BARD BrachySource Iodine125 radioactive seeds consist of a… recallZ-1702-2018 · Bard Brachytherapy, Inc. · initiated Apr 10, 2018
Class II Terminated All recalls posted in April 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bard Brachytherapy, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Bard Medical Division is recalling the BARD BrachySource Iodine125 Radioactive Seeds in Mick Applicators and BARD BrachySource Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. Because the incorrect dosage level radioactive seeds were mixed up and sent to two hospitals. One Hospital ordered BrachySource Iodine125 0.36 mCi radioactive seeds but instead received BrachySource Iodine125 0.28 mCi radioactive seeds and vice versa for the other hospital affected.
FDA root cause category: Labeling mix-ups
What the company is doing
The notification to the customer was completed by Bard Medical Customer Service via a telephone call to alert them of the incorrect product shipped on April 10, 2018. Bard Medical Division is recalling the BARD BrachySource Iodine125 Radioactive Seeds in Mick Applicators and BARD BrachySource Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. The incorrect dosage level of radioactive seeds was sent to 2 hospitals. One hospital ordered 0.36 mCi radioactive seeds but instead received 0.28 mCi radioactive seeds intended for the other hospital. For further questions, please call (770) 385-0562.
Recall details
- Recall number
- Z-1702-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 10, 2018
- Posted by the FDA
- Apr 27, 2018
- Recalling firm
- Bard Brachytherapy, Inc.
- Quantity
- 70 Seeds at 0.28mCi and 100 seeds at 0.36mCi
- Product code
- KXK
- FDA submissions
- K093663
- Distribution
- US Distribution to the states of : California and Georgia
Product description
BARD BrachySource Iodine125 radioactive seeds consist of a welded titanium capsule containing the I125 absorbed onto a nickel / copper-coated, gold-cored aluminum wire (reference Figure 1 below). The I125 has a half-life of 59.6 days and decays by electron capture with the emission of characteristic photons and Auger electrons. The principal photon emissions are 27.4 and 31 keV x-rays and 35.5 keV gamma. The titanium wall of the BrachySource Seed Implants absorbs the electrons.
Code information (models, lots, serial numbers)
Product Code:1251CSR, BrachySource¿ Seed Implants in Re-Usable Mick¿ Applicators (60 seeds at 0.28mCi) with Lot BBCP0079 Product Code: 1251CSRA, BrachySource¿ Seed Implants in Shielded Mick¿ Applicators (10 seeds at 0.28mCi) with Lot BBCQ0026 Product Code: PS1251LS, Loose BrachySource¿ Seed Implants (with Lot BBCQ0050) Spacers (with Lot 23498276) 100 Seeds at 0.36 mCi.
Other devices under the same FDA submission
8 devices cleared under K093663 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- QuickLink™, BrachySource™ 1251QCSBard Brachytherapy, Inc. · Source, brachytherapy, radionuclide5 reports
- BrachySource™ PS1251LSBard Brachytherapy, Inc. · Source, brachytherapy, radionuclide4 reports
- QuickLink™, BrachySource™ 1251QCLBard Brachytherapy, Inc. · Source, brachytherapy, radionuclide4 reports
- BrachySource™ PS1251CSBard Brachytherapy, Inc. · Source, brachytherapy, radionuclide3 reports
- BrachySource™ 1251CSRBard Brachytherapy, Inc. · Source, brachytherapy, radionuclide2 reports
- BrachySource™ PS1251CBard Brachytherapy, Inc. · Source, brachytherapy, radionuclide1 report
- BrachySource™ 1251CSRABard Brachytherapy, Inc. · Source, brachytherapy, radionuclide1 report
- BrachySource™, SourceLink™ 1251LINKBard Brachytherapy, Inc. · Source, brachytherapy, radionuclide1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page