RX imola RX4900Recalls & reported problems

Randox Laboratories Limited · Rx Imola is an automated clinical chemistry analyser complete with dedicated analyser software. · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Sep 6, 2023Open, Classified

    RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

    Z-0200-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Apr 28, 2021Terminated

    An issue was identified where the software froze during processing of commands, which resulted in no results displayed.

    Z-2013-2021 · Root cause: Software design · Recall details · FDA record ↗

  3. May 4, 2018Terminated

    Abnormal premature termination of the software could affect the use of the analyser when running patient samples. The impact being a delay in reporting test results.

    Z-2059-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗