JOURNEY BCS OXINIUM Knee Femoral Components, Catalog… recallZ-0002-2019 · Smith & Nephew, Inc. · initiated Jun 13, 2018
Class II Open, Classified All recalls posted in October 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The data in the National Joint Registry of England, Wales and Northern Ireland (NJREWNI) and Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) indicate that the first generation JOURNEY BCS Knee System has a revision rate over 1.5 times the primary total knee arthroplasty device class average revision rates reported in those registries.
FDA root cause category: Other
What the company is doing
The Field Safety Notice (FSN) sent to risk managers on 06/13/2018 asks customers to inspect inventory, locate and quarantine affected, unused Journey BCS Knee CoCr Femoral Components and JOURNEY BCS OXINIUM Knee Femoral Components. Affected product subject of the voluntary removal is to be returned to Smith & Nephew. All consignees are asked to complete an FSN Response Form, even if they do not have product to return. The Field Safety Notice, communicated with surgeons that these first generation JOURNEY BCS tibial inserts should only be used for polyethylene exchange revision of the first generation JOURNEY BCS total knee constructs where the femoral component and tibial baseplate are well-fixed. This notification, addressed to implanting surgeons, requesting their receipt and acknowledgement of the FSN.
Recall details
- Recall number
- Z-0002-2019
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jun 13, 2018
- Posted by the FDA
- Oct 1, 2018
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- 42,050 units
- Product code
- JWH
- FDA submissions
- K042515
- Distribution
- Worldwide distribution, including US nationwide and Puerto Rico, Austria, United Arab Emirates, Australia, Belgium, Canada, China, Switzerland, Germany, Denmark, Finland, France, Great Britain, India, Spain, Italy, Netherlands, Portugal, Sweden, Venezuela, South Africa.
Product description
JOURNEY BCS OXINIUM Knee Femoral Components, Catalog Numbers: 74021110, 74021111, 74021112, 74021113, 74021114, 74021115, 74021116, 74021117, 74021118, 74021119, 74021120, 74021121, 74021122, 74021123, 74021124, 74021125, 74021126, 74021127, 74021128, 74021129, 74021130, 74021131, 74021132, 74021133, 74021134, 74021135, 74021136, 74021137, 74021138, 74021139, 74021140, 74021141, 74021142, 74021143, 74021144, 74021145, 74021146, 74021147, 74021148, 74021149
Code information (models, lots, serial numbers)
All lots
Other devices under the same FDA submission
74 devices cleared under K042515 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- JOURNEY 74022213Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER48 reports
- JOURNEY 74024835Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER42 reports
- JOURNEY 74022223Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER40 reports
- JOURNEY 74024832Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER29 reports
- JOURNEY 74022216Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER28 reports
- JOURNEY 74022224Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER28 reports
- JOURNEY 74023251Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER28 reports
- JOURNEY 74022214Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER26 reports
- JOURNEY 74022225Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER25 reports
- JOURNEY 74023271Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER19 reports
- JOURNEY 74023261Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- JOURNEY 74022215Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER16 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80313).
- Z-0001-2019 · Open, ClassifiedJourney BCS Knee CoCr Femoral Components, Catalog Numbers: 74021210, 74021211, 74021212, 74021213, 74021214, 74021215, 74021216, 74021217, 74021218, 74021219, 74021220, 74021221, 7…
- Z-0003-2019 · Open, ClassifiedJOURNEY BCS Knee Conventional Polyethylene Inserts: 74023111, 74023112, 74023113, 74023114, 74023115, 74023116, 74023117, 74023118, 74023121, 74023122, 74023123, 74023124, 7402312…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page