Skip to content

Ingenuity CTRecalls & reported problems across all models

Philips Medical Systems Nederland B.V. · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name at least one Ingenuity CT model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
14
Deaths reported
1
Injuries reported
5
Recalls
6

We found 3 models of Ingenuity CT by Philips Medical Systems Nederland B.V. in the FDA's device registry (GUDID). Together they are named in 14 adverse event reports in the FDA's MAUDE database, received between Dec 21, 2017 and Dec 11, 2025 — classified as 1 death, 5 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Crack” (2 reports).

6 FDA recall records name at least one of these models by its device identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Ingenuity CT recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 11, 2026Open, Classified

    The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

    Z-2652-2026 · Root cause: Radiation Control for Health and Safety Act · names 3 of these models · Recall details · FDA record ↗

  2. Aug 13, 2025Open, Classified

    The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

    Z-2600-2025 · Root cause: Process control · names 3 of these models · Recall details · FDA record ↗

  3. Jun 8, 2021Terminated

    When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

    Z-2315-2021 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  4. May 1, 2021Open, Classified

    Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

    Z-1984-2021 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  5. Mar 28, 2018Terminated

    A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.

    Z-2275-2018 · Root cause: Other · Recall details · FDA record ↗

  6. Mar 28, 2018Terminated

    A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.

    Z-2274-2018 · Root cause: Other · Recall details · FDA record ↗

Most reported Ingenuity CT problems

Device problems

  • Crack2
  • Detachment of Device or Device Component2
  • Unintended System Motion2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Device Handling Problem1
  • Environmental Particulates1
  • Failure to Calibrate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Laceration(s)4
  • Hepatitis2
  • Insufficient Information2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Fasciitis1
  • No Consequences Or Impact To Patient1
  • No Information1

Ingenuity CT models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Ingenuity CT adverse event reports

The 14 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Philips Medical Systems Nederland B.V. products

All Philips Medical Systems Nederland B.V. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page