Skip to content

RX Imola (RX4900) recallZ-2059-2018 · Randox Laboratories · initiated May 4, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Randox Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abnormal premature termination of the software could affect the use of the analyser when running patient samples. The impact being a delay in reporting test results.

FDA root cause category: Under Investigation by firm

What the company is doing

All customers will be contacted directly by email and will be followed up with once every 2 weeks over a period of 6 weeks. Customers are advised to update the software in all analyzers. For further questions, please call 1 (866)472-6369.

Recall details

Recall number
Z-2059-2018
Status
Terminated
Initiated by the company
May 4, 2018
Recalling firm
Randox Laboratories
Quantity
41 analyzers
Product code
JJE
FDA submissions
K052914
Distribution
US Distribution including Puerto Rico and the state of : West Virginia.

Product description

RX Imola (RX4900) For Professional Use for the quantitative in vitro determination of various clinical chemistry tests.

Code information (models, lots, serial numbers)

GTIN 05055273206104. All serial numbers.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page