Percepta Quad CRT-P MRI SureScan recallZ-2413-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated Jun 21, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
FDA root cause category: Device Design
What the company is doing
On 21-Jun-2018, Medtronic began notifying US physicians and health care professionals of this issue and provided Patient Management recommendations to ensure patient safety. A Software Update to SW040 Version 8.1 will be deployed to all Medtronic 2090 and Encore clinic and hospital programmers via the Software Distribution Network or USB. Field Representatives with customers who have Percepta CRT-P patients will be sent a USB device loaded with the updated software. In consultation with an Independent Physician Quality Panel, Medtronic recommends the following actions for physicians following patients: - Contact your local Medtronic Representative and schedule installation of the updated Percepta CRT-P Application Software (SW040 Version 8.1) onto Medtronic 2090 and Encore clinic and hospital programmers. - For a patient whose Percepta CRT-P device has experienced a Reset Alert or Observation: Consider scheduling an in-clinic device interrogation as soon as possible for the patients device to receive the automatic update. - For a patient whose Percepta CRT-P device has not experienced a Reset Alert or Observation: At their next scheduled in-clinic device interrogation, the patients device will receive the automatic update. Please complete the enclosed Clinician Confirmation Certificate and return via email to [email protected]. For questions, contact your local Medtronic Representative or Medtronic Technical Services at 800-505-4636.
Recall details
- Recall number
- Z-2413-2018
- Status
- Terminated
- Initiated by the company
- Jun 21, 2018
- Posted by the FDA
- Jul 10, 2018
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Quantity
- 6712 units
- Product code
- NKE
- FDA submissions
- P010015
- Distribution
- Worldwide Distribution - US Nationwide
Product description
Percepta Quad CRT-P MRI SureScan, REF W4TR01 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and regulate the patient s heart rate by providing single or dual chamber rate responsive bradycardia pacing, sequential biventricular pacing, and atrial tachyarrhythmia therapies. These devices are indicated for heart failure patients who may have atrial arrhythmias or AV block.
Code information (models, lots, serial numbers)
Show the full code information (74k characters)
GTIN/UPN: (a) 00643169735644 Serial Numbers: RNP202329H RNP206542H RNP204079H RNP204773H RNP204816H RNP206419H RNP206432H RNP206445H RNP206456H RNP206491H RNP206511H RNP206515H RNP206522H RNP206527H RNP206528H RNP206531H RNP206541H RNP206545H RNP206550H RNP206551H RNP206552H RNP206555H RNP206560H RNP206561H RNP206568H RNP206576H RNP206577H RNP206578H RNP206580H RNP206581H RNP206586H RNP206587H RNP206574H RNP206595H RNP206598H RNP206579H RNP206606H RNP206610H RNP206582H RNP206584H RNP206585H RNP206611H RNP206614H RNP206588H RNP206615H RNP206590H RNP206591H RNP206592H RNP206593H RNP206616H RNP206596H RNP206597H RNP206618H RNP206599H RNP206601H RNP206602H RNP206605H RNP206627H RNP206629H RNP206630H RNP206612H RNP206635H RNP206638H RNP206639H RNP206640H RNP206620H RNP206621H RNP206622H RNP206641H RNP206643H RNP206647H RNP206631H RNP206632H RNP206633H RNP206655H RNP206660H RNP206663H RNP206664H RNP206642H RNP206674H RNP206675H RNP206651H RNP206682H RNP206690H RNP206661H RNP206693H RNP206707H RNP206710H RNP206711H RNP206678H RNP206712H RNP206716H RNP206688H RNP206719H RNP206725H RNP206734H RNP206769H RNP206773H RNP206775H RNP206776H RNP206176H RNP206338H RNP206415H RNP206416H RNP206417H RNP206418H RNP206420H RNP206421H RNP206423H RNP206424H RNP206430H RNP206431H RNP206433H RNP206435H RNP206441H RNP206442H RNP206443H RNP206444H RNP206447H RNP206448H RNP206450H RNP206452H RNP206453H RNP206455H RNP206459H RNP206460H RNP206461H RNP206462H RNP206463H RNP206464H RNP206465H RNP206467H RNP206468H RNP206469H RNP206470H RNP206473H RNP206475H RNP206478H RNP206479H RNP206481H RNP206482H RNP206485H RNP206486H RNP206487H RNP206488H RNP206489H RNP206490H RNP206493H RNP206496H RNP206497H RNP206498H RNP206501H RNP206502H RNP206503H RNP206504H RNP206505H RNP206506H RNP206508H RNP206509H RNP206510H RNP206512H RNP206513H RNP206514H RNP206517H RNP206521H RNP206523H RNP206524H RNP206525H RNP206526H RNP206529H RNP206530H RNP206533H RNP206534H RNP206535H RNP206537H RNP206538H RNP206539H RNP206544H RNP206546H RNP206548H RNP206556H RNP206563H RNP206565H RNP206566H RNP206567H RNP206569H RNP206570H RNP206771H RNP206772H RNP206654H RNP206687H RNP206717H RNP206743H RNP206751H RNP203505H RNP205985H RNP205989H RNP206042H RNP206046H RNP206047H RNP206052H RNP206055H RNP206056H RNP206057H RNP206061H RNP206062H RNP206067H RNP206073H RNP206075H RNP206083H RNP206085H RNP206089H RNP206093H RNP206101H RNP206110H RNP206111H RNP206117H RNP206127H RNP206129H RNP206132H RNP206144H RNP206152H RNP206157H RNP206161H RNP206164H RNP206165H RNP206166H RNP206167H RNP206168H RNP206169H RNP206172H RNP206203H RNP206205H RNP206209H RNP206211H RNP206212H RNP206214H RNP206222H RNP206225H RNP206244H RNP206245H RNP206248H RNP206250H RNP206253H RNP206254H RNP206255H RNP206257H RNP206259H RNP206268H RNP206270H RNP206273H RNP206274H RNP206284H RNP206290H RNP206305H RNP206307H RNP206312H RNP206317H RNP206321H RNP206324H RNP206328H RNP206335H RNP206337H RNP206346H RNP206348H RNP206355H RNP206360H RNP206372H RNP206378H RNP206384H RNP206392H RNP206395H RNP206397H RNP206401H RNP206402H RNP206410H RNP200362H RNP200497H RNP201286H RNP201695H RNP202308H RNP202582H RNP202776H RNP202795H RNP202871H RNP203540H RNP203921H RNP204072H RNP204498H RNP204676H RNP205117H RNP206171H RNP206178H RNP206179H RNP206181H RNP206184H RNP206186H RNP206187H RNP206188H RNP206190H RNP206191H RNP206192H RNP206193H RNP206195H RNP206196H RNP206197H RNP206198H RNP206199H RNP206200H RNP206202H RNP206204H RNP206206H RNP206207H RNP206208H RNP206210H RNP206213H RNP206215H RNP206216H RNP206217H RNP206218H RNP206219H RNP206220H RNP206221H RNP206223H RNP206224H RNP206226H RNP206227H RNP206228H RNP206229H RNP206230H RNP206232H RNP206233H RNP206234H RNP206235H RNP206236H RNP206237H RNP206239H RNP206240H RNP206242H RNP206243H RNP206247H RNP206249H RNP206251H RNP206252H RNP206256H RNP206258H RNP206260H RNP206261H RNP206262H RNP206263H RNP206264H RNP206265H RNP206266H RNP206269H RNP206271H RNP206272H RNP206275H RNP206276H RNP206277H RNP206278H RNP206279H RNP206280H RNP206281H RNP206282H RNP206283H RNP206285H RNP206286H RNP206287H RNP206288H RNP206289H RNP206291H RNP206292H RNP206293H RNP206294H RNP206295H RNP206297H RNP206298H RNP206299H RNP206300H RNP206301H RNP206302H RNP206303H RNP206304H RNP206306H RNP206308H RNP206310H RNP206311H RNP206313H RNP206314H RNP206315H RNP206316H RNP206318H RNP206319H RNP206320H RNP206322H RNP206323H RNP206325H RNP206326H RNP206327H RNP206329H RNP206331H RNP206332H RNP206333H RNP206334H RNP206340H RNP206342H RNP206343H RNP206344H RNP206345H RNP206347H RNP206349H RNP206350H RNP206351H RNP206352H RNP206353H RNP206354H RNP206357H RNP206358H RNP206359H RNP206362H RNP206363H RNP206364H RNP206366H RNP206367H RNP206368H RNP206369H RNP206370H RNP206371H RNP206373H RNP206374H RNP206375H RNP206376H RNP206377H RNP206379H RNP206380H RNP206381H RNP206382H RNP206383H RNP206385H RNP206387H RNP206388H RNP206389H RNP206390H RNP206391H RNP206393H RNP206394H RNP206396H RNP206398H RNP206400H RNP206403H RNP206404H RNP206405H RNP206406H RNP206407H RNP206413H RNP200714H RNP200716H RNP200717H RNP200719H RNP200722H RNP200723H RNP200728H RNP200731H RNP200733H RNP200734H RNP200735H RNP200736H RNP200739H RNP200741H RNP200742H RNP200744H RNP200748H RNP200750H RNP200751H RNP200752H RNP200754H RNP200755H RNP200759H RNP200765H RNP200775H RNP200779H RNP200782H RNP200783H RNP200784H RNP200786H RNP200791H RNP200796H RNP200797H RNP200810H RNP201810H RNP202796H RNP202806H RNP202841H RNP202861H RNP202872H RNP202876H RNP202884H RNP206589H RNP203684H RNP203753H RNP206518H RNP206557H RNP206765H RNP206766H RNP206767H RNP206770H RNP206774H RNP206617H RNP206623H RNP206625H RNP206626H RNP206628H RNP206634H RNP206636H RNP206644H RNP206645H RNP206646H RNP206649H RNP206652H RNP206653H RNP206656H RNP206657H RNP206658H RNP206659H RNP206673H RNP206676H RNP206679H RNP206680H RNP206681H RNP206685H RNP206689H RNP206691H RNP206692H RNP206694H RNP206695H RNP206697H RNP206698H RNP206699H RNP206700H RNP206708H RNP206709H RNP206715H RNP206718H RNP206720H RNP206721H RNP206722H RNP206723H RNP206724H RNP206727H RNP206729H RNP206730H RNP206731H RNP206732H RNP206733H RNP206735H RNP206737H RNP206738H RNP206739H RNP206740H RNP206741H RNP206742H RNP206744H RNP206745H RNP206746H RNP206747H RNP206748H RNP206749H RNP206750H RNP206752H RNP206753H RNP206754H RNP206755H RNP206756H RNP206757H RNP206758H RNP206759H RNP206760H RNP206761H RNP206763H RNP205899H RNP206070H RNP205710H RNP206003H RNP206148H RNP202687H RNP204843H RNP205831H RNP205832H RNP204358H RNP204915H RNP204487H RNP204298H RNP204875H RNP205569H RNP202970H RNP204117H RNP201219H RNP204116H RNP201233H RNP202291H RNP205277H RNP205677H RNP202619H RNP206125H RNP205993H RNP206000H RNP205575H RNP201021H RNP205299H RNP204810H RNP203282H RNP204267H RNP205921H RNP202959H RNP205842H RNP205844H RNP203708H RNP204092H RNP205943H RNP201328H RNP202426H RNP204827H RNP203741H RNP203797H RNP204903H RNP204427H RNP205053H RNP203564H RNP204657H RNP204699H RNP205948H RNP204351H RNP200263H RNP204800H RNP205365H RNP204689H RNP200368H RNP201197H RNP204547H RNP204557H RNP205069H RNP204060H RNP201913H RNP205885H RNP205886H RNP205843H RNP204360H RNP202534H RNP205766H RNP200342H RNP205540H RNP204442H RNP206065H RNP203007H RNP205963H RNP204855H RNP203641H RNP204073H RNP204997H RNP205741H RNP203654H RNP204424H RNP202743H RNP204874H RNP205708H RNP205719H RNP204597H RNP202852H RNP206131H RNP202189H RNP203243H RNP203651H RNP205649H RNP203949H RNP205012H RNP201997H RNP201361H RNP202525H RNP203599H RNP203601H RNP203605H RNP202620H RNP205872H RNP204314H RNP204974H RNP206058H RNP204621H RNP202594H RNP204103H RNP204509H RNP204870H RNP205990H RNP206064H RNP201357H RNP204422H RNP204817H RNP202843H RNP203010H RNP204721H RNP204722H RNP202969H RNP202431H RNP202448H RNP202675H RNP202975H RNP204936H RNP204594H RNP206069H RNP201500H RNP206050H RNP205754H RNP202236H RNP206053H RNP206088H RNP204651H RNP204707H RNP206091H RNP205838H RNP205777H RNP205900H RNP205903H RNP206074H RNP205829H RNP205877H RNP204706H RNP203227H RNP205688H RNP205691H RNP206149H RNP202483H RNP202484H RNP202088H RNP205869H RNP205870H RNP205961H RNP205924H RNP206060H RNP206158H RNP206071H RNP202369H RNP206092H RNP204891H RNP205566H RNP204818H RNP204415H RNP203035H RNP204999H RNP204111H RNP205686H RNP201248H RNP203988H RNP201151H RNP206114H RNP205733H RNP205858H RNP206116H RNP206119H RNP205851H RNP204682H RNP206137H RNP204932H RNP200593H RNP202747H RNP204696H RNP205635H RNP205461H RNP201593H RNP205893H RNP205897H RNP206134H RNP206151H RNP203602H RNP203603H RNP204261H RNP204472H RNP202604H RNP204327H RNP205778H RNP200587H RNP203851H RNP202040H RNP204690H RNP206133H RNP205804H RNP204885H RNP201113H RNP205717H RNP203250H RNP204053H RNP204880H RNP204988H RNP205815H RNP204736H RNP205049H RNP204697H RNP204865H RNP204772H RNP201231H RNP203305H RNP204985H RNP205334H RNP204258H RNP204675H RNP202580H RNP202699H RNP202366H RNP203232H RNP204858H RNP202052H RNP204297H RNP205930H RNP203286H RNP203287H RNP203781H RNP206086H RNP204324H RNP200616H RNP205884H RNP206126H RNP201389H RNP203778H RNP203121H RNP204907H RNP204811H RNP205038H RNP205888H RNP205738H RNP202579H RNP204478H RNP201788H RNP201077H RNP201581H RNP203547H RNP204708H RNP204833H RNP203023H RNP200461H RNP204583H RNP206118H RNP206122H RNP205636H RNP205679H RNP205680H RNP205681H RNP204343H RNP205664H RNP203135H RNP206150H RNP204416H RNP205806H RNP206090H RNP204757H RNP200758H RNP202393H RNP205939H RNP203986H RNP205420H RNP204337H RNP205809H RNP203594H RNP205702H RNP204878H RNP205400H RNP205469H RNP205923H RNP206112H RNP205772H RNP204916H RNP205915H RNP206155H RNP206084H RNP204653H RNP201848H RNP202298H RNP205795H RNP204417H RNP205043H RNP206138H RNP205890H RNP204086H RNP203030H RNP204363H RNP206130H RNP206051H RNP204530H RNP204862H RNP204770H RNP203493H RNP203577H RNP203647H RNP203649H RNP205019H RNP205044H RNP205045H RNP205901H RNP205902H RNP205904H RNP205422H RNP205982H RNP204001H RNP202126H RNP204993H RNP203850H RNP203270H RNP204863H RNP204376H RNP204896H RNP203793H RNP204419H RNP205812H RNP205698H RNP205847H RNP204283H RNP206159H RNP202775H RNP203447H RNP202144H RNP201068H RNP201069H RNP201071H RNP204762H RNP204933H RNP205784H RNP203765H RNP203769H RNP204074H RNP202892H RNP203264H RNP205931H RNP200230H RNP201322H RNP201323H RNP201566H RNP203826H RNP203827H RNP205504H RNP200356H RNP201448H RNP202199H RNP203046H RNP203364H RNP203774H RNP204248H RNP200420H RNP202565H RNP202566H RNP202828H RNP203233H RNP200186H RNP201757H RNP200018H RNP200023H RNP200076H RNP200129H RNP200142H RNP200143H RNP200144H RNP200206H RNP200211H RNP200247H RNP200288H RNP200307H RNP200309H RNP200412H RNP200413H RNP200546H RNP200841H RNP200842H RNP200843H RNP200844H RNP200845H RNP200846H RNP200847H RNP200848H RNP200849H RNP200850H RNP200851H RNP200853H RNP200854H RNP200855H RNP200856H RNP200858H RNP200859H RNP200860H RNP200861H RNP200862H RNP200863H RNP200864H RNP200865H RNP200866H RNP200867H RNP200868H RNP200869H RNP200870H RNP200872H RNP200873H RNP200874H RNP200875H RNP200877H RNP200878H RNP200879H RNP200880H RNP200881H RNP200882H RNP202903H RNP202905H RNP202923H RNP202925H RNP203005H RNP203059H RNP203060H RNP203061H RNP203063H RNP203064H RNP203065H RNP203066H RNP203069H RNP203070H RNP203071H RNP203073H RNP203074H RNP203075H RNP203077H RNP203078H RNP204684H RNP205071H RNP205072H RNP205073H RNP205075H RNP205076H RNP205077H RNP205078H RNP205079H RNP205080H RNP205081H RNP205082H RNP205083H RNP205085H RNP205086H RNP205087H RNP205089H RNP205091H RNP205092H RNP205093H RNP205094H RNP205095H RNP205096H RNP205097H RNP205098H RNP205099H RNP205100H RNP205101H RNP205102H RNP205103H RNP205104H RNP205105H RNP205106H RNP205108H RNP205109H RNP205112H RNP205115H RNP205122H RNP205123H RNP205124H RNP205125H RNP205722H RNP200019H RNP200021H RNP200133H RNP200292H RNP201063H RNP201337H RNP201892H RNP202572H RNP203212H RNP203632H RNP204231H RNP204359H RNP204428H RNP204640H RNP204889H RNP205625H RNP205627H RNP205962H RNP201165H RNP200346H RNP200349H RNP201870H RNP202216H RNP202611H RNP202716H RNP203638H RNP203891H RNP203895H RNP203916H RNP200055H RNP200399H RNP200566H RNP200808H RNP201009H RNP201900H RNP202102H RNP202256H RNP203507H RNP203643H RNP203749H RNP204748H RNP204919H RNP203573H RNP204502H RNP200889H RNP204971H RNP201416H RNP202410H RNP202474H RNP202643H RNP202760H RNP201358H RNP201366H RNP202134H RNP202247H RNP202252H RNP203528H RNP203816H RNP204035H RNP205707H RNP200440H RNP201546H RNP200338H RNP200378H RNP200379H RNP200380H RNP200488H RNP200563H RNP200682H RNP200973H RNP200987H RNP200995H RNP201060H RNP201083H RNP201084H RNP201085H RNP201086H RNP201087H RNP201088H RNP201089H RNP201090H RNP201092H RNP201103H RNP201104H RNP201105H RNP201106H RNP201107H RNP201108H RNP201109H RNP203139H RNP203141H RNP203142H RNP203143H RNP203144H RNP203145H RNP203147H RNP203149H RNP203152H RNP203153H RNP203154H RNP203155H RNP203157H RNP203158H RNP203159H RNP204249H RNP204448H RNP204465H RNP204485H RNP204495H RNP204616H RNP204617H RNP204917H RNP204925H RNP204926H RNP204928H RNP205484H RNP205490H RNP205499H RNP205500H RNP205502H RNP205508H RNP205511H RNP205512H RNP205513H RNP205515H RNP205516H RNP201047H RNP201048H RNP201370H RNP201372H RNP201519H RNP201632H RNP201791H RNP201874H RNP200181H RNP204274H RNP204591H RNP204743H RNP201065H RNP202531H RNP200678H RNP200679H RNP200680H RNP200920H RNP200921H RNP200922H RNP200924H RNP200925H RNP201981H RNP201984H RNP202326H RNP202361H RNP202362H RNP202364H RNP202365H RNP202771H RNP203167H RNP203168H RNP203221H RNP203526H RNP203550H RNP203553H RNP203829H RNP203830H RNP204367H RNP204373H RNP204606H RNP204831H RNP204832H RNP204894H RNP205065H RNP205132H RNP205404H RNP205411H RNP205412H RNP205414H RNP205415H RNP205416H RNP205426H RNP205427H RNP205429H RNP205435H RNP203968H RNP203969H RNP203974H RNP200599H RNP200604H RNP201094H RNP201380H RNP202785H RNP203481H RNP203952H RNP204709H RNP204753H RNP201639H RNP203728H RNP204450H RNP202455H RNP204924H RNP201208H RNP202770H RNP203899H RNP200477H RNP200485H RNP201455H RNP201457H RNP201461H RNP201463H RNP201467H RNP201498H RNP203831H RNP203847H RNP203852H RNP204268H RNP204339H RNP204458H RNP204512H RNP205214H RNP205525H RNP201637H RNP200020H RNP200046H RNP200071H RNP200072H RNP200165H RNP200184H RNP200221H RNP200265H RNP200325H RNP200392H RNP201073H RNP201074H RNP201075H RNP201078H RNP201079H RNP201080H RNP201285H RNP201287H RNP201320H RNP201375H RNP201888H RNP203099H RNP203101H RNP203102H RNP203103H RNP203104H RNP203106H RNP203107H RNP203108H RNP203109H RNP203110H RNP203111H RNP203112H RNP203113H RNP203114H RNP203115H RNP204959H RNP204960H RNP204961H RNP204962H RNP204963H RNP204964H RNP204965H RNP204966H RNP205020H RNP205021H RNP205022H RNP205023H RNP205024H RNP205025H RNP205026H RNP205027H RNP205028H RNP205029H RNP205030H RNP205031H RNP205032H RNP205033H RNP205034H RNP205035H RNP205036H RNP201076H RNP201438H RNP201441H RNP202521H RNP204046H RNP204488H RNP200529H RNP201035H RNP201246H RNP201313H RNP202814H RNP203002H RNP203337H RNP203338H RNP203725H RNP201066H RNP204685H RNP201225H RNP202169H RNP203984H RNP202698H RNP200568H RNP201558H RNP202441H RNP202877H RNP203596H RNP200492H RNP202863H RNP202868H RNP204025H RNP201437H RNP202885H RNP205460H RNP204436H RNP204469H RNP205974H RNP203926H RNP200122H RNP200197H RNP200489H RNP200512H RNP201005H RNP201368H RNP201702H RNP202022H RNP202533H RNP202578H RNP200355H RNP200634H RNP201153H RNP203675H RNP203912H RNP203998H RNP204007H RNP204293H RNP205056H RNP200096H RNP200654H RNP201684H RNP201698H RNP202215H RNP202288H RNP203055H RNP203056H RNP203176H RNP203558H RNP203784H RNP203862H RNP203875H RNP204075H RNP204097H RNP204799H RNP200336H RNP201774H RNP200562H RNP200658H RNP200685H RNP200686H RNP201516H RNP201580H RNP202156H RNP202157H RNP202158H RNP202159H RNP203126H RNP203127H RNP201251H RNP202849H RNP204548H RNP202859H RNP203123H RNP202568H RNP205709H RNP200558H RNP200694H RNP201445H RNP202207H RNP202819H RNP203497H RNP204679H RNP200510H RNP200554H RNP201297H RNP201298H RNP201425H RNP202434H RNP203752H RNP203955H RNP204316H RNP200464H RNP200471H RNP200560H RNP200738H RNP201061H RNP204317H RNP202599H RNP202613H RNP202652H RNP202850H RNP200052H RNP204411H RNP203688H RNP204714H RNP204977H RNP204978H RNP205017H RNP200533H RNP200592H RNP200693H RNP202694H RNP203236H RNP204201H RNP200081H RNP200382H RNP200384H RNP200386H RNP200387H RNP200551H RNP200552H RNP201039H RNP201269H RNP201382H RNP201620H RNP201646H RNP201775H RNP202090H RNP202091H RNP202200H RNP202203H RNP202206H RNP202214H RNP202368H RNP202524H RNP202672H RNP202811H RNP202812H RNP202813H RNP203358H RNP203615H RNP203619H RNP203621H RNP204110H RNP204438H RNP204734H RNP204866H RNP204867H RNP204910H RNP204911H RNP205041H RNP202164H RNP202670H RNP205352H RNP200373H RNP201342H RNP202165H RNP203052H RNP203757H RNP201220H RNP201700H RNP201544H RNP202371H RNP202384H RNP202446H RNP202824H RNP202867H RNP201296H RNP202722H RNP203824H RNP204024H RNP204295H RNP204477H RNP204711H RNP204979H RNP202997H RNP201443H RNP204702H RNP205705H RNP205282H RNP201131H RNP201611H RNP201683H RNP201886H RNP202002H RNP202015H RNP202614H RNP202756H RNP202786H RNP202940H RNP203003H RNP203184H RNP203545H RNP203610H RNP203616H RNP204195H RNP204247H RNP204276H RNP204400H RNP204420H RNP205225H RNP205226H RNP205227H RNP201339H RNP202130H RNP200756H RNP201211H RNP201212H RNP200811H RNP202185H RNP202818H RNP203700H RNP200289H RNP200293H RNP200668H RNP201746H RNP202223H RNP204500H RNP204128H RNP201423H RNP201552H RNP201567H RNP201675H RNP202403H RNP202419H RNP202456H RNP202457H RNP202510H RNP202860H RNP202873H RNP202913H RNP202915H RNP203169H RNP203717H RNP203909H RNP204246H RNP204259H RNP204584H RNP204790H RNP205728H RNP206045H RNP202842H RNP203218H RNP204238H RNP205653H RNP200761H RNP201465H RNP201739H RNP202353H RNP202447H RNP203485H RNP203487H RNP204054H RNP205388H RNP200131H RNP201577H RNP202400H RNP202723H RNP203886H RNP204405H RNP205775H RNP200367H RNP201899H RNP203017H RNP203018H RNP201276H RNP203785H RNP200591H RNP200607H RNP200614H RNP201687H RNP201688H RNP202746H RNP203697H RNP203771H RNP202938H RNP200337H RNP204237H RNP205008H RNP204255H RNP205651H RNP200709H RNP200711H RNP202106H RNP202971H RNP203789H RNP200531H RNP201897H RNP202662H RNP202680H RNP202816H RNP202870H RNP203021H RNP203405H RNP203696H RNP203719H RNP203855H RNP203861H RNP203991H RNP204047H RNP204048H RNP204089H RNP204754H RNP205823H RNP201120H RNP201122H RNP200157H RNP200433H RNP201017H RNP201333H RNP201439H RNP201506H RNP202145H RNP202197H RNP202241H RNP202375H RNP202489H RNP202574H RNP202853H RNP203583H RNP204348H RNP204922H RNP205839H RNP205937H RNP200683H RNP203517H RNP203817H RNP200697H RNP202621H RNP203356H RNP203977H RNP204285H RNP205825H RNP200626H RNP204589H RNP205477H RNP205479H RNP205480H RNP205481H RNP205483H RNP205485H RNP205486H RNP205487H RNP205488H RNP205491H RNP200038H RNP200041H RNP200125H RNP200474H RNP200838H RNP201367H RNP201539H RNP202986H RNP203011H RNP203263H RNP203766H RNP204085H RNP204680H RNP200251H RNP201499H RNP201725H RNP202503H RNP200990H RNP202485H RNP203665H RNP201759H RNP200417H RNP201241H RNP201444H RNP201919H RNP202402H RNP202719H RNP202898H RNP203536H RNP203963H RNP204739H RNP204746H RNP200927H RNP201766H RNP200671H RNP200931H RNP203027H RNP204279H RNP201557H RNP200327H RNP200051H RNP204391H RNP202418H RNP202421H RNP202532H RNP202753H RNP202758H RNP204132H RNP201517H RNP202397H RNP200524H RNP200635H RNP201594H RNP200303H RNP200653H RNP201148H RNP202626H RNP202798H RNP200171H RNP202357H RNP202358H RNP202359H RNP203185H RNP203195H RNP203196H RNP203198H RNP203441H RNP203578H RNP203611H RNP203612H RNP203613H RNP203733H RNP203738H RNP204098H RNP204782H RNP204879H RNP205714H RNP203687H RNP200698H RNP202592H RNP203623H RNP205807H RNP200028H RNP200033H RNP200475H RNP201196H RNP202535H RNP203228H RNP203759H RNP203857H RNP203860H RNP204041H RNP205863H RNP200262H RNP201712H RNP203859H RNP204095H RNP205783H RNP203807H RNP205940H RNP200151H RNP200152H RNP201345H RNP201493H RNP202452H RNP202772H RNP202948H RNP204364H RNP204838H RNP205129H RNP205889H RNP200269H RNP200518H RNP201040H RNP201610H RNP202909H RNP201206H RNP204645H RNP204650H RNP200297H RNP200788H RNP202186H RNP202709H RNP202950H RNP201081H RNP201082H RNP202444H RNP204561H RNP200045H RNP200049H RNP202833H RNP203132H RNP204056H RNP203866H RNP204368H RNP204846H RNP200665H RNP202563H RNP202705H RNP201326H RNP200108H RNP202712H RNP204104H RNP205796H RNP201163H RNP201659H RNP201664H RNP201781H RNP202101H RNP202272H RNP202682H RNP203224H RNP203786H RNP202148H RNP202307H RNP203008H RNP203798H RNP203869H RNP204011H RNP204051H RNP204576H RNP204627H RNP204628H RNP204681H RNP204794H RNP201132H RNP201484H RNP202985H RNP200628H RNP201691H RNP201713H RNP200993H RNP201024H RNP201618H RNP202430H RNP202856H RNP204695H RNP205358H RNP200037H RNP200042H RNP200043H RNP201037H RNP202179H RNP202181H RNP202879H RNP204434H RNP204532H RNP204533H RNP202257H RNP203634H RNP201125H RNP202355H RNP202585H RNP202995H RNP204384H RNP205408H RNP205534H RNP202725H RNP200225H RNP201335H RNP201421H RNP200503H RNP200578H RNP200688H RNP201042H RNP202143H RNP202187H RNP202304H RNP203239H RNP203531H RNP203723H RNP203892H RNP204213H RNP204508H RNP204779H RNP204901H RNP204902H RNP204994H RNP205005H RNP205399H RNP205565H RNP205828H RNP200350H RNP200351H RNP200613H RNP201887H RNP202282H RNP202284H RNP202285H RNP202286H RNP202287H RNP201170H RNP201545H RNP202664H RNP203020H RNP200057H RNP201721H RNP202220H RNP201232H RNP201234H RNP201238H RNP201242H RNP200895H RNP200896H RNP200897H RNP202529H RNP203509H RNP203669H RNP203670H RNP203672H RNP203676H RNP205284H RNP205285H RNP205286H RNP205287H RNP205288H RNP201418H RNP202201H RNP200438H RNP203525H RNP204304H RNP200702H RNP201202H RNP201258H RNP201502H RNP201909H RNP202439H RNP200311H RNP200445H RNP201327H RNP202450H RNP202844H RNP202846H RNP204021H RNP204560H RNP204563H RNP204929H RNP201240H RNP201811H RNP202113H RNP202115H RNP202178H RNP202697H RNP203468H RNP203470H RNP200167H RNP200319H RNP200540H RNP200663H RNP200787H RNP200803H RNP201310H RNP201447H RNP201528H RNP201634H RNP202109H RNP202110H RNP202255H RNP203240H RNP203251H RNP204328H RNP204402H RNP204497H RNP204588H RNP204808H RNP205757H RNP205761H RNP200343H RNP200345H RNP200785H RNP200802H RNP201029H RNP201419H RNP201432H RNP201521H RNP201743H RNP201859H RNP202246H RNP202605H RNP202855H RNP202907H RNP203014H RNP203037H RNP203253H RNP203259H RNP204646H RNP205917H RNP204236H RNP202881H RNP200084H RNP202634H RNP202637H RNP203821H RNP200077H RNP200154H RNP200287H RNP200495H RNP200500H RNP200519H RNP200561H RNP200667H RNP200695H RNP200778H RNP200781H RNP201207H RNP201243H RNP203079H RNP203080H RNP203081H RNP203082H RNP203083H RNP203085H RNP203086H RNP203087H RNP203088H RNP203089H RNP203090H RNP203091H RNP203092H RNP203093H RNP203094H RNP203095H RNP203097H RNP203098H RNP203706H RNP204468H RNP204479H RNP204520H RNP205320H RNP205321H RNP205322H RNP205327H RNP205329H RNP205330H RNP205331H RNP201642H RNP202142H RNP202398H RNP203837H RNP201299H RNP202603H RNP202462H RNP202900H RNP202989H RNP203439H RNP203608H RNP204243H RNP202155H RNP200370H RNP200455H RNP201381H RNP201755H RNP201910H RNP201992H RNP202512H RNP202514H RNP203409H RNP203747H RNP203953H RNP204077H RNP204275H RNP204277H RNP204294H RNP204329H RNP204332H RNP205704H RNP201914H RNP206128H RNP200308H RNP200523H RNP200900H RNP200901H RNP200902H RNP204973H RNP202763H RNP202904H RNP204654H RNP202571H RNP200298H RNP200300H RNP200301H RNP200302H RNP202443H RNP202463H RNP203897H RNP205860H RNP205861H RNP201321H RNP200235H RNP202404H RNP203919H RNP203920H RNP203922H RNP204049H RNP204260H RNP204471H RNP204526H RNP204605H RNP204613H RNP204694H RNP204712H RNP204719H RNP205780H RNP205781H RNP205891H RNP205892H RNP205898H RNP200027H RNP200285H RNP201002H RNP201003H RNP201018H RNP201260H RNP201831H RNP201836H RNP201881H RNP201882H RNP201883H RNP202193H RNP202198H RNP202202H RNP202204H RNP202208H RNP202990H RNP203276H RNP203737H RNP203839H RNP203842H RNP203843H RNP203845H RNP203846H RNP205751H RNP205760H RNP205762H RNP200078H RNP200310H RNP200429H RNP200648H RNP200676H RNP201006H RNP201252H RNP201409H RNP202792H RNP205665H RNP205667H RNP201150H RNP200541H RNP204362H RNP203928H RNP203763H RNP200656H RNP200812H RNP201572H RNP201795H RNP201871H RNP202118H RNP202121H RNP202124H RNP202127H RNP203900H RNP203902H RNP203907H RNP203914H RNP200039H RNP200040H RNP200163H RNP200457H RNP200484H RNP201218H RNP201609H RNP201676H RNP201701H RNP202021H RNP202217H RNP202541H RNP202769H RNP202968H RNP203271H RNP203588H RNP203888H RNP204190H RNP204344H RNP204619H RNP204869H RNP205357H RNP205495H RNP205711H RNP202653H RNP201756H RNP203567H RNP200069H RNP200075H RNP200083H RNP200573H RNP201798H RNP202047H RNP202882H RNP205874H RNP200191H RNP202265H RNP203343H RNP202832H RNP203999H RNP204710H RNP204814H RNP200044H RNP200649H RNP202025H RNP202667H RNP202685H RNP202738H RNP203076H RNP203746H RNP204492H RNP204615H RNP204840H RNP200726H RNP201272H RNP201289H RNP201204H RNP202596H RNP200996H RNP201277H RNP201556H RNP201614H RNP202391H RNP202414H RNP203446H RNP203496H RNP201660H RNP202956H RNP202269H RNP200606H RNP200771H RNP201119H RNP201144H RNP201200H RNP202227H RNP202591H RNP203790H RNP203894H RNP203957H RNP203992H RNP204038H RNP204090H RNP204338H RNP204345H RNP204369H RNP204386H RNP205734H RNP205736H RNP205737H RNP205740H RNP205743H RNP205746H RNP200118H RNP200120H RNP201462H RNP201911H RNP202659H RNP203326H RNP201406H RNP200590H RNP200582H RNP202433H RNP202435H RNP203435H RNP203467H RNP203560H RNP203664H RNP202496H RNP204026H RNP204751H RNP200034H RNP200113H RNP201186H RNP201307H RNP201631H RNP201663H RNP202559H RNP202616H RNP202617H RNP202827H RNP202865H RNP203853H RNP203873H RNP205439H RNP205440H RNP205445H RNP205451H RNP205452H RNP205454H RNP205455H RNP205457H RNP205458H RNP205459H RNP200402H RNP204288H RNP205014H RNP204637H RNP200315H RNP200651H RNP202983H RNP200258H RNP200260H RNP200570H RNP201043H RNP201141H RNP201183H RNP201224H RNP201451H RNP201596H RNP201606H RNP201624H RNP201792H RNP201797H RNP202239H RNP202577H RNP202650H RNP202955H RNP203188H RNP203189H RNP203190H RNP203748H RNP204747H RNP204821H RNP205470H RNP205473H RNP200257H RNP201718H RNP200284H RNP200286H RNP200421H RNP200422H RNP200426H RNP200444H RNP200740H RNP200806H RNP200899H RNP201046H RNP201134H RNP201431H RNP201602H RNP201727H RNP201731H RNP201876H RNP202372H RNP202376H RNP202377H RNP202378H RNP202388H RNP202390H RNP202428H RNP202917H RNP203025H RNP203201H RNP203247H RNP203543H RNP203645H RNP203882H RNP204244H RNP204271H RNP204392H RNP204788H RNP204857H RNP204860H RNP204864H RNP205619H RNP205706H RNP205820H RNP200404H RNP200579H RNP204266H RNP200099H RNP200102H RNP200190H RNP201253H RNP201312H RNP203896H RNP205659H RNP200200H RNP200344H RNP200557H RNP201173H RNP201600H RNP201709H RNP201998H RNP202548H RNP203432H RNP203438H RNP204002H RNP200198H RNP200597H RNP201235H RNP201239H RNP201247H RNP201250H RNP201256H RNP201264H RNP201265H RNP201280H RNP201290H RNP201292H RNP201860H RNP201862H RNP201996H RNP203806H RNP204029H RNP204031H RNP204032H RNP204033H RNP204123H RNP204740H RNP205316H RNP205317H RNP205318H RNP205319H RNP205323H RNP205324H RNP205325H RNP205326H RNP205328H RNP205332H RNP205333H RNP205335H RNP205339H RNP205340H RNP205354H RNP205379H RNP205383H RNP205384H RNP205390H RNP205391H RNP205395H RNP205396H RNP205403H RNP205405H RNP205537H RNP206160H RNP203931H RNP200890H RNP200891H RNP201189H RNP203970H RNP205701H RNP200701H RNP201317H RNP202138H RNP202381H RNP203034H RNP203054H RNP203890H RNP203932H RNP203971H RNP204792H RNP202839H RNP202840H RNP202958H RNP200814H RNP201127H RNP201228H RNP201474H RNP203532H RNP203652H RNP203653H RNP200226H RNP200696H RNP200763H RNP201364H RNP204394H RNP200505H RNP200470H RNP200486H RNP200499H RNP201095H RNP201098H RNP201100H RNP201101H RNP201102H RNP201124H RNP201377H RNP201921H RNP202166H RNP202170H RNP202930H RNP202932H RNP203808H RNP203814H RNP205837H RNP202778H RNP203960H RNP204290H RNP200114H RNP201346H RNP201371H RNP201476H RNP200279H RNP201643H RNP202692H RNP202748H RNP203151H RNP203810H RNP204815H RNP200266H RNP200801H RNP201041H RNP202051H RNP203724H RNP204352H RNP204908H RNP205692H RNP205696H RNP200655H RNP201835H RNP201816H RNP200312H RNP201723H RNP200603H RNP201533H RNP201782H RNP202182H RNP202191H RNP203170H RNP203171H RNP203775H RNP204193H RNP204310H RNP204378H RNP204877H RNP200140H RNP200585H RNP201501H RNP201523H RNP201524H RNP201537H RNP201694H RNP201830H RNP201890H RNP201904H RNP202080H RNP202209H RNP203156H RNP203180H RNP203207H RNP203249H RNP203322H RNP204131H RNP204185H RNP204194H RNP204404H RNP204480H RNP204481H RNP204483H RNP204528H RNP204567H RNP204570H RNP204897H RNP205368H RNP201724H RNP200322H RNP200323H RNP200473H RNP200586H RNP200817H RNP201053H RNP201182H RNP201267H RNP201570H RNP201603H RNP201607H RNP201648H RNP201655H RNP201678H RNP201680H RNP201685H RNP201692H RNP201758H RNP201784H RNP201808H RNP201850H RNP201907H RNP201923H RNP201924H RNP201985H RNP202062H RNP202104H RNP202105H RNP202173H RNP202343H RNP202389H RNP202432H RNP202464H RNP202537H RNP202720H RNP202741H RNP202750H RNP202908H RNP202910H RNP202943H RNP202954H RNP202974H RNP202992H RNP203269H RNP203280H RNP203298H RNP203541H RNP203552H RNP203802H RNP203805H RNP203835H RNP203863H RNP203918H RNP204078H RNP204196H RNP204232H RNP204235H RNP204281H RNP204282H RNP204451H RNP204523H RNP204622H RNP204726H RNP204786H RNP204900H RNP204920H RNP205678H RNP205791H RNP205905H RNP206002H RNP206043H RNP202382H RNP200415H RNP201275H RNP205449H RNP200437H RNP200542H RNP200776H RNP201184H RNP202607H RNP205047H RNP201301H RNP204771H RNP200393H RNP200456H RNP201149H RNP201937H RNP202542H RNP203512H RNP204013H RNP204016H RNP204634H RNP204822H RNP204841H RNP204859H RNP201349H RNP201796H RNP202642H RNP203917H RNP204109H RNP204575H RNP201511H RNP204828H RNP204601H RNP205749H RNP200245H RNP200805H RNP200807H RNP201031H RNP201032H RNP201585H RNP202174H RNP202177H RNP202320H RNP202768H RNP204750H RNP205039H RNP205799H RNP201271H RNP201867H RNP204221H RNP205964H RNP200070H RNP200101H RNP200348H RNP200405H RNP200407H RNP200675H RNP200720H RNP200892H RNP202053H RNP202055H RNP202056H RNP202057H RNP202058H RNP202061H RNP202063H RNP202065H RNP202066H RNP202068H RNP202070H RNP202071H RNP201129H RNP201169H RNP201647H RNP201697H RNP201729H RNP201752H RNP202168H RNP200388H RNP201195H RNP202669H RNP204871H RNP200153H RNP201025H RNP202663H RNP200605H RNP201541H RNP201592H RNP202422H RNP202423H RNP202425H RNP202618H RNP203712H RNP203716H RNP204107H RNP204713H RNP204967H RNP200659H RNP204715H RNP202999H RNP200439H RNP200441H RNP200479H RNP202677H RNP202688H RNP203213H RNP203219H RNP200079H RNP200080H RNP200391H RNP200408H RNP201332H RNP201360H RNP201686H RNP202736H RNP203836H RNP200458H RNP201403H RNP201404H RNP201405H RNP203483H RNP203484H RNP202468H RNP203346H RNP203354H RNP203357H RNP203559H RNP203691H RNP203692H RNP203693H RNP200588H RNP200622H RNP200674H RNP200737H RNP201279H RNP202718H RNP202779H RNP202928H RNP203867H RNP204220H RNP204542H RNP206136H RNP203644H RNP202162H RNP200555H RNP201001H RNP201198H RNP201390H RNP201799H RNP203022H RNP201510H RNP202360H RNP200306H RNP202153H RNP202573H RNP203548H RNP203878H RNP205684H RNP200727H RNP200813H RNP201118H RNP202513H RNP202520H RNP202609H RNP202739H RNP203767H RNP203768H RNP203964H RNP204456H RNP205826H RNP200576H RNP200577H RNP201579H RNP202103H RNP202345H RNP202413H RNP203587H RNP204431H RNP204802H RNP204851H RNP205006H RNP205817H RNP205849H RNP205868H RNP206076H RNP204836H RNP200138H RNP200580H RNP200636H RNP200721H RNP200839H RNP201158H RNP201857H RNP201915H RNP202593H RNP202993H RNP203373H RNP203569H RNP203595H RNP204188H RNP204347H RNP204490H RNP204536H RNP204546H RNP205009H RNP201056H RNP200508H RNP204890H RNP203690H RNP204071H RNP202961H RNP205669H RNP201483H RNP203561H RNP203934H RNP200938H RNP201742H RNP202244H RNP203938H RNP204969H RNP200930H RNP202401H RNP201329H RNP201359H RNP202271H RNP204126H RNP201629H RNP202937H RNP203029H RNP204720H RNP205058H RNP201303H RNP204299H RNP204631H RNP200053H RNP200201H RNP200320H RNP201319H RNP201348H RNP201512H RNP201587H RNP202112H RNP202226H RNP202657H RNP202658H RNP202762H RNP202766H RNP202767H RNP202848H RNP203518H RNP203726H RNP204380H RNP204552H RNP204683H RNP205846H RNP202557H RNP200216H RNP200715H RNP201373H RNP202547H RNP202919H RNP203012H RNP203796H RNP203876H RNP204223H RNP204253H RNP204980H RNP205046H RNP205203H RNP205209H RNP205350H RNP205607H RNP205610H RNP205797H RNP205800H RNP205880H RNP205883H RNP203235H RNP204017H RNP203335H RNP200160H RNP200321H RNP200589H RNP201130H RNP201430H RNP201459H RNP201717H RNP202686H RNP202784H RNP203194H RNP203751H RNP203975H RNP204441H RNP204595H RNP204686H RNP200082H RNP200159H RNP200175H RNP200657H RNP200923H RNP201011H RNP201012H RNP201030H RNP201343H RNP201662H RNP201719H RNP201849H RNP202420H RNP202536H RNP202538H RNP202924H RNP203743H RNP203959H RNP204050H RNP204228H RNP204568H RNP204975H RNP205342H RNP205343H RNP205345H RNP205346H RNP205347H RNP200389H RNP200545H RNP201156H RNP202242H RNP202744H RNP202752H RNP203555H RNP205131H RNP200383H RNP200095H RNP202810H RNP204898H RNP205720H RNP202691H RNP201773H RNP204837H RNP200254H RNP200998H RNP201410H RNP201690H RNP201846H RNP202367H RNP203191H RNP204066H RNP200252H RNP200507H RNP201064H RNP201838H RNP202089H RNP202171H RNP204020H RNP204179H RNP204443H RNP204596H RNP205054H RNP200062H RNP200064H RNP202774H RNP204176H RNP205190H RNP205194H RNP205198H RNP203656H RNP203657H RNP203679H RNP205004H RNP200333H RNP200335H RNP201449H RNP201450H RNP201452H RNP201453H RNP201520H RNP201627H RNP203884H RNP203511H RNP200048H RNP201893H RNP200732H RNP201137H RNP201522H RNP201549H RNP201563H RNP201578H RNP201994H RNP202044H RNP202075H RNP202633H RNP202668H RNP202982H RNP203261H RNP203331H RNP203334H RNP203341H RNP203355H RNP203460H RNP203521H RNP203864H RNP203872H RNP205356H RNP200491H RNP200905H RNP200906H RNP200907H RNP201010H RNP202454H RNP202820H RNP202822H RNP202831H RNP203495H RNP203520H RNP204307H RNP204572H RNP204705H RNP204787H RNP205729H RNP205731H RNP205745H RNP205845H RNP200115H RNP201376H RNP201612H RNP202048H RNP203284H RNP204055H RNP204081H RNP204130H RNP204241H RNP204531H RNP204545H RNP204738H RNP205113H RNP205114H RNP205116H RNP201925H RNP200371H RNP200188H RNP200210H RNP200217H RNP200246H RNP200480H RNP200621H RNP203304H RNP203309H RNP200914H RNP200915H RNP200916H RNP200917H RNP200324H RNP201008H RNP202026H RNP202188H RNP202874H RNP204061H RNP200248H RNP201116H RNP203279H RNP203883H RNP202132H RNP201133H RNP201583H RNP202374H RNP201099H RNP201605H RNP202511H RNP202707H RNP203575H RNP202561H RNP201050H RNP201551H RNP202267H RNP203125H RNP203906H RNP202449H RNP202470H RNP202493H RNP202649H RNP203038H RNP204693H RNP205003H RNP205007H RNP205672H RNP205942H RNP202250H RNP202492H RNP203009H RNP203028H RNP204356H RNP204749H RNP204861H RNP205786H RNP201395H RNP203677H RNP205571H RNP205572H RNP201412H RNP202883H RNP200894H RNP201259H RNP201458H RNP202639H RNP202640H RNP202808H RNP202851H RNP203453H RNP203478H RNP203572H RNP203800H RNP204096H RNP204918H RNP202817H RNP202984H RNP203580H RNP204065H RNP204306H RNP204506H RNP204610H RNP204796H RNP204798H RNP205866H RNP205879H RNP201509H RNP201562H RNP202219H RNP202392H RNP202755H RNP203452H RNP203941H RNP204113H RNP204599H RNP201933H RNP201934H RNP201935H RNP201938H RNP201939H RNP203946H RNP201429H RNP201804H RNP202504H RNP202764H RNP203492H RNP203609H RNP204180H RNP204300H RNP204302H RNP204703H RNP204742H RNP204758H RNP204892H RNP201230H RNP202554H RNP203825H RNP203854H RNP203948H RNP204624H RNP204756H RNP204760H RNP204872H RNP205752H RNP201284H RNP202830H RNP200119H RNP200747H RNP201394H RNP201906H RNP204499H RNP204577H RNP200029H RNP200030H RNP200199H RNP200396H RNP200608H RNP200749H RNP200966H RNP200967H RNP200968H RNP200969H RNP200970H RNP200971H RNP200972H RNP200974H RNP200975H RNP200976H RNP200977H RNP200978H RNP202517H RNP202809H RNP203197H RNP204491H RNP204493H RNP204611H RNP205620H RNP205621H RNP205622H RNP205623H RNP205624H RNP205626H RNP205630H RNP205631H RNP201413H RNP202459H RNP200232H RNP200261H RNP200318H RNP200469H RNP201160H RNP201167H RNP201703H RNP201932H RNP204027H RNP204114H RNP204777H RNP204803H RNP204913H RNP204914H RNP200168H RNP203372H RNP204987H RNP205764H RNP201138H RNP201853H RNP201110H RNP201111H RNP203427H RNP203428H RNP200109H RNP200147H RNP201538H RNP202918H RNP200112H RNP200452H RNP200453H RNP200799H RNP200926H RNP200994H RNP201330H RNP201391H RNP201793H RNP202154H RNP202385H RNP202638H RNP202759H RNP203929H RNP203958H RNP204088H RNP204574H RNP204785H RNP205037H RNP205126H RNP200689H RNP200746H RNP201894H RNP202387H RNP202623H RNP202787H RNP203466H RNP204101H RNP201051H RNP203940H RNP204125H RNP204521H RNP205824H RNP205859H RNP200032H RNP203238H RNP204659H RNP200374H RNP201351H RNP202412H RNP202558H RNP205419H RNP205482H RNP205518H RNP205520H RNP205530H RNP203935H RNP204187H RNP205732H RNP200764H RNP201479H RNP202137H RNP204389H RNP200189H RNP201806H RNP201896H RNP201898H RNP202136H RNP203036H RNP203458H RNP203744H RNP203776H RNP203780H RNP203782H RNP203791H RNP204540H RNP204593H RNP205689H RNP205690H RNP205855H RNP205944H RNP205637H RNP200237H RNP200238H RNP200241H RNP200910H RNP200911H RNP200912H RNP200913H RNP205769H RNP201294H RNP202107H RNP200629H RNP201178H RNP202576H RNP204333H RNP204496H RNP200565H RNP200050H RNP200177H RNP200401H RNP200687H RNP201986H RNP202394H RNP202396H RNP202711H RNP203255H RNP203508H RNP204701H RNP202210H RNP203777H RNP201527H RNP202380H RNP203833H RNP202306H RNP204287H RNP200502H RNP201649H RNP201760H RNP202622H RNP203533H RNP203686H RNP203812H RNP204062H RNP204211H RNP204410H RNP204453H RNP200369H RNP200762H RNP200903H RNP201045H RNP201392H RNP201414H RNP201543H RNP201616H RNP201817H RNP201903H RNP202645H RNP202693H RNP202962H RNP203502H RNP203516H RNP203760H RNP203905H RNP203954H RNP204312H RNP204334H RNP204346H RNP204501H RNP204795H RNP205252H RNP205867H RNP205951H RNP200025H RNP200074H RNP200124H RNP200815H RNP200816H RNP200818H RNP200819H RNP200821H RNP200822H RNP200823H RNP200824H RNP200825H RNP200826H RNP200827H RNP200828H RNP200829H RNP200830H RNP200831H RNP200832H RNP200833H RNP200834H RNP200835H RNP200836H RNP202281H RNP203421H RNP203423H RNP203424H RNP203425H RNP203429H RNP203430H RNP203433H RNP203437H RNP203440H RNP203442H RNP203443H RNP203444H RNP203449H RNP203461H RNP203462H RNP204856H RNP205608H RNP205723H RNP205724H RNP205725H RNP205755H RNP205756H RNP205848H RNP205984H RNP205986H RNP206120H RNP206121H RNP206139H RNP200753H RNP200960H RNP200961H RNP200962H RNP200963H RNP200964H RNP200209H RNP201027H RNP201471H RNP201705H RNP202491H RNP202595H RNP204342H RNP201058H RNP203626H RNP205528H RNP205531H RNP202001H RNP202866H RNP204724H RNP204819H RNP205050H RNP205371H RNP203026H RNP201344H RNP200158H RNP200174H RNP200334H RNP200543H RNP200681H RNP201155H RNP201485H RNP201487H RNP201488H RNP201489H RNP201490H RNP202417H RNP202726H RNP202728H RNP202729H RNP202730H RNP202731H RNP203116H RNP203117H RNP203118H RNP203119H RNP203967H RNP203990H RNP204069H RNP204408H RNP204439H RNP204440H RNP204517H RNP204990H RNP205687H RNP205694H RNP205695H RNP205712H RNP205819H RNP205864H RNP205953H RNP205988H RNP206072H RNP206077H RNP200620H RNP201014H RNP201015H RNP201480H RNP201481H RNP202734H RNP204064H RNP204514H RNP204527H RNP204555H RNP204632H RNP204633H RNP204797H RNP204884H RNP205378H RNP205380H RNP205382H RNP205386H RNP205387H RNP203592H RNP203698H RNP204014H RNP204881H RNP205574H RNP205577H RNP205579H RNP205581H RNP205583H RNP205585H RNP205586H RNP203762H RNP203978H RNP200353H RNP200354H RNP200416H RNP201315H RNP203721H RNP203729H RNP203734H RNP204284H RNP200105H RNP200107H RNP200229H RNP200377H RNP201121H RNP201472H RNP201475H RNP201486H RNP201495H RNP202567H RNP202689H RNP203137H RNP203388H RNP203392H RNP203400H RNP203401H RNP203406H RNP203417H RNP203422H RNP203431H RNP203514H RNP202838H RNP201411H RNP204028H RNP204100H RNP201532H RNP200789H RNP200121H RNP200463H RNP203639H RNP203642H RNP203648H RNP204642H RNP204643H RNP200220H RNP200466H RNP200820H RNP200837H RNP200852H RNP201096H RNP201097H RNP201288H RNP201291H RNP201293H RNP201334H RNP201336H RNP205493H RNP205494H RNP205497H RNP205498H RNP205503H RNP200809H RNP201573H RNP202415H RNP202458H RNP202646H RNP203058H RNP204059H RNP205673H RNP200155H RNP200267H RNP200798H RNP201222H RNP202152H RNP202222H RNP202794H RNP202949H RNP203530H RNP203933H RNP204984H RNP200056H RNP200058H RNP200243H RNP200549H RNP200571H RNP200883H RNP201540H RNP201559H RNP201564H RNP201641H RNP201681H RNP201778H RNP201805H RNP201834H RNP202176H RNP202526H RNP202527H RNP202829H RNP203000H RNP203551H RNP203742H RNP203915H RNP204044H RNP204240H RNP204868H RNP205067H RNP205841H RNP205887H RNP200283H RNP200424H RNP203947H RNP200669H RNP200670H RNP201477H RNP201657H RNP201754H RNP202546H RNP203818H RNP204119H RNP201744H RNP201745H RNP201790H RNP201807H RNP201823H RNP204812H RNP200525H RNP200493H RNP202427H RNP200207H RNP200760H RNP200928H RNP201880H RNP202671H RNP204008H RNP201927H RNP203965H RNP200270H RNP201615H RNP201722H RNP202586H RNP203527H RNP203985H RNP201324H RNP202487H RNP200410H RNP200712H RNP202476H RNP202710H RNP202931H RNP203482H RNP203709H RNP204121H RNP204813H RNP205920H RNP205938H RNP205957H RNP206059H RNP204406H RNP204698H RNP200645H RNP202805H RNP203254H RNP202801H RNP203140H RNP203925H RNP200234H RNP202509H RNP203773H RNP203923H RNP200194H RNP201019H RNP201020H RNP201237H RNP201468H RNP201505H RNP201741H RNP201825H RNP201841H RNP202005H RNP202317H RNP202327H RNP202481H RNP202921H RNP202946H RNP204118H RNP204254H RNP204457H RNP204895H RNP202248H RNP200264H RNP201494H RNP204423H RNP205652H RNP200442H RNP202555H RNP204538H RNP201482H RNP202570H RNP200148H RNP200637H RNP200840H RNP200893H RNP201508H RNP203237H RNP205011H RNP205697H RNP204382H RNP201036H RNP201067H RNP202518H RNP203607H RNP204648H RNP205853H RNP200326H RNP200328H RNP200329H RNP200330H RNP200332H RNP200615H RNP200625H RNP203910H RNP203911H RNP203913H RNP202262H RNP202880H RNP204412H RNP204413H RNP202556H RNP202998H RNP203701H RNP203703H RNP203993H RNP204504H RNP204608H RNP204725H RNP205618H RNP205918H RNP200187H RNP201174H RNP201889H RNP202224H RNP202690H RNP202994H RNP204755H RNP201941H RNP200137H RNP200139H RNP201192H RNP201193H RNP201384H RNP202041H RNP202379H RNP202539H RNP204513H RNP204820H RNP204887H RNP205016H RNP200172H RNP200617H RNP203995H RNP204251H RNP200085H RNP200087H RNP201504H RNP203166H RNP204636H RNP202804H RNP203360H RNP203363H RNP200632H RNP200060H RNP202167H RNP202951H RNP202952H RNP203622H RNP203956H RNP204519H RNP205057H RNP202233H RNP202344H RNP203136H RNP200256H RNP205803H RNP200164H RNP201164H RNP202243H RNP203755H RNP204292H RNP205674H RNP201114H RNP202072H RNP202666H RNP204525H RNP200375H RNP201665H RNP201666H RNP201667H RNP201669H RNP201674H RNP203764H RNP204012H RNP204409H RNP200705H RNP201227H RNP201229H RNP205529H RNP202098H RNP202114H RNP202116H RNP205955H RNP202655H RNP204776H RNP204789H RNP204801H RNP204804H RNP204805H RNP205907H RNP205463H RNP205465H RNP205471H RNP205906H RNP205908H RNP205909H RNP205910H RNP205911H RNP205912H RNP205913H RNP205914H RNP203885H RNP206163H RNP200804H RNP201397H RNP201398H RNP201399H RNP201400H RNP201401H RNP201402H RNP201628H RNP201777H RNP201989H RNP202087H RNP202147H RNP202641H RNP203129H RNP203130H RNP203131H RNP205506H RNP200313H RNP200884H RNP200885H RNP200886H RNP200887H RNP200888H RNP201576H RNP202000H RNP202009H RNP202011H RNP202013H RNP202014H RNP202016H RNP202023H RNP202024H RNP202328H RNP202330H RNP202332H RNP202333H RNP202335H RNP202336H RNP202337H RNP202340H RNP202341H RNP202342H RNP204981H RNP204982H RNP204983H RNP205475H RNP206066H RNP206078H RNP206095H RNP206098H RNP206099H RNP206100H RNP206103H RNP206104H RNP206105H RNP206106H RNP206107H RNP206109H RNP202789H RNP202891H RNP202894H RNP202895H RNP203042H RNP203043H RNP203044H RNP203262H RNP203265H RNP203268H RNP203273H RNP203274H RNP203275H RNP203292H RNP203293H RNP203294H RNP203295H RNP203296H RNP203297H RNP203300H RNP203301H RNP203302H RNP203303H RNP203306H RNP203307H RNP203308H RNP203310H RNP203311H RNP203312H RNP203313H RNP203315H RNP203317H RNP203321H RNP204661H RNP204664H RNP204666H RNP204667H RNP204668H RNP204669H RNP204670H RNP204672H RNP204673H RNP204844H RNP204845H RNP204847H RNP204848H RNP204849H RNP205133H RNP205134H RNP205135H RNP205136H RNP205137H RNP205138H RNP205139H RNP205142H RNP205143H RNP205144H RNP205146H RNP205147H RNP205150H RNP205151H RNP205155H RNP205156H RNP205157H RNP205161H RNP205162H RNP205163H RNP205164H RNP205165H RNP205168H RNP205169H RNP205171H RNP205172H RNP205173H RNP205174H RNP205177H RNP205178H RNP205179H RNP205182H RNP205183H RNP205184H RNP205186H RNP205187H RNP205189H RNP205191H RNP205192H RNP205195H RNP205196H RNP205197H RNP205204H RNP205205H RNP205206H RNP205207H RNP205212H RNP205216H RNP205217H RNP205218H RNP205224H RNP205228H RNP205229H RNP205233H RNP205235H RNP205236H RNP205239H RNP205241H RNP205242H RNP205243H RNP205244H RNP205245H RNP205248H RNP205249H RNP205251H RNP205255H RNP205256H RNP205257H RNP205262H RNP205264H RNP205266H RNP205268H RNP205269H RNP205270H RNP205271H RNP205272H RNP205278H RNP205279H RNP205281H RNP205289H RNP205291H RNP205294H RNP205297H RNP205349H RNP205351H RNP205353H RNP205355H RNP200641H RNP206170H RNP206173H RNP206174H RNP206182H RNP206183H RNP206185H RNP206189H RNP206201H RNP205042H RNP205456H RNP205671H RNP206115H RNP203950H RNP203989H RNP203997H RNP206135H RNP203803H RNP205362H RNP202735H RNP205881H RNP202477H RNP202673H RNP203813H RNP202544H RNP206044H RNP206081H RNP206156H RNP206175H RNP206194H RNP206619H RNP206650H RNP206686H RNP206713H RNP206728H RNP206764H RNP206572H RNP206594H RNP206600H RNP206603H RNP206608H RNP206609H RNP206662H RNP206666H RNP206667H RNP206670H RNP206672H RNP206677H RNP206683H RNP206714H RNP204474H RNP201658H RNP200496H RNP201308H RNP203045H RNP201613H RNP203739H RNP203838H RNP202569H RNP204549H RNP201436H RNP201726H RNP203631H RNP201147H RNP201201H RNP201650H RNP201652H RNP201780H RNP201843H RNP201844H RNP201847H RNP201856H RNP201863H RNP202082H RNP202084H RNP202847H RNP203283H RNP203445H RNP203745H RNP203804H RNP204379H RNP204407H RNP204462H RNP204768H RNP205753H RNP205821H RNP200446H RNP200447H RNP201007H RNP201123H RNP201318H RNP201466H RNP201492H RNP201536H RNP201591H RNP202117H RNP202119H RNP202122H RNP202129H RNP202279H RNP204022H RNP204906H RNP205587H RNP205588H RNP205589H RNP205591H RNP205592H RNP205594H RNP204677H RNP203874H RNP204301H RNP201772H RNP200222H RNP200501H RNP201503H RNP202073H RNP204921H RNP205199H RNP205369H RNP205813H RNP205830H RNP200162H RNP200236H RNP200271H RNP200277H RNP200278H RNP200395H RNP200398H RNP200490H RNP201216H RNP201464H RNP201584H RNP201869H RNP203901H RNP204042H RNP204809H RNP204852H RNP200250H RNP200317H RNP201117H RNP201203H RNP204091H RNP200231H RNP200253H RNP200272H RNP201128H RNP201350H RNP202054H RNP202060H RNP202964H RNP203811H RNP203820H RNP203856H RNP204040H RNP204357H RNP204578H RNP204784H RNP204998H RNP205773H RNP200713H RNP205726H RNP204612H RNP203122H RNP200136H RNP200423H RNP200427H RNP201154H RNP201802H RNP202479H RNP205478H RNP205693H RNP205810H RNP200425H RNP202039H RNP203172H RNP204286H RNP201993H RNP203001H RNP203434H RNP203966H RNP205875H RNP200376H RNP201454H RNP202289H RNP203202H RNP200219H RNP200364H RNP200366H RNP200572H RNP201617H RNP201626H RNP201635H RNP201714H RNP200703H RNP201172H RNP201530H RNP201656H RNP202386H RNP204076H RNP204554H RNP204923H RNP202733H RNP205648H RNP205666H RNP205668H RNP205121H RNP205811H RNP200959H RNP201638H RNP204484H RNP201161H RNP201162H RNP202978H RNP201244H RNP202972H RNP202953H RNP203513H RNP205873H RNP205721H RNP205727H RNP204806H RNP201526H RNP201895H RNP204996H RNP204403H RNP205683H RNP204781H RNP205130H RNP202800H RNP205854H RNP203474H RNP201226H RNP202562H RNP203996H RNP202234H RNP202277H RNP203134H RNP205758H RNP201352H RNP201633H RNP202837H RNP204102H RNP204551H RNP204905H RNP206154H RNP204039H RNP204603H RNP204609H RNP204834H RNP205747H RNP204687H RNP204970H RNP200090H RNP200093H RNP200180H RNP200182H RNP200476H RNP202988H RNP203571H RNP203579H RNP203581H RNP203582H RNP204507H RNP204752H RNP205149H RNP205153H RNP205154H RNP205158H RNP205159H RNP205166H RNP205167H RNP205170H RNP203841H RNP200089H RNP205175H RNP204426H RNP205916H RNP204626H RNP201049H RNP201159H RNP204731H RNP202140H RNP204658H RNP200627H RNP201875H RNP202480H RNP200537H RNP201028H RNP201534H RNP203736H RNP200169H RNP200514H RNP200706H RNP201000H RNP202028H RNP202030H RNP202583H RNP203436H RNP205309H RNP205310H RNP205313H RNP205314H RNP200478H RNP201991H RNP202660H RNP202790H RNP203702H RNP203879H RNP204108H RNP204120H RNP204335H RNP205603H RNP205787H RNP205788H RNP205789H RNP204189H RNP205336H RNP205337H RNP205338H RNP205341H RNP201353H RNP204233H RNP200359H RNP200360H RNP200394H RNP200918H RNP200919H RNP202083H RNP202086H RNP202261H RNP200035H RNP200494H RNP201710H RNP201801H RNP205972H RNP205975H RNP205977H RNP205978H RNP205980H RNP205850H RNP204393H RNP200215H RNP201249H RNP201568H RNP202627H RNP202745H RNP203164H RNP203662H RNP204824H RNP202864H RNP200111H RNP200183H RNP200482H RNP200602H RNP202802H RNP202942H RNP203260H RNP203499H RNP203500H RNP203501H RNP203503H RNP204401H RNP204473H RNP205822H RNP206079H RNP201054H RNP204446H RNP205857H RNP205919H RNP202973H RNP204432H RNP201300H RNP205966H RNP205967H RNP203849H RNP204759H RNP205818H RNP204229H RNP204482H RNP205793H RNP202979H RNP204586H RNP200208H RNP203586H RNP205798H RNP203678H RNP205716H RNP203944H RNP205936H RNP203574H RNP205570H RNP200432H RNP204835H RNP202793H RNP203519H RNP205000H RNP203815H RNP201044H RNP204058H RNP202550H RNP203590H RNP202135H RNP204361H RNP204590H RNP204124H RNP205926H RNP205928H RNP205929H RNP204106H RNP202765H RNP204745H RNP201022H RNP201771H RNP202442H RNP204341H RNP204995H RNP201407H RNP202717H RNP205015H RNP203203H RNP203200H RNP203412H RNP202027H RNP202615H RNP202740H RNP203410H RNP204203H RNP204355H RNP204618H RNP204623H RNP201210H RNP202724H RNP203600H RNP203865H RNP204435H RNP202799H RNP203057H RNP201191H RNP201473H RNP201789H RNP202713H RNP200097H RNP204823H RNP206145H RNP206146H RNP206147H RNP204972H RNP200291H RNP200483H RNP201179H RNP201589H RNP201982H RNP201983H RNP201987H RNP202033H RNP202035H RNP202036H RNP202037H RNP202038H RNP200623H RNP201213H RNP200059H RNP200126H RNP200127H RNP200179H RNP200185H RNP200224H RNP200227H RNP200239H RNP200296H RNP200400H RNP200443H RNP200450H RNP200451H RNP200454H RNP200550H RNP200553H RNP200672H RNP200700H RNP200792H RNP200793H RNP200794H RNP200795H RNP201145H RNP201166H RNP201672H RNP201735H RNP201854H RNP201855H RNP202008H RNP202019H RNP202067H RNP202300H RNP202461H RNP202478H RNP202581H RNP202629H RNP202630H RNP202631H RNP202644H RNP202674H RNP202695H RNP202700H RNP202854H RNP202875H RNP202893H RNP202920H RNP202939H RNP202965H RNP203041H RNP203049H RNP203050H RNP203053H RNP203072H RNP203084H RNP203096H RNP203100H RNP203105H RNP203128H RNP204023H RNP204245H RNP204270H RNP204305H RNP204430H RNP204452H RNP204550H RNP204553H RNP204556H RNP204558H RNP204562H RNP204565H RNP204566H RNP204569H RNP204571H RNP204573H RNP205213H RNP205215H RNP205220H RNP205221H RNP205222H RNP205223H RNP200294H RNP200372H RNP200535H RNP202469H RNP202696H RNP202715H RNP202935H RNP203889H RNP204544H RNP204585H RNP204587H RNP204829H RNP205776H RNP204115H RNP200692H RNP200929H RNP201338H RNP201340H RNP201440H RNP201529H RNP204639H RNP205201H RNP205941H RNP204765H RNP204766H RNP204767H RNP201236H RNP201314H RNP201542H RNP202460H RNP202727H RNP202754H RNP203223H RNP203404H RNP204638H RNP205987H RNP201995H RNP201023H RNP200699H RNP201575H RNP202149H RNP202150H RNP202151H RNP202869H RNP203206H RNP203208H RNP203209H RNP205645H RNP205646H RNP204225H RNP204539H RNP202651H RNP202654H RNP201255H RNP201194H RNP205181H RNP202495H RNP204445H RNP203454H RNP201261H RNP201263H RNP201456H RNP201548H RNP201762H RNP201905H RNP201931H RNP202046H RNP202049H RNP202050H RNP203242H RNP203244H RNP203246H RNP203248H RNP203285H RNP203288H RNP205632H RNP205633H RNP205805H RNP206005H RNP201052H RNP201268H RNP201386H RNP202160H RNP202543H RNP202845H RNP202896H RNP202899H RNP203951H RNP203982H RNP204313H RNP204723H RNP204793H RNP204883H RNP205647H RNP205790H RNP201751H RNP203004H RNP203480H RNP203694H RNP204272H RNP205782H RNP204449H RNP204839H RNP203459H RNP202515H RNP203903H RNP205111H RNP205650H RNP204202H RNP203704H RNP200564H RNP204691H RNP204559H RNP200173H RNP202273H RNP200575H RNP200047H RNP201139H RNP201177H RNP203120H RNP203471H RNP203783H RNP203898H RNP203972H RNP204991H RNP201355H RNP201569H RNP202661H RNP204579H RNP200467H RNP201595H RNP202815H RNP202947H RNP203792H RNP205865H RNP200643H RNP200644H RNP200646H RNP200652H RNP200281H RNP200282H RNP201644H RNP202258H RNP202683H RNP202933H RNP203554H RNP203633H RNP205770H RNP205774H RNP200100H RNP200647H RNP200666H RNP201737H RNP204989H RNP204084H RNP200259H RNP200357H RNP200409H RNP200418H RNP200434H RNP200583H RNP200601H RNP200743H RNP200745H RNP201262H RNP201283H RNP201302H RNP201645H RNP201732H RNP201786H RNP201800H RNP201826H RNP201929H RNP201930H RNP202029H RNP202043H RNP202078H RNP202081H RNP202094H RNP202553H RNP202679H RNP202936H RNP203040H RNP203488H RNP203562H RNP203754H RNP203893H RNP203936H RNP203937H RNP203945H RNP204206H RNP204652H RNP204826H RNP205118H RNP205374H RNP205376H RNP205816H RNP205969H RNP205973H RNP200123H RNP200642H RNP200997H RNP201059H RNP201434H RNP201571H RNP202231H RNP202290H RNP202292H RNP202293H RNP202294H RNP202295H RNP202297H RNP202299H RNP202301H RNP202309H RNP202310H RNP202681H RNP203387H RNP203393H RNP203397H RNP203402H RNP201491H RNP200633H RNP202370H RNP203211H RNP203650H RNP200704H RNP200707H RNP200708H RNP205808H RNP201026H RNP201426H RNP202184H RNP202606H RNP202647H RNP203537H RNP203685H RNP204105H RNP205595H RNP200116H RNP200569H RNP201555H RNP201598H RNP201601H RNP200790H RNP200631H RNP201033H RNP201531H RNP200609H RNP202648H RNP203568H RNP205767H RNP205768H RNP200584H RNP202253H RNP202782H RNP201363H RNP202656H RNP203039H RNP203266H RNP203510H RNP204600H RNP201747H RNP200729H RNP203133H RNP204607H RNP205443H RNP200228H RNP200567H RNP201146H RNP201417H RNP202045H RNP202190H RNP202192H RNP202451H RNP204366H RNP204581H RNP200150H RNP200428H RNP200430H RNP200538H RNP200539H RNP201442H RNP201470H RNP201640H RNP202354H RNP202523H RNP202560H RNP202632H RNP202781H RNP202803H RNP203019H RNP203663H RNP203711H RNP204433H RNP204655H RNP204656H RNP204660H RNP204662H RNP204665H RNP204671H RNP204674H RNP204700H RNP204716H RNP204718H RNP203772H RNP203924H RNP205063H RNP202788H RNP200406H RNP202575H RNP202916H RNP201749H RNP203146H RNP205505H RNP205835H RNP201661H RNP201753H RNP202773H RNP203823H RNP204377H RNP204414H RNP203456H RNP200660H RNP202608H RNP202926H RNP203961H RNP204842H RNP200766H RNP200767H RNP200768H RNP200769H RNP200770H RNP204454H RNP204455H RNP204460H RNP204467H RNP204470H RNP204036H RNP202473H RNP201209H RNP204582H RNP204927H RNP200331H RNP200465H RNP201152H RNP201435H RNP202245H RNP202346H RNP202348H RNP202349H RNP202350H RNP203138H RNP203426H RNP203779H RNP204365H RNP204934H RNP205643H RNP201387H RNP202092H RNP202540H RNP204336H RNP205894H RNP205895H RNP205960H RNP203290H RNP200639H RNP200985H RNP200988H RNP200989H RNP200991H RNP204704H RNP205611H RNP205612H RNP205614H RNP205615H RNP205617H RNP200098H RNP200193H RNP200249H RNP200468H RNP200898H RNP201281H RNP201306H RNP201608H RNP201763H RNP201764H RNP201842H RNP202482H RNP202506H RNP203718H RNP204129H RNP204340H RNP204893H RNP205070H RNP205110H RNP205119H RNP200204H RNP200419H RNP200640H RNP201004H RNP201273H RNP201446H RNP201496H RNP201840H RNP201891H RNP202218H RNP202225H RNP202237H RNP202238H RNP202356H RNP202587H RNP202597H RNP203006H RNP203451H RNP203457H RNP203522H RNP203627H RNP204353H RNP202249H RNP206153H RNP204273H RNP200275H RNP204425H RNP205246H RNP205247H RNP205250H RNP205254H RNP205258H RNP205259H RNP205260H RNP205261H RNP205545H RNP205549H RNP205550H RNP205553H RNP205556H RNP205562H RNP205567H RNP205568H RNP205578H RNP205580H RNP205584H RNP205590H RNP200316H RNP201180H RNP201181H RNP201369H RNP202590H RNP204000H RNP203267H RNP203682H RNP203683H RNP205739H RNP205956H RNP206123H RNP205750H RNP200146H RNP200233H RNP201199H RNP201622H RNP201625H RNP201879H RNP202195H RNP202472H RNP202835H RNP203549H RNP203727H RNP203795H RNP204242H RNP204535H RNP204730H RNP204732H RNP204774H RNP205682H RNP205949H RNP203032H RNP200431H RNP200638H RNP200725H RNP200757H RNP200980H RNP200981H RNP200982H RNP200983H RNP200984H RNP201168H RNP201217H RNP201673H RNP202085H RNP203173H RNP203174H RNP203175H RNP203177H RNP203178H RNP203179H RNP203182H RNP203186H RNP203187H RNP204886H RNP205140H RNP205141H RNP205596H RNP205598H RNP205599H RNP205600H RNP205601H RNP202486H RNP203983H RNP205265H RNP205273H RNP205274H RNP205275H RNP205280H RNP205283H RNP205290H RNP205292H RNP205293H RNP205296H RNP205301H RNP205308H RNP205311H RNP205312H RNP205315H RNP205538H RNP205539H RNP205542H RNP205543H RNP205544H RNP205526H RNP205660H RNP205661H RNP205662H RNP205663H RNP203614H RNP204503H RNP205366H RNP205367H RNP205370H RNP205372H RNP205373H RNP205375H RNP205377H RNP205300H RNP205302H RNP205303H RNP205304H RNP205305H RNP205306H RNP205606H RNP203761H RNP203981H RNP204199H RNP205074H RNP205507H RNP205509H RNP205517H RNP205519H RNP205521H RNP205522H RNP205523H RNP205524H RNP205527H RNP205532H RNP205533H RNP205535H RNP205536H RNP205546H RNP205547H RNP205551H RNP205552H RNP205555H RNP205563H RNP200276H RNP200509H RNP200515H RNP201016H RNP201507H RNP202749H RNP203591H RNP200134H RNP200135H RNP203051H RNP202196H RNP203707H RNP204383H RNP203750H RNP203124H RNP204082H RNP200205H RNP200581H RNP201574H RNP201750H RNP201785H RNP201916H RNP202211H RNP202436H RNP202437H RNP202438H RNP203689H RNP204976H RNP201582H RNP200066H RNP200504H RNP201185H RNP202636H RNP202889H RNP202906H RNP203222H RNP203881H RNP203994H RNP204005H RNP204080H RNP202665H RNP200772H RNP200773H RNP200774H RNP201424H RNP202598H RNP203016H RNP203673H RNP205634H RNP201388H RNP201393H RNP202093H RNP202096H RNP202600H RNP201215H RNP203681H RNP204278H RNP205060H RNP201668H RNP200530H RNP201671H RNP204764H RNP201654H RNP201679H RNP200677H RNP200684H RNP201912H RNP203234H RNP202079H RNP202232H RNP201561H RNP201711H RNP200363H RNP200435H RNP200520H RNP201187H RNP201311H RNP201699H RNP201706H RNP202180H RNP202528H RNP202624H RNP202737H RNP202862H RNP203455H RNP203489H RNP205801H RNP200448H RNP201765H RNP202213H RNP203659H RNP200063H RNP201062H RNP203720H RNP203822H RNP204127H RNP200244H RNP200170H RNP200436H RNP202530H RNP202676H RNP202678H RNP204094H RNP200242H RNP200414H RNP201379H RNP202319H RNP202322H RNP200352H RNP200534H RNP202732H RNP205950H RNP200548H RNP200610H RNP200730H RNP202507H RNP203730H RNP203731H RNP201809H RNP200574H RNP203556H RNP201630H RNP201651H RNP206063H RNP205952H RNP201988H RNP203770H RNP202704H RNP204602H RNP201908H RNP202228H RNP206080H RNP204070H RNP204537H RNP204830H RNP204850H RNP203557H RNP202006H RNP203544H RNP204057H RNP202505H RNP203848H RNP205970H RNP203539H RNP205657H RNP203735H RNP205604H RNP205992H RNP205945H RNP205685H RNP202791H RNP203870H RNP203887H RNP200358H RNP204688H RNP202212H RNP204529H RNP205001H RNP201374H RNP205090H RNP203204H RNP203498H RNP204854H RNP204511H RNP204234H RNP203333H RNP203871H RNP203210H RNP205474H RNP205193H RNP204183H RNP205394H RNP204063H RNP204783H RNP203640H RNP204882H RNP205933H RNP205946H RNP205176H RNP205871H RNP206108H RNP200385H RNP202017H RNP202018H RNP202612H RNP202966H RNP204371H RNP204763H RNP205180H RNP205700H RNP204778H RNP204775H RNP201599H RNP202494H RNP203486H RNP205655H RNP205675H RNP200240H RNP202074H RNP202076H RNP205188H RNP205814H RNP204876H RNP204912H RNP201057H RNP201175H RNP202821H RNP202980H RNP201205H RNP205927H RNP200176H RNP204209H RNP205991H RNP201142H RNP202944H RNP203942H RNP204262H RNP202589H RNP204192H RNP205344H RNP200161H RNP200449H RNP200521H RNP200800H RNP201176H RNP201535H RNP201553H RNP201565H RNP202406H RNP203411H RNP203494H RNP205765H RNP205958H RNP200598H RNP201354H RNP203756H RNP204204H RNP203257H RNP202334H RNP202552H RNP204598H RNP204604H RNP204037H RNP204524H RNP203606H RNP205840H RNP204291H RNP204308H RNP202467H RNP203705H RNP205658H RNP203788H RNP202516H RNP204873H RNP201396H RNP204931H RNP205876H RNP202628H RNP200506H RNP205676H RNP200339H RNP204354H RNP203589H RNP205010H RNP200513H RNP201013H RNP204930H RNP201926H RNP203214H RNP203216H RNP203217H RNP203220H RNP203225H RNP203226H RNP203230H RNP203231H RNP203324H RNP203327H RNP203328H RNP203329H RNP203330H RNP203546H RNP203563H RNP205359H RNP205360H RNP205361H RNP205363H RNP205364H RNP205934H RNP201427H RNP203033H RNP203047H RNP204398H RNP204399H RNP203629H RNP204761H RNP203199H RNP202922H RNP204437H RNP205932H RNP204067H RNP204068H RNP205393H RNP204486H RNP205744H RNP204744H RNP204649H RNP204122H RNP203566H RNP202780H RNP204257H RNP204388H RNP206140H RNP202163H RNP203024H RNP202280H RNP200141H RNP200156H RNP200347H RNP201126H RNP201331H RNP201347H RNP201677H RNP201693H RNP201779H RNP201873H RNP202490H RNP202498H RNP202714H RNP202878H RNP202897H RNP202901H RNP202902H RNP204205H RNP204207H RNP204208H RNP204210H RNP204212H RNP204214H RNP204216H RNP204217H RNP204218H RNP204219H RNP204476H RNP205735H RNP204375H RNP205742H RNP205971H RNP205407H RNP205128H RNP205145H RNP205406H RNP203787H RNP201597H RNP203671H RNP204807H RNP205852H RNP204309H RNP201135H RNP201803H RNP205605H RNP201858H RNP202146H RNP202761H RNP202887H RNP202941H RNP203598H RNP201833H RNP202230H RNP202807H RNP201518H RNP202125H RNP202445H RNP203339H RNP203576H RNP201619H RNP201112H RNP201707H RNP202914H RNP203278H RNP205200H RNP203252H RNP203732H RNP203979H RNP205670H RNP204853H RNP205018H RNP202229H RNP204009H RNP205771H RNP204093H RNP205833H RNP202927H RNP203880H RNP202888H RNP201055H RNP201341H RNP202836H RNP203758H RNP202858H RNP203031H RNP204396H RNP204641H RNP204899H RNP204992H RNP205656H RNP202131H RNP202141H RNP205718H RNP200128H RNP202967H RNP203655H RNP203658H RNP203660H RNP203661H RNP200130H RNP201748H RNP201922H RNP203666H RNP200777H RNP201140H RNP202172H RNP202602H RNP202886H RNP203858H RNP204418H RNP205925H RNP204265H RNP200544H RNP200547H RNP200904H RNP200999H RNP201828H RNP201829H RNP202221H RNP202823H RNP202825H RNP203504H RNP203973H RNP205703H RNP205748H RNP205827H RNP205983H RNP205979H RNP205629H RNP201794H RNP205785H RNP204543H RNP201550H RNP201554H RNP203801H RNP200932H RNP200934H RNP200935H RNP200936H RNP200937H RNP203832H RNP203834H RNP200304H RNP200305H RNP201221H RNP201254H RNP201257H RNP201266H RNP201270H RNP203160H RNP203161H RNP203162H RNP203163H RNP203165H RNP204717H RNP203930H RNP202240H RNP204263H RNP202077H RNP200724H RNP202139H RNP202471H RNP202564H RNP205447H RNP201728H RNP203192H RNP203229H RNP202797H RNP200690H RNP200022H RNP200280H RNP200498H RNP200526H RNP200527H RNP200611H RNP200612H RNP200618H RNP200619H RNP201171H RNP201586H RNP201670H RNP201708H RNP201720H RNP201734H RNP201736H RNP201740H RNP201767H RNP201787H RNP201861H RNP201865H RNP201902H RNP202003H RNP202007H RNP202020H RNP202097H RNP202100H RNP202549H RNP202551H RNP202584H RNP202601H RNP202701H RNP202702H RNP203241H RNP203272H RNP203291H RNP204252H RNP204289H RNP204580H RNP204592H RNP204727H RNP204733H RNP205185H RNP205954H RNP202183H RNP201827H RNP204280H RNP204935H RNP201282H RNP201304H RNP201362H RNP201513H RNP201514H RNP201515H RNP201588H RNP201590H RNP201824H RNP201999H RNP202834H RNP202857H RNP202945H RNP203215H RNP203277H RNP203585H RNP203617H RNP203618H RNP203620H RNP203624H RNP203625H RNP203630H RNP203635H RNP203636H RNP203637H RNP205231H RNP205232H RNP205234H RNP205237H RNP205238H RNP205593H RNP205597H RNP205602H RNP205609H RNP205616H RNP205834H RNP205013H RNP202912H RNP204083H RNP203289H RNP200290H RNP205642H RNP203148H RNP204006H RNP202996H RNP203490H RNP203534H RNP204226H RNP204534H RNP200196H RNP203323H RNP204198H RNP204629H RNP202522H RNP202499H RNP202991H RNP205040H RNP203939H RNP205947H RNP203597H RNP204030H RNP200255H RNP200314H RNP201072H RNP201851H RNP201901H RNP201936H RNP202500H RNP202502H RNP202751H RNP202826H RNP202963H RNP203181H RNP203314H RNP203319H RNP203819H RNP204541H RNP205467H RNP205794H RNP204034H RNP202545H RNP203015H RNP200460H RNP201188H RNP202311H RNP202312H RNP202314H RNP202316H RNP202325H RNP205398H RNP200487H RNP201761H RNP202742H RNP203809H RNP205476H RNP204635H RNP202976H RNP202721H RNP204516H RNP204303H RNP204620H RNP205882H RNP205959H RNP201245H RNP204614H RNP205127H RNP205253H RNP204387H RNP204644H RNP200092H RNP202111H RNP201460H RNP201837H RNP203067H RNP205348H RNP200166H RNP200214H RNP201378H RNP201383H RNP201415H RNP201420H RNP201422H RNP201547H RNP200381H RNP200462H RNP202266H RNP202703H RNP203013H RNP203713H RNP203715H RNP200065H RNP200939H RNP200940H RNP200941H RNP200942H RNP200943H RNP200944H RNP200945H RNP200946H RNP200947H RNP200948H RNP200949H RNP200950H RNP200951H RNP200952H RNP200953H RNP200954H RNP200955H RNP200956H RNP200957H RNP200958H RNP204909H RNP205048H RNP205052H RNP205055H RNP205059H RNP205061H RNP205062H RNP205066H RNP205068H RNP205424H RNP205425H RNP205428H RNP205430H RNP205431H RNP205432H RNP205433H RNP205434H RNP205438H RNP205441H RNP205442H RNP205444H RNP205446H RNP205450H RNP205453H RNP205462H RNP200273H RNP201385H RNP201428H RNP201433H RNP202259H RNP202351H RNP202352H RNP203418H RNP203419H RNP203420H RNP205410H RNP205413H RNP205417H RNP200908H RNP200909H RNP202263H RNP202264H RNP205496H RNP201682H RNP200594H RNP200061H RNP206414H RNP206425H RNP206427H RNP206428H RNP206437H RNP206438H RNP206440H RNP206446H RNP206449H RNP206451H RNP206454H RNP206457H RNP206458H RNP206471H RNP206474H RNP206476H RNP206480H RNP206484H RNP206492H RNP206500H RNP206519H RNP206520H RNP206540H RNP206549H RNP206558H RNP206562H RNP206564H RNP206571H RNP206696H RNP206701H RNP206768H RNP201408H RNP202042H RNP203408H RNP203491H RNP204505H RNP204986H RNP205397H RNP205654H RNP205792H RNP205802H RNP206001H (b) 00643169891418 Serial Numbers: RNP602176S RNP602177S RNP602178S RNP602179S RNP602180S RNP602181S RNP602182S RNP601674S RNP601910S RNP601996S RNP602054S RNP602132S RNP601552S RNP601804S RNP602034S RNP601862S RNP601969S RNP601812S RNP601917S RNP602038S RNP601793S RNP601939S RNP602113S RNP602138S RNP601820S RNP601806S RNP601658S RNP602158S RNP601963S RNP601614S RNP601962S RNP602035S RNP602021S RNP601965S RNP601749S RNP601805S RNP602033S RNP601947S RNP602134S RNP601715S RNP602129S RNP602144S RNP601645S RNP601601S RNP601602S RNP601603S RNP601604S RNP601605S RNP601577S RNP601997S RNP601998S RNP602142S RNP602143S RNP601684S RNP602130S RNP602145S RNP602146S RNP601670S RNP602127S RNP602069S RNP601782S RNP601506S RNP601507S RNP601508S RNP601912S RNP601913S RNP601957S RNP601958S RNP602058S RNP602059S RNP602018S RNP601966S RNP601613S RNP601537S RNP601538S RNP601539S RNP601542S RNP601544S RNP601547S RNP601548S RNP601549S RNP601550S RNP601551S RNP601990S RNP601991S RNP601992S RNP601993S RNP601994S RNP601796S RNP601967S RNP602017S RNP601496S RNP602064S RNP601553S RNP601534S RNP601762S RNP601641S RNP601867S RNP601620S RNP601864S RNP601865S RNP601933S RNP601934S RNP601937S RNP601938S RNP601980S RNP602032S RNP601974S RNP601946S RNP602031S RNP601834S RNP601469S RNP601803S RNP601843S RNP601844S RNP601470S RNP601964S RNP602001S RNP601953S RNP602137S RNP601916S RNP601897S RNP601900S RNP601740S RNP601904S RNP602101S RNP602165S RNP601866S RNP602084S RNP602087S RNP601905S RNP601935S RNP602077S RNP601841S RNP601984S RNP601986S RNP601787S RNP601919S RNP602140S RNP601807S RNP602076S RNP602118S RNP601598S RNP601764S RNP602125S RNP601883S RNP601902S RNP601653S RNP602133S RNP601795S RNP601876S RNP601871S RNP602131S RNP601754S RNP601813S RNP602102S RNP601999S RNP601557S RNP601558S RNP601559S RNP601560S RNP601706S RNP602072S RNP601630S RNP601959S RNP601637S RNP601775S RNP602071S RNP601961S RNP601931S RNP601742S RNP601509S RNP601510S RNP601511S RNP601512S RNP601513S RNP601515S RNP601516S RNP601517S RNP601518S RNP601519S RNP601520S RNP601521S RNP601522S RNP601523S RNP601524S RNP601799S RNP602088S RNP602089S RNP602090S RNP602091S RNP602010S RNP602039S RNP602040S RNP602046S RNP601949S RNP601652S RNP602053S RNP602061S RNP601468S RNP601878S RNP601842S RNP601504S RNP601952S RNP602128S RNP602085S RNP602044S RNP601632S RNP601654S RNP601845S RNP601663S RNP602051S RNP601623S RNP601624S RNP601661S RNP601532S RNP601535S RNP601536S RNP601540S RNP601541S RNP601543S RNP601545S RNP602011S RNP602029S RNP601627S RNP601628S RNP601606S RNP602167S RNP601752S RNP601771S RNP601772S RNP602164S RNP601581S RNP601650S RNP601685S RNP601689S RNP601690S RNP601691S RNP601692S RNP601693S RNP601694S RNP601695S RNP601696S RNP601697S RNP601698S RNP601699S RNP601700S RNP601701S RNP601702S RNP601703S RNP601704S RNP601705S RNP601707S RNP601709S RNP601710S RNP601711S RNP601712S RNP601713S RNP601714S RNP601625S RNP601639S RNP601640S RNP601868S RNP601492S RNP601500S RNP602117S RNP601495S RNP601766S RNP601642S RNP601753S RNP601863S RNP601972S RNP602156S RNP601622S RNP601655S RNP602096S RNP601599S RNP601672S RNP601750S RNP601756S RNP601786S RNP602121S RNP602099S RNP601626S RNP602043S RNP601688S RNP601860S RNP601960S RNP601600S RNP601846S RNP601847S RNP601848S RNP601976S RNP601977S RNP601978S RNP602024S RNP601773S RNP601561S RNP601562S RNP601563S RNP601564S RNP601565S RNP601566S RNP601567S RNP601568S RNP601569S RNP601570S RNP601629S RNP601819S RNP601800S RNP602097S RNP602080S RNP601631S RNP601829S RNP601836S RNP601896S RNP601988S RNP601621S RNP601636S RNP602152S RNP602163S RNP602020S RNP602025S RNP601664S RNP601683S RNP601839S RNP601885S RNP601955S RNP601922S RNP601923S RNP602094S RNP602000S RNP602170S RNP602030S RNP602047S RNP601792S RNP601851S RNP601968S RNP602037S RNP602135S RNP601585S RNP601765S RNP601932S RNP601920S RNP601673S RNP601675S RNP601676S RNP601677S RNP601678S RNP601679S RNP601467S RNP601574S RNP601575S RNP601616S RNP601617S RNP601781S RNP601481S RNP601485S RNP601850S RNP601901S RNP602042S RNP601546S RNP601584S RNP602083S RNP601667S RNP601884S RNP601911S RNP602109S RNP601853S RNP601854S RNP601855S RNP601856S RNP601857S RNP601979S RNP601571S RNP601572S RNP601533S RNP601582S RNP601797S RNP602049S RNP602086S RNP602126S RNP601668S RNP601647S RNP601887S RNP601888S RNP601889S RNP601666S RNP601643S RNP601482S RNP601757S RNP601776S RNP602045S RNP601680S RNP601651S RNP601759S RNP601760S RNP602078S RNP602116S RNP601791S RNP601948S RNP601951S RNP601970S RNP601971S RNP602073S RNP602074S RNP601950S RNP602062S RNP602068S RNP601798S RNP602005S RNP602006S RNP602007S RNP601573S RNP601936S RNP602002S RNP601784S RNP601768S RNP601769S RNP601493S RNP601497S RNP601498S RNP601499S RNP601501S RNP601925S RNP601926S RNP601927S RNP601928S RNP601929S RNP601716S RNP601717S RNP601718S RNP601719S RNP601720S RNP601721S RNP601722S RNP601723S RNP601724S RNP601725S RNP601903S RNP601881S RNP601669S RNP601708S RNP602056S RNP601838S RNP601985S RNP601578S RNP601580S RNP602149S RNP601989S RNP601892S RNP601895S RNP601898S RNP601924S RNP601983S RNP602041S RNP602095S RNP602147S RNP601817S RNP601943S RNP602027S RNP602159S RNP601780S RNP601505S RNP601659S RNP601869S RNP601802S RNP601987S RNP602013S RNP602148S RNP602150S RNP602151S RNP601665S RNP601671S RNP601859S RNP601861S RNP601894S RNP601514S RNP601525S RNP601814S RNP601788S RNP602015S RNP602066S RNP602075S RNP602111S RNP602012S RNP602098S RNP602065S RNP602160S RNP602161S RNP601483S RNP601484S RNP601487S RNP601489S RNP601494S RNP601915S RNP601886S RNP602060S RNP602104S RNP602105S RNP602106S RNP601930S RNP601583S RNP601638S RNP601660S RNP602174S RNP601751S RNP601840S RNP601808S RNP601648S RNP601815S RNP601872S RNP601832S RNP601789S RNP601941S RNP601833S RNP601810S RNP602022S RNP601858S RNP602112S RNP602026S RNP601662S RNP602122S RNP602162S RNP601811S RNP601586S RNP601587S RNP601824S RNP602114S RNP601576S RNP601801S RNP601981S RNP601503S RNP601879S RNP602100S RNP601758S RNP601755S RNP602079S RNP601526S RNP601527S RNP601528S RNP601530S RNP601531S RNP601770S RNP601774S RNP601777S RNP601681S RNP601682S RNP601686S RNP601687S RNP602157S RNP602057S RNP601490S RNP601491S RNP602019S RNP601726S RNP601727S RNP601728S RNP601729S RNP601973S RNP602008S RNP602173S RNP601907S RNP601944S RNP602009S RNP602082S RNP601555S RNP601730S RNP601731S RNP601816S RNP601945S RNP602048S RNP602153S RNP602154S RNP602052S RNP601615S RNP601732S RNP601733S RNP601747S RNP601748S RNP601818S RNP602168S RNP602169S RNP601837S RNP602067S RNP601763S RNP601790S RNP601849S RNP601646S RNP601828S RNP601778S RNP602036S RNP601657S RNP602070S RNP601735S RNP601736S RNP601737S RNP601738S RNP601739S RNP601741S RNP601743S RNP601744S RNP601745S RNP601746S RNP601827S RNP601982S RNP602124S RNP601882S RNP601908S RNP602063S RNP601831S RNP602081S RNP601877S RNP601906S RNP602003S RNP601588S RNP601589S RNP601590S RNP601591S RNP601592S RNP601593S RNP601594S RNP601595S RNP601596S RNP601597S RNP601633S RNP601635S RNP601821S RNP601890S RNP601891S RNP601940S RNP601785S RNP601556S RNP601612S RNP601644S RNP601942S RNP601826S RNP602115S RNP602139S RNP602050S RNP602023S RNP602141S RNP601830S RNP601649S RNP601825S RNP601823S RNP601809S RNP602103S RNP602175S RNP601893S RNP601975S RNP601607S RNP601608S RNP601609S RNP601610S RNP601611S RNP601783S RNP602028S RNP601822S RNP602092S RNP601761S RNP601619S RNP601880S RNP601874S RNP602093S RNP601870S RNP601502S RNP601914S RNP602014S RNP601873S RNP601471S RNP601472S RNP601473S RNP601529S RNP602123S RNP602119S RNP601995S RNP601618S RNP602004S RNP602110S RNP602166S RNP601779S RNP601474S RNP601475S RNP601476S RNP601477S RNP601478S RNP601479S RNP601480S RNP601875S RNP601835S RNP601486S RNP601488S RNP601767S RNP602136S RNP602172S RNP602107S RNP602108S RNP602120S RNP601794S RNP601899S RNP601909S RNP601918S RNP602055S
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Percepta™ Quad CRT-P MRI SureScan™ W4TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall28 deaths897 reports
- Percepta™ Quad CRT-P MRI SureScan™ W4TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall8 deaths703 reports
Product lines
Other devices under the same FDA submission
20 devices cleared under P010015 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ATTAIN® BIPOLAR OTW 419488Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode15 deaths730 reports
- ATTAIN® BIPOLAR OTW 419478Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode5 deaths464 reports
- Percepta™ CRT-P MRI SureScan™ W1TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall4 deaths300 reports
- Percepta™ CRT-P MRI SureScan™ W1TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall2 deaths218 reports
- CareLink SmartSync™ Percepta Serena Solara Application D00U004Medtronic, Inc. · Pacemaker/icd/crt non-implanted components4 recalls0 reports
- Consulta® CRT-P C4TR01Medtronic, Inc. · Pulse generator, permanent, implantable4 deaths179 reports
- Viva™ CRT-P C6TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 death151 reports
- Solara™ Quad CRT-P MRI SureScan™ W4TR03Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 deaths140 reports
- Serena™ Quad CRT-P MRI SureScan™ W4TR02Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)3 deaths118 reports
- Solara™ Quad CRT-P MRI SureScan™ W4TR03Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)5 deaths102 reports
- Serena™ Quad CRT-P MRI SureScan™ W4TR02Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 death90 reports
- Solara™ CRT-P MRI SureScan™ W1TR03Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)3 deaths76 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80433).
- Z-2412-2018 · TerminatedPercepta CRT-P MRI SureScan, REF W1TR01 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and …
- Z-2414-2018 · TerminatedPercepta CRT-P MRI SureScan, REF W1TR04 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and …
- Z-2415-2018 · TerminatedPercepta Quad CRT-P MRI SureScan, REF W4TR04 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page