Medical device recalls: May 2018
The FDA posted 161 medical device recall records in May 2018, from 25 recall events by 24 companies. 42 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Smith & Nephew, Inc. · 2 events, 3 records
- Covidien Medtronic · 1 event, 40 records
- Vyaire Medical · 1 event, 35 records, 35 Class I
- Siemens Healthcare Diagnostics, Inc. · 1 event, 24 records
- Boston Scientific Corporation · 1 event, 11 records
- Philips Medical Systems Nederlands · 1 event, 7 records
- Cook Inc. · 1 event, 6 records
- Baxter Healthcare Corporation · 1 event, 5 records
FDA root cause
- Process control · 50
- Component design/selection · 35
- Error in labeling · 24
- Device Design · 21
- Equipment maintenance · 9
- Nonconforming Material/Component · 8
- Software design · 5
- Employee error · 5
Class I recalls (42)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1903-2018 · Heartware · posted May 21, 2018 · Open, Classified
Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or…
- Z-1774-2018 · Abbott · posted May 17, 2018 · Terminated
Reports of outflow graft twist occlusions. Patients whose devices experience outflow graft occlusions will experience a persistent low flow alarm. Outflow graft occlusions can result in serious adverse events such as hemodynamic compromise, thrombus, and…
- Z-1701-2018 · Micro Therapeutics Inc, Dba Ev3 Neurovascular · posted May 16, 2018 · Terminated
There is a potential issue for all MindFrame Capture devices to partially detach or separate from the delivery wire.
- Z-1547-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1548-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1549-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1550-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1551-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1552-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1553-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- AirLife Resuscitation, Pediatric, with mask, 40inch oxygen reservoir tubing, pressure-relief valve ↗Class IZ-1554-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1555-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1556-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1557-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1558-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1559-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1560-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1561-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1562-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- AirLife Resuscitation, Pediatric, with mask, 28inch large bore tubing, pressure-relief valve ↗Class IZ-1563-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1564-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1565-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1566-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1567-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1568-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1569-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1570-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1571-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1572-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1573-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1574-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1575-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1576-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1577-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1578-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1579-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1580-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- AirLife Resuscitation, Infant, with mask, oxygen reservoir bag, pressure-relief valve, PEEP valve ↗Class IZ-1581-2018 · Vyaire Medical · posted May 10, 2018 · Terminated
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
- Z-1481-2018 · Draeger Medical, Inc. · posted May 8, 2018 · Terminated
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
- Z-1482-2018 · Draeger Medical, Inc. · posted May 8, 2018 · Terminated
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
- Z-1483-2018 · Draeger Medical, Inc. · posted May 8, 2018 · Terminated
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
- Z-1484-2018 · Draeger Medical, Inc. · posted May 8, 2018 · Terminated
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
Class II recalls (119)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1947-2018 · Respire Medical · posted May 29, 2018 · Terminated
Potential for device breakage during use
- Z-1948-2018 · Respire Medical · posted May 29, 2018 · Terminated
Potential for device breakage during use
- Z-1949-2018 · Respire Medical · posted May 29, 2018 · Terminated
Potential for device breakage during use
- Z-1950-2018 · Respire Medical · posted May 29, 2018 · Terminated
Potential for device breakage during use
- Z-1951-2018 · Respire Medical · posted May 29, 2018 · Terminated
Potential for device breakage during use
- Z-2107-2018 · MicroAire Surgical Instruments, LLC · posted May 25, 2018 · Terminated
The Instructions for Use (IFU) is incorrect with regards to sterilization parameters and must be replaced.
- Z-2267-2018 · Smith & Nephew, Inc. · posted May 21, 2018 · Terminated
The affected devices are manufactured to the same design specification as products intended for sale and use in the US but do not contain the FDA approved labeling (i.e., IFU and Patient Information Brochure). The affected devices were distributed in the…
- Z-2268-2018 · Smith & Nephew, Inc. · posted May 21, 2018 · Terminated
The affected devices are manufactured to the same design specification as products intended for sale and use in the US but do not contain the FDA approved labeling (i.e., IFU and Patient Information Brochure). The affected devices were distributed in the…
- Z-1901-2018 · LivaNova USA · posted May 18, 2018 · Terminated
Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.
- Z-1891-2018 · Volcano Corporation · posted May 17, 2018 · Terminated
The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.
- Z-1814-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1815-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1816-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1817-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1818-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1819-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1820-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1821-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1822-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1823-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1824-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1825-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1826-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1828-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1829-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1830-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1831-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- BOX KIT00713 KIT BY PASSClass IIZ-1832-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- BOX KIT00714 KIT SLEEVEClass IIZ-1833-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- BOX KIT1171C KIT LAP CYSTECTOMYClass IIZ-1834-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1837-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 1)KIT4184 APEN LOBECTOMY KIT X0Class IIZ-1838-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1839-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1840-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1842-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1843-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1844-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 1) BOX LAPZD ZESTAW DO SGClass IIZ-1847-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1848-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1849-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- MGBX ZESTAW GB LS X1Class IIZ-1850-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- BOX PST03533 KIT APP STAPLERClass IIZ-1852-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1854-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1855-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1858-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 1)BOX SLEEVEGK3Class IIZ-1859-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1862-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- BOX VATS1SCDX DO ZABIEGOW VATSClass IIZ-1863-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1865-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- BOX VWR ZESTAW DO VATS KLINClass IIZ-1866-2018 · Covidien Medtronic · posted May 16, 2018 · Terminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- Z-1883-2018 · Coltene Whaledent Inc · posted May 14, 2018 · Terminated
The products are labeled with an incorrect product number.
- Z-1802-2018 · Bard Peripheral Vascular Inc · posted May 11, 2018 · Terminated
Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present…
- Z-1803-2018 · Bard Peripheral Vascular Inc · posted May 11, 2018 · Terminated
Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present…
- Z-1751-2018 · Cook Inc. · posted May 9, 2018 · Terminated
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Z-1752-2018 · Cook Inc. · posted May 9, 2018 · Terminated
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Z-1753-2018 · Cook Inc. · posted May 9, 2018 · Terminated
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Z-1754-2018 · Cook Inc. · posted May 9, 2018 · Terminated
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Z-1755-2018 · Cook Inc. · posted May 9, 2018 · Terminated
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Z-1756-2018 · Cook Inc. · posted May 9, 2018 · Terminated
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Syngo.Via VB20A model 1049610 ↗Class IIZ-1778-2018 · Siemens Medical Solutions USA, Inc · posted May 9, 2018 · Terminated
Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still…
- Z-1921-2018 · Smith & Nephew, Inc. · posted May 9, 2018 · Terminated
Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.
- Z-2185-2018 · Roche Diabetes Care, Inc. · posted May 9, 2018 · Terminated
The recalling firm identified four US lots of test strips out of specification. Affected strips show increased potential for strip errors prior to applying blood, strips not detected once inserted, or biased results, which may not be detectable. In the…
- Z-2186-2018 · Roche Diabetes Care, Inc. · posted May 9, 2018 · Terminated
The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur,…
- Z-2187-2018 · Roche Diabetes Care, Inc. · posted May 9, 2018 · Terminated
The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur,…
- Z-2428-2018 · bioMerieux, Inc. · posted May 9, 2018 · Terminated
Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS and mini VIDAS system due to a…
- Z-2429-2018 · bioMerieux, Inc. · posted May 9, 2018 · Terminated
Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS and mini VIDAS system due to a…
- Z-1705-2018 · Zimmer Biomet, Inc. · posted May 7, 2018 · Terminated
The recalling firm has confirmed that the Outer Plate component exhibits an excessive chamfer on the threading after deburring operations. This excessive chamfer results in non-conforming product where the threads of the outer plate component have limited…
- Z-1616-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Z-1617-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Capio CLClass IIZ-1618-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Z-1619-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Capio Standard Box 4 (Capio PC)Class IIZ-1620-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Capio RPClass IIZ-1621-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Uphold LITE w/ Capio SLIMClass IIZ-1622-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Posterior LITE w/ Capio SLIM ↗Class IIZ-1623-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Z-1624-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
- PFR Kit- Pinnacle, Posterior ↗Class IIZ-1625-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Uphold Vaginal Support System ↗Class IIZ-1626-2018 · Boston Scientific Corporation · posted May 4, 2018 · Terminated
Potential for Capio sutures to break and /or detach.
- Z-1663-2018 · Baxter Healthcare Corporation · posted May 4, 2018 · Terminated
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr…
- Z-1664-2018 · Baxter Healthcare Corporation · posted May 4, 2018 · Terminated
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr…
- Z-1665-2018 · Baxter Healthcare Corporation · posted May 4, 2018 · Terminated
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr…
- Z-1666-2018 · Baxter Healthcare Corporation · posted May 4, 2018 · Terminated
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr…
- Z-1667-2018 · Baxter Healthcare Corporation · posted May 4, 2018 · Terminated
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr…
- Z-1668-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at…
- Z-1669-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at…
- Z-1670-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1671-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1672-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1673-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at…
- Z-1674-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1675-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1676-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1677-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- IMMULITE /IMMULITE 1000 CEA ↗Class IIZ-1678-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1679-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1680-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1681-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1682-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1683-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- IMMULITE /IMMULITE 1000 TG ↗Class IIZ-1684-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1685-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1686-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- ADVIA Centaur CyclosporineClass IIZ-1687-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at…
- ADVIA Centaur DHEA-SO4 ↗Class IIZ-1688-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at…
- ADVIA Centaur Folate (FOL)Class IIZ-1689-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- ADVIA Centaur HBc IgM ↗Class IIZ-1690-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- Z-1691-2018 · Siemens Healthcare Diagnostics, Inc. · posted May 4, 2018 · Terminated
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE…
- MultiDim Eleva 708032 ↗Class IIZ-1603-2018 · Philips Medical Systems Nederlands · posted May 3, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- 708033 URODiagnost ↗Class IIZ-1604-2018 · Philips Medical Systems Nederlands · posted May 3, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- MultiDiagnost Eleva 708036 ↗Class IIZ-1605-2018 · Philips Medical Systems Nederlands · posted May 3, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- Z-1606-2018 · Philips Medical Systems Nederlands · posted May 3, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- Z-1607-2018 · Philips Medical Systems Nederlands · posted May 3, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- Z-1608-2018 · Philips Medical Systems Nederlands · posted May 3, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- Z-1609-2018 · Philips Medical Systems Nederlands · posted May 3, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- Z-1610-2018 · Roche Diagnostics Corporation · posted May 3, 2018 · Terminated
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
- Z-1611-2018 · Roche Diagnostics Corporation · posted May 3, 2018 · Terminated
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
- Z-1612-2018 · Roche Diagnostics Corporation · posted May 3, 2018 · Terminated
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
- Z-1613-2018 · Roche Diagnostics Corporation · posted May 3, 2018 · Terminated
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
- Z-1389-2018 · Stryker Medical Division of Stryker Corporation · posted May 1, 2018 · Terminated
The power load floor plates which hold the ambulance cot anchors in place during patient transport, may be broken.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.