Endo GIA EGIA45AMT (UDI 10884523003154)Recalls & reported problems
Covidien LP · Articulation Reload with Tri-Staple Technology · Class 2 · Implant
All models of Endo GIA (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Endo GIA reload with Tri-Staple technology is a cartridge of surgical staples used with a stapler. Its 2,969 adverse event reports mostly describe the reload failing to fire, misfiring, or failing to form staples, with some reports of parts breaking. The reports include 10 deaths and 324 injuries. No recalls name this device. Reports fell to 140 over the last year, from 244 the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,965
- Deaths reported
- 10
- Injuries reported
- 325
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2,965 adverse event reports that list this exact device (DI 10884523003154), received between May 23, 2016 and Sep 28, 2026 — classified as 10 deaths, 325 injuries, 2,630 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (1,583 reports).
73 other recalls cover devices cleared under the same FDA submission (K111825); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire1,583
- Misfire435
- Failure to Form Staple401
- Adverse Event Without Identified Device or Use Problem248
- Difficult to Open or Close179
- Break158
- Entrapment of Device151
- Human-Device Interface Problem123
Patient problems
- No Clinical Signs, Symptoms or Conditions1,140
- No Known Impact Or Consequence To Patient621
- No Consequences Or Impact To Patient574
- Tissue Breakdown166
- Insufficient Information141
- No Code Available141
- Unspecified Tissue Injury119
- Hemorrhage/Blood Loss/Bleeding95
Recent adverse event reports
The 25 most recent of 2,965 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Misfire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50924898 ↗MalfunctionSep 23, 2026Misfire
According to the reporter, during a laparoscopic appendectomy procedure, the knife advanced first. A new handle and reload were used to resolve the issue. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50864422 ↗MalfunctionSep 22, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859384 ↗MalfunctionSep 18, 2026Failure to Fire; Expulsion
According to the reporter, during a thoracoscopic right partial lung resection, during parenchymal resection, after closing the jaws, the green button was pressed but the firing could not be done at all. The staples came out around the proximal side of the jaws. A new reload was used to resolve the issue. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50816978 ↗InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50599844 ↗MalfunctionAug 26, 2026Difficult to Open or Close
Patient problems: Insufficient Information
MAUDE report 50460223 ↗MalfunctionAug 24, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50421776 ↗MalfunctionAug 24, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50420767 ↗MalfunctionAug 24, 2026Misfire; Insufficient Device Problem Information
Patient problems: Insufficient Information
MAUDE report 50418419 ↗MalfunctionAug 24, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50416595 ↗MalfunctionAug 17, 2026Failure to Fire; Expulsion
According to the reporter, during a laparoscopic rectal resection, when the jaws were closed and after pressing the green button, the firing could not be performed at all. A stapling near the proximal side of the cartridge's jaws was found after the firing attempt. A new cartridge was used to resolve the issue. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50275606 ↗MalfunctionAug 16, 2026Misfire
It was reported that during a procedure, the reload had a firing issue and was reported as not working. A new reload was used to resolve the issue. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50267715 ↗MalfunctionAug 14, 2026Entrapment of Device; Misfire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50251333 ↗MalfunctionAug 12, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50221597 ↗MalfunctionAug 12, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50220212 ↗MalfunctionAug 12, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50220203 ↗MalfunctionAug 12, 2026Entrapment of Device; Human-Device Interface Problem
Patient problems: Tissue Breakdown
MAUDE report 50219740 ↗MalfunctionAug 11, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Failure to Anastomose; Tissue Breakdown
MAUDE report 50204928 ↗MalfunctionJul 30, 2026Failure to Form Staple
Patient problems: Failure to Anastomose; Tissue Breakdown
MAUDE report 50093780 ↗MalfunctionJul 7, 2026Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC COLECTOMY, THE JAWS WERE CLOSED AND THE GREEN BUTTON WAS PRESSED BUT THE FIRING COULD NOT BE DONE AT ALL. A NEW RELOAD WAS USED WITH THE SAME HANDLE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25734783 ↗MalfunctionJun 29, 2026Break; Entrapment of Device
IT WAS REPORTED THAT, DURING A LUNG SEGMENTECTOMY PROCEDURE, THE RELOAD CONNECTED TO THE ADAPTER BECAME STUCK CLOSED UPON FIRING AND WOULD NOT OPEN, REMAINING LODGED IN THE PATIENT. IT WAS ALSO REPORTED THAT THE RELOAD SHAFT BROKE, AND THE BROKEN PART DID NOT DISENGAGE. THE SURGICAL TIME WAS EXTENDED BY 40 MINUTES. THE DEVICES WERE REPLACED WITH ANOTHER DEVICE TO RESOLVE THE ISSUE.
Patient problems: Tissue Breakdown
MAUDE report 25661629 ↗MalfunctionJun 15, 2026Failure to Fire; Insufficient Device Problem Information
ACCORDING TO THE REPORTER DURING A THORACOSCOPIC LOBECTOMY, AT THE TIME OF LOBAR TRANSECTION, THE STAPLER DID NOT FIRE. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 25498026 ↗MalfunctionJun 5, 2026Retraction Problem
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, AFTER FIRING TO RESECT THE RECTUM, THE KNIFE BLADE WAS DIFFICULT TO RETRACT. THE SURGEON SQUEEZED THE HANDLE OF THE STAPLER AGAIN TO RETRACT THE KNIFE TO RESOLVE THE ISSUE, AND THE HANDLE WORKED NORMALLY WITH A NEW RELOAD. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25411281 ↗MalfunctionJun 4, 2026Entrapment of Device; Retraction Problem; Misfire
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE FIRING WAS INTERRUPTED AT THE BEGINNING, IT WAS ALSO DIFFICULT TO RETRACT THE KNIFE BLADE AND THE DEVICE WAS LOCKED ON TISSUE. ADDITIONALLY, AT THE ONSET OF THE CARTRIDGE FIRING SEQUENCE, WHEN THE ERROR OCCURRED, THE FOLLOWING INFORMATION WAS DISPLAYED ON THE SCREEN: WITHIN THE POWERED STAPLER BLOCK, THE STATUS APPEARED IN GREEN; WITHIN THE ADAPTER BLOCK, IT APPEARED IN BLACK WITH A GREEN BORDER; AND WITHIN THE CARTRIDGE BLOCK, A CARTRIDGE ICON WAS DISPLAYED ALONGSIDE A YELLOW TRIANGLE CONTAINING A SYMBOL INDICATING THAT THE TISSUE HAD DRIED OUT, THEREBY NECESSITATING MANUAL SUTURING USING THE DEVICE. THE SURGEON WAS COMPELLED TO PERFORM A TISSUE RESECTION ON THE BRONCHUS, LARGER THAN NECESSARY IN ORDER TO REMOVE THE LOAD AND ADAPTER, WHICH LED TO EXTENDED SURGICAL TIME OF MORE THAN THREE HOURS.
Patient problems: Tissue Breakdown
MAUDE report 25403170 ↗MalfunctionJun 3, 2026Failure to Fire
ACCORDING TO THE REPORTER, DURING A GALLBLADDER PROCEDURE, AFTER PRESSING THE GREEN BUTTON, THE HANDLE COULD NOT BE SQUEEZE AND THE FIRING COULD NOT BE DONE. A NEW HANDLE AND RELOAD WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25375519 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111825. One submission often covers many models, so check the product and lot details in each record.
- Z-0748-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection
- Z-0747-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or tra
- Z-0005-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit 45 mm - 2.5mm, Item Code 030454 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric
- Z-0008-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 60 mm - 3.5 mm, Item Code 030458 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0003-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 3.5 mm, Item Code 030452 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0007-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit, Item Code 030457 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic
- Z-0006-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 3.5 mm, Item Code 030455 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0004-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 2.0 mm, Item Code 030453 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0001-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Gray Universal Articulating Loading Unit 30 mm - 2.0 mm, Item Code 030450 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gyn
- Z-0002-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 2.5 mm, Item Code 030451 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0009-2021 · Aug 12, 2020 · Open, ClassifiedCovidien Best Practices Procedure Kits containing Auto Suture Endo GIA Universal Loading Units, Parent Codes 00Z2760 - Product Usage: The Endo GIA universal staplers have applicat
- Z-1264-2020 · Jan 17, 2020 · TerminatedCovidien Endo GIA Ultra Universal Stapler 12 mm-intended for resection, transection in general surgery Item Code: EGIAUSTND
- Z-1836-2019 · May 17, 2019 · Open, ClassifiedSIGNIA TRI-STAPLE 2.0 CURVED TIP 60MM MEDIUM/ THIC , SIG60CTAMT
- Z-1828-2019 · May 17, 2019 · Open, ClassifiedEGIA 60 ARTICULATING EXTRA THICK SULU, Item code EGIA60AXT
- Z-1832-2019 · May 17, 2019 · Open, ClassifiedSIGNIA TRI-STAPLE 2.0 30MM MEDIUM/ THICK RELOAD , SIG30AMT
…and 58 more.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
- ECHELON FLEX PVE35AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE29 deaths1,880 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page