ACUSON SC2000 Ultrasound System. The firm name on the… recallZ-1200-2018 · Siemens Medical Solutions USA, Inc. · initiated Mar 5, 2018

Class II Terminated All recalls posted in March 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment.

FDA root cause category: Device Design

What the company is doing

Undated recall letters were issued via email (if they had an email address for the customer) or via certified mail (if no email address was available) on 3/5/2018.

Recall details

Recall number
Z-1200-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 5, 2018
Posted by the FDA
Mar 26, 2018
Recalling firm
Siemens Medical Solutions USA, Inc.
Quantity
229 systems
Product code
IYN
FDA submissions
K170315
Distribution
Distribution was made to medical facilities in AZ, CA, CT, FL, IL, IN, KS, KY, MI, MN, MO, NJ, NY, NC, OH, PA, TX, VA, and WA. There was no military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Bosnia Herzeg, Colombia, Croatia, Denmark, France, Germany, India, Ireland, Israel, Italy, Japan, Mexico, New Zealand, Philippines, Poland, Republic Korea, Romania, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, and Vietnam.

Product description

ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo Grove, IL.

Code information (models, lots, serial numbers)

Software versions VB20A, VB20B, or VB20C that have a common physio module (CPM) (known by customers as the ECG (physio) hardware) installed at revision 4

Other devices under the same FDA submission

1 device cleared under K170315 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.