Endo GIA 45 mm Articulating Medium/Thick Reload with… recallZ-2503-2018 · Covidien Medtronic · initiated May 22, 2018
Class II Open, Classified All recalls posted in July 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien Medtronic or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
FDA root cause category: Process control
What the company is doing
On May 22, 2018, all U.S. consignees received an UPDATED URGENT MEDICAL DEVICE RECALL and RETURN FORM notice via Federal Express or certified. The initial notice was sent to impacted customers on April 26, 2018. Customers are advised to take the following required actions: 1. Immediately discontinue use, and quarantine affected products listed 2. Return affected product as indicated. 3. Complete the Recalled Product Return Form even if you do not have inventory and submit by either method below: * Product purchased directly from Medtronic: [email protected] or fax to (800) 895-6140. *Product purchased through distributor: [email protected] or fax it to (203) 492-7719.
Recall details
- Recall number
- Z-2503-2018
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- May 22, 2018
- Posted by the FDA
- Jul 19, 2018
- Recalling firm
- Covidien Medtronic
- Product code
- GDW
- FDA submissions
- K111825
- Distribution
- Nationwide U.S.A. Internationally to: Albania, Algeria, Austria, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Iceland, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Malta, Martinique, Moldova, Morocco, Namibia, Netherlands, New Caledonia, Norway, Pakistan, Poland, Portugal, Reunion, Romania, Russian federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arab Emirates & United Kingdom
Product description
Endo GIA 45 mm Articulating Medium/Thick Reload with Tri-Staple Technology, Item Code EGIA45AMT
Code information (models, lots, serial numbers)
N7L0521KX, N7M0047KX, N7M0592KX, N7M1001KX, N7L0975KX, N7M0434KX, N7M0683KX, N8A0203KX, N7L1057KX, N7M0528KX, N7M0937KX, N8A0677KX, N7L1098KX, N7M0529KX, N7M0949KX, N8C0144KX & N7L1099KX
Other devices under the same FDA submission
77 devices cleared under K111825 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- Endo GIA EGIA45AVMCovidien LP · STAPLE, IMPLANTABLE7 deaths1,583 reports
- Endo GIA EGIA60AVMCovidien LP · STAPLE, IMPLANTABLE9 deaths1,438 reports
- GIA GIA8038SCovidien LP · STAPLE, IMPLANTABLE12 deaths1,237 reports
- Endo GIA 030458Covidien LP · STAPLE, IMPLANTABLE1 recall1 death948 reports
- Endo GIA EGIA60AXTCovidien LP · STAPLE, IMPLANTABLE4 deaths987 reports
- Endo GIA 030415Covidien LP · STAPLE, IMPLANTABLE973 reports
- Endo GIA 030449Covidien LP · STAPLE, IMPLANTABLE2 deaths716 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80423).
- Z-2504-2018 · Open, ClassifiedEndo GIA 60 mm Articulating Medium/Thick Reload with Tri-StapleTechnology, Item Code EGIA60AMT
- Z-2505-2018 · Open, ClassifiedEndo GIA 60 mm Extra Thick Black Articulating Reload with Tri-Staple Technology, Item Code EGIA60AXT
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page