Medical device recalls: April 2018
The FDA posted 43 medical device recall records in April 2018, from 20 recall events by 18 companies.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips Electronics North America Corporation · 2 events, 14 records
- Teleflex Medical Europe Ltd · 2 events, 5 records
- C.R. Bard, Inc. · 1 event, 3 records
- Datascope Corporation · 1 event, 3 records
- Terumo Cardiovascular Systems Corporation · 1 event, 3 records
- Cook Inc. · 1 event, 2 records
- DePuy Orthopaedics, Inc. · 1 event, 2 records
- Bard Brachytherapy, Inc. · 1 event, 1 record
FDA root cause
- Nonconforming Material/Component · 14
- Process control · 6
- Error in labeling · 4
- Employee error · 4
- Device Design · 4
- Labeling design · 3
- Other · 2
- Under Investigation by firm · 1
Class II recalls (42)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- ATTUNE Revision CRS Femoral LT SZ 5 Cemented, Catalog No. 150440105, Femoral Revision Implant, KneeClass IIZ-1703-2018 · DePuy Orthopaedics, Inc. · posted Apr 30, 2018 · Terminated
There may be burrs on the extraction hole threads.
- ATTUNE Revision CRS Femoral RT SZ 5 Cemented, Catalog No. 150440205, Femoral Revision Implant, KneeClass IIZ-1704-2018 · DePuy Orthopaedics, Inc. · posted Apr 30, 2018 · Terminated
There may be burrs on the extraction hole threads.
- Z-1702-2018 · Bard Brachytherapy, Inc. · posted Apr 27, 2018 · Terminated
Bard Medical Division is recalling the BARD BrachySource Iodine125 Radioactive Seeds in Mick Applicators and BARD BrachySource Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. Because the incorrect dosage level radioactive seeds were mixed…
- Z-1894-2018 · Beckman Coulter Inc. · posted Apr 27, 2018 · Terminated
The Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.
- Z-1523-2018 · Sofradim Production · posted Apr 26, 2018 · Terminated
There have been patient reports of abdominal hernia recurrence following hernia repair. The recalling firm has added a statement to the device's Instructions for Use to make surgeons aware of risks in using the device in Transversus Abdominis Muscle…
- Z-1513-2018 · BioFire Diagnostics, LLC · posted Apr 25, 2018 · Terminated
There is an increased risk of false positive Proteus results when the product is used with specific types of BD BACTEC blood culture bottles.
- Z-1361-2018 · Crimson Trace Corporation · posted Apr 23, 2018 · Terminated
A previously "in specification" Finished Product with a Model 25-0657 diode had power output higher than 5 mW.
- Z-1463-2018 · Integra LifeSciences Corp. · posted Apr 20, 2018 · Terminated
Incidents of stem trial breakage were reported to the firm suggesting that these fractures all occurred during insertion/impaction or extraction of the humeral stem trial whiling preparing the humeral canal and/or trialing. All cases resulted in a delay in…
- Medtronic DxTerity Diagnostic Catheter, JL 4.0, REF DXT5JL40, marked with 6F color code on box.Class IIZ-1479-2018 · Medtronic Vascular · posted Apr 20, 2018 · Terminated
The color coded French size indicator on the outer box may incorrectly reflect a 6 French size, rather than the correct 5 French size. The catheters inside the carton are the correct DXT5JL40 item, the inner pouch has correct 5F labeling, information and…
- Z-1444-2018 · TMJ Solutions Inc · posted Apr 19, 2018 · Terminated
This device was mislabeled and is not the correct component for this patient.
- Z-1416-2018 · Terumo Cardiovascular Systems Corporation · posted Apr 18, 2018 · Terminated
Update to cooler-heater cleaning instructions.
- Sarns TCM, Catalog No. 15747 ↗Class IIZ-1417-2018 · Terumo Cardiovascular Systems Corporation · posted Apr 18, 2018 · Terminated
Update to cooler-heater cleaning instructions.
- Z-1418-2018 · Terumo Cardiovascular Systems Corporation · posted Apr 18, 2018 · Terminated
Update to cooler-heater cleaning instructions.
- Z-1430-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Z-1431-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Intera 1.0T ↗Class IIZ-1432-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Intera 1.5T ↗Class IIZ-1433-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Panorama 1.0T, Model 781250 ↗Class IIZ-1434-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Z-1435-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Z-1436-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Z-1437-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- T5-NT, Model 78104 ↗Class IIZ-1438-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- ACS-NT, Model 78108 ↗Class IIZ-1439-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- T10-NT, Model 78107 ↗Class IIZ-1440-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Multiva systems ↗Class IIZ-1441-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- T5 Model 78104 ↗Class IIZ-1442-2018 · Philips Electronics North America Corporation · posted Apr 18, 2018 · Terminated
Potential risk for helium gas inside the MR examination room during a magnet quench
- Z-1403-2018 · Teleflex Medical Europe Ltd · posted Apr 17, 2018 · Terminated
These one-way valves may disconnect at the joint between the two components that make up the device, which would cause an interruption of ventilation to the patient.
- Z-1383-2018 · Datascope Corporation · posted Apr 16, 2018 · Terminated
The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
- Z-1384-2018 · Datascope Corporation · posted Apr 16, 2018 · Terminated
The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
- Z-1385-2018 · Datascope Corporation · posted Apr 16, 2018 · Terminated
The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
- Z-1377-2018 · Cook Inc. · posted Apr 13, 2018 · Terminated
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
- Z-1378-2018 · Cook Inc. · posted Apr 13, 2018 · Terminated
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
- Latex Rebreathing Bags: ↗Class IIZ-1367-2018 · Teleflex Medical Europe Ltd · posted Apr 12, 2018 · Terminated
These products contain latex, but the following statement is not included on any level of product packaging: Caution: This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
- Rectal Catheters ↗Class IIZ-1368-2018 · Teleflex Medical Europe Ltd · posted Apr 12, 2018 · Terminated
These products contain latex, but the following statement is not included on any level of product packaging: Caution: This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
- Z-1369-2018 · Teleflex Medical Europe Ltd · posted Apr 12, 2018 · Terminated
These products contain latex, but the following statement is not included on any level of product packaging: Caution: This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
- Baker Jejunostomy Tubes: ↗Class IIZ-1370-2018 · Teleflex Medical Europe Ltd · posted Apr 12, 2018 · Terminated
These products contain latex, but the following statement is not included on any level of product packaging: Caution: This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
- VOLISTA StandOP Surgical LightClass IIZ-1349-2018 · Getinge US Sales LLC · posted Apr 10, 2018 · Terminated
The manufacturer received complaints indicating that the device's central handle holder detaches easily from the handle holder ring on the light head during clinical use. In the event of such detachment, particles from the handle holder ring could fall…
- IntelliVue X3 Patient Monitor. ↗Class IIZ-1315-2018 · Philips Electronics North America Corporation · posted Apr 6, 2018 · Terminated
The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting.…
- Z-1307-2018 · C.R. Bard, Inc. · posted Apr 4, 2018 · Terminated
Several instances of a catheter caught in the sterile barrier seal and in some instances cut catheters were observed.
- Z-1308-2018 · C.R. Bard, Inc. · posted Apr 4, 2018 · Terminated
Several instances of a catheter caught in the sterile barrier seal and in some instances cut catheters were observed.
- Z-1309-2018 · C.R. Bard, Inc. · posted Apr 4, 2018 · Terminated
Several instances of a catheter caught in the sterile barrier seal and in some instances cut catheters were observed.
- Z-1289-2018 · Invivo Corporation · posted Apr 2, 2018 · Terminated
Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered correctly on the DynaCad Client. The defect causes the pharmacokinetic (PK) color maps to display incorrectly when viewed from remote DynaCAD client…
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1322-2018 · Clinical Diagnostic Systems · posted Apr 10, 2018 · Terminated
VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.