Ball Tip Guide Wire 2.0mmx60cm recallZ-2235-2018 · Zimmer Biomet, Inc. · initiated May 16, 2018
Class II Terminated All recalls posted in June 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 16, 2018 on May 16, 2018, to affected consignees via email and FedEx. The letter describes the product, problem and actions to be taken. Consignees were informed of the recall and instructed to take the following actions: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form must be returned even if you do not have affected products at your facility. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. For each return, send a copy of Attachment 1 to [email protected] b. International returns, please request an IRA by emailing [email protected] c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals included with the email notification sent to your facility, which includes a list of hospitals that have already been notified of this recall. b. Identify whether there are any additional hospitals that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format
Recall details
- Recall number
- Z-2235-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 16, 2018
- Posted by the FDA
- Jun 1, 2018
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 26
- Product code
- HTY
- FDA submissions
- K831005
- Distribution
- Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.
Product description
BALL TIP GUIDE WIRE 2.0MMX60CM; 469020 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
Code information (models, lots, serial numbers)
271231
Other devices under the same FDA submission
2 devices cleared under K831005 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- OMNIMAX 99-5824Biomet Microfixation, Inc · PIN, FIXATION, SMOOTH1 report
- LAG SCREW GUIDEWIRE 14-440052Biomet Orthopedics, LLC · PIN, FIXATION, SMOOTH1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 80193).
- Z-2230-2018 · TerminatedDRILL PT GUIDE WIRE 3.2MMX98CM, 469360 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended…
- Z-2231-2018 · TerminatedDRILL PT GUIDE WIRE 2MM X 60CM; 469320 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended…
- Z-2232-2018 · TerminatedBALL TIP GUIDE WIRE 3.2MMX98CM; 469060 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended…
- Z-2233-2018 · TerminatedBALL TIP GUIDE WIRE 3.2MMX55CM; 469055 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended…
- Z-2234-2018 · TerminatedGUIDE WIRE w/ BALL TIP 2.0MMX98CM; 469040 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Inten…
- Z-2236-2018 · TerminatedG-PIN 2.8BL 3.2PL 60CM 2PK; 467265 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for…
- Z-2237-2018 · TerminatedG-PIN 2.8BL 3.2PL 98CM 2PK; 467260 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for…
- Z-2238-2018 · TerminatedG-PIN 2.8BL 3.2PL 98CM 2PK; 467260 NL DRIVING GUIDE 4.4MM X 80CM; 467220 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nai…
- Z-2239-2018 · TerminatedBEAD TIP GD WIRE 3.0MM X 60CM; 29402 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended …
- Z-2240-2018 · Terminated3.2MM X 510MM COCR THD TIP; 14-443053 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended …
- Z-2241-2018 · TerminatedRECON DRILL 6MM X 439MM; 14-443023 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for…
- Z-2242-2018 · Terminated3.2MM X 560MM COCR THD TIP; 14-441053 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page