BACT/ALERT VIRTUO recallZ-0229-2019 · bioMerieux, Inc. · initiated Jun 27, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from bioMerieux, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells P24 - P27, where the calibration standards typically reside. If the calibration standards are not returned to the designated locations, an anomaly may occur.

FDA root cause category: Software design

What the company is doing

bioMerieux were initially informing customers with on-site visits beginning on 06/27/2018. The letter informed customers of the anomaly and that local Field Service Engineers were confirming the calibration standards are in the designated bottle cell locations. Actions requested included to ensure the recall notice was distributed to applicable personnel within the organization, notify bioMerieux support if a patient/test bottle is loaded into cells P23-P27, print or view bottle reports to ensure accurate bottle results, store the recall letter with bioMerieux documentation, and complete and return the Acknowledgement Form. A second notification was initiated on about 08/08/2018. The letter was mailed via FedEx Overnight to ensure all customers were informed of the recall prior to the local Field Service Engineers performing their visit and device examination. Instructions included to ensure the recall notification was distributed to all appropriate personnel in the organization, to be vigilant when reviewing alarms for anonymous and/or duplicate bottles, if the customer becomes aware of the patient/test bottles located in cells P24-P27 to contact their local bioMerieux support, to view and/or print bottle reports to ensure accurate bottle results, to store the recall notification with bioMerieux, Inc. BACT/ALERT VIRTUO documentation, and to complete and return the Acknowledgement Form in Attachment A. Questions can be directed to loca bioMerieux, Inc. Clinical Custome Service representative at (800) 682-2666 or Industry Customer Service representative at (800) 634-7656.

Recall details

Recall number
Z-0229-2019
Status
Terminated
Initiated by the company
Jun 27, 2018
Recalling firm
bioMerieux, Inc.
Quantity
151 units
Product code
MDB
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KS, LA, MI, MO, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SD, TN, TX, UT, VT, WV. Distributed internationally to Austria, Australia, Belgium, Brazil, Canada, Switzerland, Chile, Colombia, Germany, Spain, France, United Kingdom, Hong Kong, India, Italy, Japan, South Korea, Netherlands, Portugal, Sweden, Singapore, Thailand, South Africa, Saudi Arabia, Philippines, Myanmar, Serbia, Lebanon, United Arab Emirates, Slovenia, Qatar, Costa Rica, Croatia, Panama, Estonia, Israel, Philippines, Malaysia, Brunei Darussalam, Vietnam, Kuwait.

Product description

BACT/ALERT VIRTUO system, B Unit, with Version R2.0 Firmware Product Usage: The BacT/ALERT VIRTU Microbial Detection System is an automated microbial test system capable of incubating, agitating, and continuously monitoring for the detection of aerobic, facultative, and anaerobic microorganism growth from blood and other normally sterile body fluids.

Code information (models, lots, serial numbers)

Reference Number 411661, UDI 03573026369774, all serial numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 80759).