Skip to content

ATTUNE recallZ-1704-2018 · DePuy Orthopaedics, Inc. · initiated Apr 16, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There may be burrs on the extraction hole threads.

FDA root cause category: Process control

What the company is doing

Notification letters distributed on 4/17/18 instructed customers to perform the following: Steps to Take Please take the following actions: " Immediately return unused affected units: Note: On March 19, 2018, one unit was removed from consignment. Two units were implanted prior to the medical device recall. No additional units are in the market. " Reconciliation Form: Complete the reconciliation form and return to your sales consultant or fax to 574-371-4939 or email to [email protected] within five (5) days of this notice. " Records: Retain a copy of the completed reconciliation form in your files along with this notice. " Additional Notifications: o Notify surgeons at your facility by providing them with a copy of this notice. Forward this notice to others in your facility that need to be informed. Clinical Implications and Patient Impact DePuy (Ireland) is not recommending prophylactic revision in the absence of symptoms. The company recommends that surgeons discuss potential clinical implications and risks with symptomatic patients that received the affected units.

Recall details

Recall number
Z-1704-2018
Status
Terminated
Initiated by the company
Apr 16, 2018
Posted by the FDA
Apr 30, 2018
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
1 unit
Product code
JWH
FDA submissions
K160700
Distribution
The products were distributed to the following US states: FL, MO, and PA.

Product description

ATTUNE Revision CRS Femoral RT SZ 5 Cemented, Catalog No. 150440205, Femoral Revision Implant, Knee

Code information (models, lots, serial numbers)

Lot Number HD6516, GTIN Number 10603295042280

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

48 devices cleared under K160700 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 79818).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page