BioMerieux mini VIDAS System in combination with Quality… recallZ-2428-2018 · bioMerieux, Inc. · initiated Jan 4, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from bioMerieux, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS and mini VIDAS system due to a decrease of pipetted volume.
FDA root cause category: Device Design
What the company is doing
Subsidiaries and distributors were initially notified of the recall on approximately 01/04/2018. U.S. customers were notified via letter on approximately 01/22/2018. On 03/28/2018 a follow-up notification was issued world-wide to customers to expand the scope. Instructions included to increase the frequency of the Quality Control VIDAS (QCV) testing to weekly rather than monthly to reduce the period of potential retrospective analyses needed, continue to conduct a visual inspection of the SPR after each run (including patient sample test and QCV test) to ensure that the dot on the SPR is still in place, continue to perform External QC testing in accordance with laboratory policy and in compliance with Governmental Regulations & Accreditation Requirements, which includes CLIA, ensure the notification letter is distributed to and reviewed by all appropriate personnel, store the notification letter with the bioMrieux VIDAS and /or miniVIDAS documentation, and complete and return the Acknowledgement Form in Attachment A by Fax to confirm receipt of the notice. For further questions, please call (314) 731-8694
Recall details
- Recall number
- Z-2428-2018
- Status
- Terminated
- Initiated by the company
- Jan 4, 2018
- Posted by the FDA
- May 9, 2018
- Recalling firm
- bioMerieux, Inc.
- Quantity
- 1,866 units
- Product code
- JJE
- FDA submissions
- K923579
- Distribution
- Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY. International distribution to Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bolivia, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Cape Verde, Chad, Chile, China / Hong Kong, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Djibouti, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guam, Guatemala, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mali, Malta, Mauretania, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Tanzania, Thailand, Togo, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Zimbabwe.
Product description
BioMerieux mini VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV) is an automated test for use on the VIDAS system to detect abnormal operation of the VIDAS and mini VIDAS instrument pipette mechanisms and optical systems. Specifically the mini VIDAS system is an immunodiagnostic system intended to be used by trained and qualified laboratory professionals, for In vitro Diagnostic (IVD) purpose, veterinary and industrial applications. The mini VIDAS system is intended to execute an immunoassay protocol and to release results according to the package insert of the VIDAS assay kits.
Code information (models, lots, serial numbers)
Description, Reference No. (UDI No.): MINI VIDAS ANALYSER EU, 99174 (03573026111021); MINI VIDAS ANALYSER US, 99175 (03573026111045); MINI VIDAS 100-120 VOLT VDE/CE, 99733 (03573026118662); MINI VIDAS 200-240 VOLT VDE/CE, 99734 (03573026118655); MINI VIDAS BLUE 110 / 220 V, 99737 (03573026140441); MINI VIDAS ARKRAY ANALYZER, 99739 (03573026140465); MINI VIDAS, 410416 (03573026351656); MINI VIDAS, W1421 (03573026029180); MINIVIDAS INDUSTRY US, 421001 (03573026553319)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- mini VIDAS® 99737Biomerieux SA · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use1 recall73 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 79993).
- Z-2429-2018 · TerminatedBioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV) is an automated test for use on the VIDAS system to detect…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page