Why it was recalled

One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm initiated their recall by letter and email on 03/07/2018. The firm requested return of the product.

Recall details

Recall number
Z-1317-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 7, 2018
Posted by the FDA
Mar 20, 2018
Recalling firm
Smith & Nephew, Inc.
Quantity
59 units
Product code
HWC
FDA submissions
K132296
Distribution
US, Puerto Rico, Malaysia

Product description

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

Code information (models, lots, serial numbers)

Lot Number 17GM04522

Other devices under the same FDA submission

7 devices cleared under K132296 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.