008128242, 0008143691, 0008165882, 0008177971, 0008399825, 0008440033, 0008471807, 0008492382, 0008506208, 0008537078, 0008582366, 0008588037, 0008603132, 0008609895, 0008090676, 0008600837, 0008621626, 0008090675, 0008479023, 0008503960, 0008548757, 0008560294, 0008582359, 0008582368, 0008603133, 0008128234, 0008490737, 0008498615, 0008584623, 0008202458, 0008441391, 0008498618, 0008140165, 0008560285, 0008140150, 0008488921, 0008104887, 0008479027, 0008490740, 0008143674, 0008397725, 0008092504, 0008153924, 0008096759, 0008492379, 0008582360, 0008102477, 0008359340, 0008582361, 0008611513, 0008150881, 0008498652, 0008525959, 0008564274, 0008092508, 0008102480, 0008132528, 0008148613, 0008164715, 0008171025, 0008181979, 0008181989, 0008185951, 0008191341, 0008195432, 0008205771, 0008212336, 0008218503, 0008231699, 0008236364, 0008347562, 0008352965, 0008374576, 0008390327, 0008401590, 0008405712, 0008409466, 0008428077, 0008479035, 0008485259, 0008503938, 0008518377, 0008560298, 0008584622, 0008589911, 0008603127, 0008606088, 0008611514, 0008614077, 0008535441, 0008128248, 0008132540, 0008150890, 0008169126, 0008181977, 0008181981, 0008185948, 0008191342, 0008208232, 0008221977, 0008240515, 0008312792, 0008357550, 0008374582, 0008399821, 0008411532, 0008428070, 0008456333, 0008479047, 0008490744, 0008503954. 0008518378. 0008584621, 0008589919. 0008609891. 0008616444. 0008537070, 0008095215, 0008117845, 0008171011, 0008189752, 0008420666, 0008448840, 0008454369, 0008483198, 0008503966, 0008518385, 0008584620, 0008589913, 0008603141, 0008128250, 0008135878, 0008185941, 0008191344, 0008498621, 0008548759, 0008579086, 0008579091, 0008589910, 0008087485, 0008104880, 0008510126, 0008436901, 0008087504, 0008095211, 0008117855, 0008143694, 0008164713, 0008171004, 0008191346, 0008220304, 0008374577, 0008381355, 0008399817, 0008407403, 0008420668, 0008436910, 0008479043, 0008492420, 0008503959, 0008515023, 0008562623, 0008582367, 0008584626, 0008589925, 0008600842, 0008614066, 0008535451, 0008095213, 0008165884, 0008181990, 0008189748, 0008202468, 0008216623, 0008333048, 0008397728, 0008479026, 0008490739, 0008510139, 0008546592, 0008560290, 0008584624, 0008589914, 0008606083, 0008403381, 0008503961, 0008582362, 0008582369, 0008169122, 0008555937, 0008580932, 0008203696, 0008121920, 0008117836, 0008132519, 0008158808, 0008169117, 0008312795, 0008498623, 0008518379, 0008582363, 0008586329, 0008603139, 0008092506, 0008128244, 0008150884, 0008170999, 0008090690, 0008104882, 0008121937, 0008148621, 0008165876, 0008175023, 0008191351, 0008195431, 0008225404, 0008240517, 0008390326, 0008401592, 0008416711, 0008456328, 0008483194, 0008492402, 0008506207, 0008562625, 0008588044, 0008535454, 0008121941, 0008143681, 0008148614, 0008153929, 0008158840, 0008173079, 0008189737, 0008195424, 0008416706, 0008440028, 0008448843, 0008458394, 0008490746, 0008503957, 0008510135, 0008589927, 0008611519, 000810 884, 0008128236, 0008150907, 0008164712, 0008181985, 0008418567, 0008436906, 0008456330, 0008492406, 0008510159, 0008560301, 0008586334, 0008589928, 0008606082, 0008614070, 0008121933, 0008132518, 0008150897, 0008175030, 0008208228, 0008335373, 0008344295, 0008510137, 0008582364, 0008589916, 0008606080, 0008580941, 0008591672, 0008173077, 0008169115, 0008098495, 0008164707, 0008498606, and 0008589924.
Other devices under the same FDA submission
3 devices cleared under K083422 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- TAIGA TA6JR40Medtronic, Inc. · CATHETER, PERCUTANEOUS1 report
- TAIGA TA6EBU35Medtronic, Inc. · CATHETER, PERCUTANEOUS1 report
- TAIGA TA6AL10Medtronic, Inc. · CATHETER, PERCUTANEOUS1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page