Ingenuity CTRecalls & reported problems across all models
Philips Medical Systems Nederland B.V. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one Ingenuity CT model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 14
- Deaths reported
- 1
- Injuries reported
- 5
- Recalls
- 6
We found 3 models of Ingenuity CT by Philips Medical Systems Nederland B.V. in the FDA's device registry (GUDID). Together they are named in 14 adverse event reports in the FDA's MAUDE database, received between Dec 21, 2017 and Dec 11, 2025 — classified as 1 death, 5 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Crack” (2 reports).
6 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Ingenuity CT recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jun 11, 2026Open, Classified
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
Z-2652-2026 · Root cause: Radiation Control for Health and Safety Act · names 3 of these models · Recall details · FDA record ↗
- Aug 13, 2025Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Z-2600-2025 · Root cause: Process control · names 3 of these models · Recall details · FDA record ↗
- Jun 8, 2021Terminated
When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
Z-2315-2021 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗
- May 1, 2021Open, Classified
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Z-1984-2021 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗
- Mar 28, 2018Terminated
A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Z-2275-2018 · Root cause: Other · Recall details · FDA record ↗
- Mar 28, 2018Terminated
A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Z-2274-2018 · Root cause: Other · Recall details · FDA record ↗
Most reported Ingenuity CT problems
Device problems
- Crack2
- Detachment of Device or Device Component2
- Unintended System Motion2
- Adverse Event Without Identified Device or Use Problem1
- Device Dislodged or Dislocated1
- Device Handling Problem1
- Environmental Particulates1
- Failure to Calibrate1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Laceration(s)4
- Hepatitis2
- Insufficient Information2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Fasciitis1
- No Consequences Or Impact To Patient1
- No Information1
Ingenuity CT models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Ingenuity CTSystem, x-ray, tomography, computed5 recalls1 death9 reports
- Ingenuity CTSystem, x-ray, tomography, computed5 recalls2 reports
- Ingenuity CTSystem, x-ray, tomography, computed3 recalls3 reports
Recent Ingenuity CT adverse event reports
The 14 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryDec 11, 2025Crack; Crack
THE CUSTOMER REPORTED THAT A TECHNICIAN SUSTAINED A MINOR CUT WHILE WIPING DOWN THE MYLAR RING COVER ON AN INGENUITY CT SYSTEM. ADDITIONAL INFORMATION CONFIRMED THAT THE TECHNICIAN WAS CLEANING BLOOD FROM THE SCANNER WHEN THEY SUSTAINED A SMALL PERCUTANEOUS PUNCTURE FROM THE BROKEN MYLAR. IT WAS CONFIRMED THAT THE EXPOSURE INVOLVED BLOOD POSITIVE FOR HEPATITIS C. AN ATTEMPT WAS MADE TO OBTAIN THE TECHNICIAN¿S POST-EXPOSURE RESULTS; HOWEVER, THE INFORMATION PROVIDED WAS THAT THE RESULTS WERE "POSITIVE," WITHOUT CLARIFICATION AS TO WHETHER THIS REFERRED TO THE SOURCE BLOOD ON THE MYLAR RING OR THE TECHNICIAN¿S OWN TEST RESULTS. DUE TO THE LACK OF CLARITY REGARDING THE TEST OUTCOME, THIS EVENT IS BEING REPORTED AS A SERIOUS INJURY.
Patient problems: Hepatitis; Laceration(s)
MAUDE report 23778729 ↗InjuryDec 9, 2025Crack; Crack
THE CUSTOMER REPORTED THAT A TECHNICIAN SUSTAINED A MINOR CUT WHILE WIPING DOWN THE MYLAR RING COVER ON AN INGENUITY CT SYSTEM. ADDITIONAL INFORMATION INDICATED THAT THE TECHNICIAN WAS CLEANING BLOOD FROM THE SCANNER WHEN THEY SUSTAINED A SMALL PERCUTANEOUS PUNCTURE FROM THE BROKEN MYLAR. IT WAS CONFIRMED THAT THE EXPOSURE INVOLVED BLOOD POSITIVE FOR HEPATITIS C. AN ATTEMPT WAS MADE TO OBTAIN THE TECHNICIAN¿S POST-EXPOSURE RESULTS; HOWEVER, THE INFORMATION PROVIDED WAS THAT THE RESULTS WERE ¿POSITIVE," WITHOUT CLARIFICATION AS TO WHETHER THIS REFERRED TO THE SOURCE BLOOD ON THE MYLAR RING OR THE TECHNICIAN¿S OWN TEST RESULTS. DUE TO THE LACK OF CLARITY REGARDING THE TEST OUTCOME, THIS EVENT IS BEING REPORTED AS A SERIOUS INJURY.
Patient problems: Hepatitis; Laceration(s)
MAUDE report 23760867 ↗InjuryNov 13, 2025Environmental Particulates; Insufficient Device Problem Information; Environmental Particulates; Insufficient Device Problem Information
PHILIPS RECEIVED A REPORT INDICATING THAT DURING AN INTERVENTIONAL CEMENTOPLASTY PROCEDURE, THE CT SYSTEM INTERMITTENTLY GENERATED AN ERROR MESSAGE, INTERRUPTING THE SCAN. DURING THE PROCEDURE, THE ERROR CAUSED THE SYSTEM TO CLOSE AND RESTART THE EXAMINATION, RESULTING IN A DELAY IN IMAGE ACQUISITION. DUE TO THE INTERRUPTION, A PORTION OF THE INJECTED CEMENT COULD NOT BE ADEQUATELY VISUALIZED AND SUBSEQUENTLY MIGRATED INTO THE PATIENT¿S INFERIOR VENA CAVA (IVC).
Patient problems: Insufficient Information; Unspecified Vascular Problem
MAUDE report 23549881 ↗MalfunctionAug 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE INTERNAL RING COVER CAME LOOSE FROM A DAMAGED ROTATING INTERNAL COMPONENT AND CAUSED A PATIENT TO EXPERIENCE A SUPERFICIAL CUT TO THEIR LEFT ELBOW WITH SLIGHT BLEEDING. THE CUT WAS TREATED WITH A BANDAGE. PATIENT OUTCOME IS LISTED AS A BRUISE EXPECTED TO FULLY HEAL, NO PAIN AND FULL FUNCTIONALITY OF THE ELBOW/ARM. THEREFORE, THE REPORTED HARM DOES NOT MEET CRITERIA FOR SERIOUS INJURY. HOWEVER, IT IS UNCERTAIN WHETHER THIS EVENT IS LIKELY TO RECUR AND CAUSE A SERIOUS INJURY. THEREFORE, BASED ON THE AVAILABLE INFORMATION THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20044338 ↗InjuryAug 1, 2023Lack of Maintenance Documentation or Guidelines; Adverse Event Without Identified Device or Use Problem; Lack of Maintenance Documentation or Guidelines; Adverse Event Without Identified Device or Use Problem
REPORTING DETERMINATION: THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS AN ALLEGATION OF SERIOUS INJURY. THIS ISSUE REPORTED WAS A FIELD SERVICE ENGINEER INJURED HIS BACK DURING REPLACEMENT OF THE X-RAY TUBE. FSE WAS DIAGNOSED WITH LUMBAR FASCITIS AND FLURBIPROFEN WAS PRESCRIBED AND REST FOR A WEEK. FSE WAS RECOVERED AND HAS BEEN RETURNED TO WORK. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN INITIALLY DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: Fasciitis
MAUDE report 17439160 ↗MalfunctionMay 31, 2023Unintended System Motion; Unintended System Motion
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THE CUSTOMER REPORTED THAT WHILE RAISING THE COUCH WITH PATIENT ON TABLE, THE COUCH SUDDENLY DROPPED TO THE FLOOR. PHILIPS FIELD SERVICE ENGINEER(FSE) INSPECTED SYSTEM AND IDENTIFIED THAT COUCH BALL SCREW WAS BROKEN. FSE REPLACED BOTH BALL BEARING AND BALL SCREW, SYSTEM WAS RETURNED TO THE CUSTOMER FOR CLINICAL USE AFTER PARTS REPLACEMENT. THERE WAS NO REPORT OF HARM. WE CONSIDERED THIS EVENT REPORTABLE OUT OF ABUNDANCE OF CAUTION. PHILIPS HAS NOW COMPLETED THE INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17033634 ↗MalfunctionMay 16, 2023Unintended System Motion; Unintended System Motion
REPORTING DETERMINATION: THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THE CUSTOMER REPORTED SYSTEM MADE A VERY LOUD NOISE AND THEN TABLE DROPPED FROM RAISED POSITION FOR APPROXIMATELY 4 INCHES WHEN PATIENT GOT ON THE TABLE. THERE WAS NO REPORT OF HARM. THE PATIENT WENT TO A DIFFERENT FACILITY (HOSPITAL SITE) TO COMPLETE STUDY. PHILIPS FIELD SERVICE ENGINEER EVALUATED THE SYSTEM AND CONFIRMED THAT THE BALL SCREW AND BEARINGS HAD FAILED. THE COUCH WAS REPLACED AT THE CUSTOMER SITE AND DEVICE WAS RETURNED TO CLINICAL USE AFTER THE SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16937719 ↗MalfunctionSep 1, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. THE ISSUE REPORTED WAS 5 OF 5 FASTENERS SECURING THE CT COUCH TO THE FLOOR WERE DISLODGED AND THAT THERE IS POTENTIAL FOR THE COUCH TO TIP. THE SYSTEM WAS NOT IN CLINICAL USE AT THE TIME. BASED ON INFORMATION KNOWN AT THIS TIME, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. THE ISSUE IS CURRENTLY UNDER INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15334262 ↗MalfunctionJun 16, 2022Packaging Problem; Packaging Problem
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. THE ISSUE REPORTED WAS THAT IMPROPER PACKING FOR APM/CPM THAT WAS SENT TO THE SERVICE ENGINEER FOR REPLACEMENT. ENGINEERING INITIAL REVIEW OF THE EVENT HAS DETERMINED THE RISK IS UNACCEPTABLE. THERE IS POTENTIAL OF HARM. THIS EVENT IS CURRENTLY UNDER INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION HAS BEEN COMPLETED. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN INITIALLY DETERMINED TO BE A REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14713030 ↗DeathOct 14, 2021No Device Output; Failure to Calibrate; No Device Output; Failure to Calibrate
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THE ISSUE REPORTED WAS THAT DURING A HEAD/ NECK CT ANGIO EXAM, THE SYSTEM HAD AN ERROR AND WOULD NOT EXPOSE. THE CUSTOMER STATED THEY STOPPED THE PATIENT SCAN. THE CUSTOMER HAS ALLEGED THAT THIS RESULTED IN A DELAYED RESCUE RESULTING IN THE PATIENT'S DEATH APPROXIMATELY 20 DAYS LATER. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN INITIALLY DETERMINED TO BE A REPORTABLE EVENT AND IS CURRENTLY UNDER INVESTIGATION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information
MAUDE report 12632486 ↗InjuryAug 23, 2021Device Handling Problem; Device Handling Problem
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THE ISSUE REPORTED WAS THAT WHILE POSITIONING AN UNCONSCIOUS PATIENT, THE PATIENT'S ARM WAS NOT SECURED AND BUMPED AGAINST THE GANTRY MYLAR RING. THIS LED TO THE MYLAR RING BEING DAMAGED CAUSING A LACERATION TO PATIENT'S ARM INJURY WHICH REQUIRED STITCHES. THERE WAS NO MALFUNCTION OF THE SYSTEM; IT IS WORKING AS SPECIFIED. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.
Patient problems: Laceration(s)
MAUDE report 12357304 ↗MalfunctionApr 15, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. THE ISSUE REPORTED WAS THAT GANTRY MYLAR RING BROKE AND MADE CONTACT WITH THE PATIENT IN THE FACE AND ELBOW. BASED ON THE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.
Patient problems: Laceration(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 11676277 ↗MalfunctionAug 12, 2020Patient Data Problem; Patient Data Problem
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THE ISSUE REPORTED WAS THAT DURING A CT GUIDED PUNCTURE, NEWLY CREATED CT IMAGES WERE NOT UPDATED AFTER THE MESSAGE "VIEWER MEMORY FULL" WAS DISPLAYED ON THE CONSOLE. THIS RESULTED IN THE PATIENT'S PORTAL VEIN WAS INADVERTENTLY PUNCTURED. THE PATIENT WAS TRANSFERRED TO ANOTHER CT SYSTEM TO EVALUATE POTENTIAL INJURY(INTERNAL BLEEDING) AS A RESULT OF THE PORTAL VEIN PUNCTURE. PATIENT OUTCOME IS UNKNOWN AT THIS TIME. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient; No Information
MAUDE report 10399367 ↗MalfunctionDec 21, 2017Noise, Audible
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. THE CUSTOMER REPORTED THAT THERE WAS NOISE FROM THE CT BOX SPEAKER DUE TO GANTRY ELECTRONICS/FAN SOUNDS WITHOUT ANY INPUT FROM THE MICROPHONE; THEREFORE, THE SPEAKER VOLUME WAS TURNED DOWN. IF THE OPERATOR IS UNABLE TO HEAR THE PATIENT DUE TO THE VOLUME BEING TURNED DOWN TO A MINIMUM, THERE IS POTENTIAL FOR INJURY TO THE PATIENT. THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. THIS EVENT IS CURRENTLY UNDER INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7139620 ↗
Other Philips Medical Systems Nederland B.V. products
- Allura Xper FD19 models · 15,187 reports · 84 recalls
- Azurion22 models · 9,199 reports · 71 recalls
- Ingenia54 models · 479 reports · 138 recalls
- IntelliSpace Cardiovascular11 models · 439 reports · 3 recalls
- MR 7000 Series2 models · 4 reports · 11 recalls
- Dynamic Coronary Roadmap3 models · 10 reports · 9 recalls
- MR 5000 Series3 models · 11 reports · 8 recalls
- Brilliance iCT2 models · 9 reports · 8 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page