Why it was recalled

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

FDA root cause category: Error in labeling

What the company is doing

Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA

Recall details

Recall number
Z-1685-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 2, 2018
Posted by the FDA
May 4, 2018
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product code
MSW
FDA submissions
K982468
Distribution
Nationwide and Canada, Mexico, and Thailand

Product description

IMMULITE 2000/IMMULITE 2000 XPi TG

Code information (models, lots, serial numbers)

TG, TG (IMMULITE/IMMULITE 1000, LKTY1(D)), (IMMULITE 2000/IMMULITE 2000 XPi, LKTY2(D))  Lot codes: TG Lot # 346 347 348 349 350 351 352 353 354 355 356 D348 D349 D350 D351 D352 D353 D354 D356

Other devices under the same FDA submission

1 device cleared under K982468 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 79505).