Medical device recalls: June 2018
The FDA posted 54 medical device recall records in June 2018, from 18 recall events by 15 companies. 2 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Zimmer Biomet, Inc. · 4 events, 16 records
- Orthofix, Inc · 1 event, 20 records
- Randox Laboratories Ltd. · 1 event, 3 records
- Becton Dickinson & Company · 1 event, 2 records
- Draeger Medical Systems, Inc. · 1 event, 2 records, 2 Class I
- Philips Medical Systems Nederlands · 1 event, 2 records
- BioMerieux SA · 1 event, 1 record
- Datex-Ohmeda, Inc. · 1 event, 1 record
FDA root cause
- Device Design · 26
- Under Investigation by firm · 22
- Nonconforming Material/Component · 2
- Vendor change control · 1
- Software design · 1
- Software change control · 1
- Radiation Control for Health and Safety Act · 1
Class I recalls (2)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2046-2018 · Draeger Medical Systems, Inc. · posted Jun 7, 2018 · Terminated
Users have misinterpreted the display for out of range measurement indicated by the blinking" ---" to mean a zero measurement.
- Z-2047-2018 · Draeger Medical Systems, Inc. · posted Jun 7, 2018 · Terminated
Users have misinterpreted the display for out of range measurement indicated by the blinking "-0-" to mean a zero measurement
Class II recalls (52)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2317-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Jun 29, 2018 · Terminated
A subset of Medtronic Model 53401 Single Chamber External Pulse Generators (EPGs) may revert from programmed settings to nominal settings during use with a patient due to an unexpected Power on Reset (POR).
- Z-2331-2018 · BioMerieux SA · posted Jun 29, 2018 · Terminated
Invalid calibration with low calibrator S1 while using the product.
- OmniDiagnost-Eleva X-ray system Angiographic and Diagnostic, Product Nos. 708026, 708027, 708028 ↗Class IIZ-2300-2018 · Philips Medical Systems Nederlands · posted Jun 27, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the Omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. This rotating movement can not be stopped by the user. This could potentially lead…
- Z-2301-2018 · Philips Medical Systems Nederlands · posted Jun 27, 2018 · Terminated
The fixation of the upper and lower tilt actuator, of the Omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. This rotating movement can not be stopped by the user. This could potentially lead…
- Z-2424-2018 · Ion Beam Applications S.A. · posted Jun 20, 2018 · Terminated
Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.
- Z-2287-2018 · Teleflex Medical Europe Ltd · posted Jun 14, 2018 · Terminated
The incorrect plastic was used to manufacture the handles of these shears, which causes the handles to melt if sterilized.
- Z-2278-2018 · Randox Laboratories Ltd. · posted Jun 13, 2018 · Terminated
The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in…
- Z-2279-2018 · Randox Laboratories Ltd. · posted Jun 13, 2018 · Terminated
The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in…
- Z-2280-2018 · Randox Laboratories Ltd. · posted Jun 13, 2018 · Terminated
The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in…
- Z-2058-2018 · Zeiss, Carl Inc · posted Jun 9, 2018 · Terminated
In certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit.
- Biograph mCT Systems: ↗Class IIZ-2248-2018 · Siemens Medical Solutions USA, Inc. · posted Jun 8, 2018 · Terminated
Siemens Medical Solutions USA, Inc., Molecular Imaging business line became aware, through our supplier that some laser cover windows may fall off a limited population of the Biograph mCT and Biograph mCT Flow systems.
- Z-2116-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2117-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Rival Reduce Kit Plates, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-012CClass IIZ-2118-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2119-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2120-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2121-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2122-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2123-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2124-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2125-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2126-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2127-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2128-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2129-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2130-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2131-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2132-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2133-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2134-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2135-2018 · Orthofix, Inc · posted Jun 7, 2018 · Terminated
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Z-2256-2018 · PENTAX of America Inc · posted Jun 5, 2018 · Terminated
The catheter alignment arrow printed onto the strain relief is not aligned with the detachment feature of the catheter connector, resulting in the potential for the catheter to be attached to the controller in the detach orientation. In this orientation,…
- Vanguard Complete Knee SystemClass IIZ-2289-2018 · Zimmer Biomet, Inc. · posted Jun 4, 2018 · Terminated
Two lots of tibial bearings were commingled. There is a possibility that a 14mm tibial bearing is packaged in a box labeled as 12 mm tibial bearing and vice versa. This may lead to extension of surgery to find another implant.
- Z-2051-2018 · Mako Surgical Corporation · posted Jun 1, 2018 · Terminated
Possible hinge pin disassociation from the Mako Onlay Insert Extractor.
- BD VACUTAINER PLUS SST Tube with HEMOGARD closure 13x75mm 3.5ml PLBL Gold, Catalog No. 367983Class IIZ-2055-2018 · Becton Dickinson & Company · posted Jun 1, 2018 · Terminated
Tubes may exhibit stopper creepout/pullout, where the stopper dissociates from the tube
- BD VACUTAINER PLUS SST Tube with HEMOGARD closure 13x100mm 5.0ml PLBL Gold, Catalog No. 367986Class IIZ-2056-2018 · Becton Dickinson & Company · posted Jun 1, 2018 · Terminated
Tubes may exhibit stopper creepout/pullout, where the stopper dissociates from the tube
- Z-2177-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented.
- Z-2183-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The label master file was errantly set up as a 63/37mm instead of a 63/67mm. There is no 63/37 size offered, and the product is laser marked with the correct size.
- Z-2230-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- DRILL PT GUIDE WIRE 2MM X 60CMClass IIZ-2231-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- BALL TIP GUIDE WIRE 3.2MMX98CMClass IIZ-2232-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- BALL TIP GUIDE WIRE 3.2MMX55CMClass IIZ-2233-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- Z-2234-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- BALL TIP GUIDE WIRE 2.0MMX60CMClass IIZ-2235-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- G-PIN 2.8BL 3.2PL 60CM 2PKClass IIZ-2236-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- G-PIN 2.8BL 3.2PL 98CM 2PKClass IIZ-2237-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- G-PIN 2.8BL 3.2PL 98CM 2PKClass IIZ-2238-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- BEAD TIP GD WIRE 3.0MM X 60CMClass IIZ-2239-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- 3.2MM X 510MM COCR THD TIPClass IIZ-2240-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- RECON DRILL 6MM X 439MMClass IIZ-2241-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- 3.2MM X 560MM COCR THD TIPClass IIZ-2242-2018 · Zimmer Biomet, Inc. · posted Jun 1, 2018 · Terminated
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the…
- Z-2266-2018 · Datex-Ohmeda, Inc. · posted Jun 1, 2018 · Terminated
Potential for the display processor to experience an unexpected failure.