Ingenuity CT recallZ-2274-2018 · Philips Medical Systems (Cleveland) Inc · initiated Mar 28, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems (Cleveland) Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.

FDA root cause category: Other

What the company is doing

Urgent Field Safety Notice letters were sent to customers on 3/28/18. The letters did not request any actions to be conducted by the customers. For Brilliance Big Bore/ Ingenuity CT / Ingenuity Core / Ingenuity Corel 28 Customers: A Philips Field Service Engineer will contact you to schedule the FCO updates at your site. Field Change Order (FCO 72800691 and 72800692) will be released to correct the issue. For Ingenuity TF PET/CT Customers: Field Change Order (FCO 88200520 and 88200511) will be released to correct this issue. A Philips Field Service Engineer will contact you to schedule the FCO updates at your site.

Recall details

Recall number
Z-2274-2018
Status
Terminated
Initiated by the company
Mar 28, 2018
Recalling firm
Philips Medical Systems (Cleveland) Inc
Quantity
28 total
Product code
JAK
FDA submissions
K160743
Distribution
The systems were distributed to the following US states: HI, KS, and MO. The systems were distributed to the following foreign countries: Algeria, Austria, Brazil, China, Indonesia, Japan, Kenya, Korea, Mexico, Poland, Turkey, and Uganda.

Product description

Philips Ingenuity Core 728321

Code information (models, lots, serial numbers)

UDI: (01)00884838059498(21)333169, Serial Number: 333169;  UDI: (01)00884838059498(21)310377, Serial Number: 310377;  UDI: (01)00884838059498(21)310375, Serial Number: 310375;  UDI: (01)00884838059498(21)333171, Serial Number: 333171;  UDI: (01)00884838059498(21)333172, Serial Number: 333172

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 80176).