BACT/ALERT VIRTUORecalls & reported problems across all models
Biomerieux , Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 41
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 2
We found 2 models of BACT/ALERT VIRTUO by Biomerieux , Inc. in the FDA's device registry (GUDID). Together they are named in 41 adverse event reports in the FDA's MAUDE database, received between Dec 13, 2017 and May 21, 2026 — classified as 2 injuries, 39 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (7 reports).
2 FDA recall records name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
BACT/ALERT VIRTUO recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jun 27, 2018Terminated
The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells P24 - P27, where the calibration standards typically reside. If the calibration standards are not returned to the designated locations, an anomaly may occur.
Z-0229-2019 · Root cause: Software design · Recall details · FDA record ↗
- Jun 27, 2018Terminated
The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells P24 - P27, where the calibration standards typically reside. If the calibration standards are not returned to the designated locations, an anomaly may occur.
Z-0228-2019 · Root cause: Software design · Recall details · FDA record ↗
Most reported BACT/ALERT VIRTUO problems
Device problems
- False Negative Result7
- Incorrect, Inadequate or Imprecise Result or Readings5
- Data Problem4
- Communication or Transmission Problem3
- Date/Time-Related Software Problem3
- Failure to Obtain Sample3
- Device-Device Incompatibility2
- False Positive Result2
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Known Impact Or Consequence To Patient9
- No Consequences Or Impact To Patient6
- No Patient Involvement2
- Injury1
- Insufficient Information1
- Patient Problem/Medical Problem1
BACT/ALERT VIRTUO models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- BACT/ALERT® VIRTUO® 411660System, blood culturing1 recall37 reports
- BACT/ALERT® VIRTUO® 411661System, blood culturing1 recall4 reports
Recent BACT/ALERT VIRTUO adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2026Difficult to Open or Close
INTENDED USE: BACT/ALERT® VIRTUO¿ MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS. ISSUE DESCRIPTION: ON (B)(6) 2026, A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX THAT ONE OF THEIR OPERATORS WAS INJURED WHEN THE WASTE TRAY DOOR OF THEIR VIRTUO, A UNIT (REF 411660, SERIAL (B)(6)) FELL ON THE OPERATOR'S FOOT. THE CUSTOMER REPORTED THAT THE OPERATOR WAS TRYING TO EMPTY THE VIRUTO'S WASTE TRY AND THE WASTE TRAY DOOR FELL ONTO THE OPERATOR'S FOOT. THE INJURED EMPLOYEE WAS SENT TO URGENT CARE, WAS PUT ON LEAVE FOR THE REST OF THE WEEK, AND THEN RETURNED TO WORK ON MODIFIED DUTY. BIOMÉRIEUX HAS REQUESTED DETAILS REGARDING THE SPECIFIC INJURY AND TREATMENT NECESSARY, BUT THE CUSTOMER DECLINED TO PROVIDE FURTHER INFORMATION. THE INJURY OCCURRED ON (B)(6) 2026 BUT WAS NOT REPORTED TO BIOMÉRIEUX UNTIL (B)(6) 2026. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE ON (B)(6) 2026 TO REPAIR THE VIRTUO. UPON ARRIVAL, THE FSE IDENTIFIED THAT THE DRAWER WAS MISSING HARD
Patient problems: Insufficient Information
MAUDE report 25240729 ↗MalfunctionMar 1, 2024Device-Device Incompatibility; Device-Device Incompatibility
INTENDED USE BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS AND FROM BLOOD PLATELET SAMPLES. IN BLOOD BANK FACILITIES, THE BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM WILL PROVIDE IN-PROCESS TESTING FOR THE PRESENCE OR ABSENCE OF MICROORGANISMS. DESCRIPTION OF THE ISSUE A CUSTOMER IN THE UNITED KINGDOM NOTIFIED BIOMÉRIEUX OF OBTAINING A DELAYED RESULT WHEN USING THE VIRTUO, A UNIT - REFERENCE 411660 (SERIAL NUMBER (B)(6). THE CUSTOMER NOTIFIED THAT HE HAS TWO VIRTUO INSTRUMENTS : A VIRTUO AT LUTON AND A VIRTUO AT BEDFORD. IT IS UNDERSTOOD THAT POSITIVE BOTTLE RESULTS FROM LUTON HAVE NOT BEEN RECEIVED TO THE LABORATORY INFORMATION SYSTEM (LIS) AT BEDFORD WHERE THE BOTTLES ARE FURTHER TESTED. THE CUSTOMER INDICATED THAT THE GRAM STAINING HAS NOT BEEN PERFORMED FOR THESE BOTTLES BECAUSE BEDFORD SITE HAS NOT RECEIVED THESE BOTTLES. THEREFORE, FOR NOW, IT IS NOT POSSIBLE TO KNOW WHETHER OR NOT THESE BOTTLES WERE A TRUE POSITIVE, OR A FALSE POSITIVE. AT THE TIM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18817638 ↗MalfunctionMar 1, 2024Device-Device Incompatibility; Device-Device Incompatibility
INTENDED USE. BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS AND FROM BLOOD PLATELET SAMPLES. IN BLOOD BANK FACILITIES, THE BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM WILL PROVIDE IN-PROCESS TESTING FOR THE PRESENCE OR ABSENCE OF MICROORGANISMS. DESCRIPTION OF THE ISSUE. A CUSTOMER IN THE UNITED KINGDOM NOTIFIED BIOMÉRIEUX OF OBTAINING A DELAYED RESULT WHEN USING THE VIRTUO, A UNIT - REFERENCE 411660 (SERIAL NUMBER# (B)(6)). THE CUSTOMER NOTIFIED THAT HE HAS TWO VIRTUO INSTRUMENTS : A VIRTUO AT LUTON AND A VIRTUO AT BEDFORD. IT IS UNDERSTOOD THAT POSITIVE BOTTLE RESULTS FROM LUTON HAVE NOT BEEN RECEIVED TO THE LABORATORY INFORMATION SYSTEM (LIS) AT BEDFORD WHERE THE BOTTLES ARE FURTHER TESTED. THE CUSTOMER INDICATED THAT THE GRAM STAINING HAS NOT BEEN PERFORMED FOR THESE BOTTLES BECAUSE BEDFORD SITE HAS NOT RECEIVED THESE BOTTLES. THEREFORE, FOR NOW, IT IS NOT POSSIBLE TO KNOW WHETHER OR NOT THESE BOTTLES WERE A TRUE POSITIVE, OR A FALSE POSITIVE. AT THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18817606 ↗MalfunctionFeb 6, 2024Mechanical Jam; Mechanical Jam
INTENDED USE: BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS AND FROM BLOOD PLATELET SAMPLES. ISSUE DESCRIPTION: A CUSTOMER IN FINLAND NOTIFIED BIOMÉRIEUX OF OBTAINING NO RESULT IN ASSOCIATION WITH THEIR VIRTUO® A UNIT (REF. (B)(4), SERIAL (B)(6). CUSTOMER REPORTED THEIR VIRTUO SECONDARY INSTRUMENT TO GIVE LOT OF ROBOT JAM ERRORS. THEY SAID IT OCCURRED SHORTLY AFTER INSTRUMENT HAS UNLOADED NEGATIVE BOTTLES. WHEN THEY OPENED THE INSTRUMENT DOOR TO FIX THE ERROR SYSTEM HAD CALIBRATION SLUG IN THE ROBOT HAND. CUSTOMER ALSO MENTIONED THAT THERE WAS A NOISE WHEN ROBOT MOVES BACK TO THE HOME POSITION (THEY ASSUME) WHICH IS SIMILAR TO THE NOISE WHEN BOTTLE IS UNLOADED TO THE RETRIEVAL AREA. THEY WERE WONDERING IF IT HITS SOMEWHERE. BIOMÉRIEUX FIELD SPECIALIST ENGINEER FOUND (B)(4) BOTTLES IN THE INSTRUMENT WHICH INSTRUMENT THOUGHT TO BE UNLOADED. GROWTH CURVE WAS ONLY CREATED FOR < 24 HOURS EVEN THOUGH BOTTLES HAVE BEEN INSIDE THE INSTRUMENT SINCE (B)(6) AND NO RES
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18648626 ↗MalfunctionJul 26, 2023False Negative Result; False Negative Result
PRODUCT DESCRIPTION ********************** BACT/ALERT® VIRTUO¿ MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS. COMPLAINT DESCRIPTION ************************* A CUSTOMER IN COLOMBIA COMPLAINED AFTER HAVING OBTAINED A FALSE NEGATIVE RESULT WITH THE VIRTUO, A UNIT (REF (B)(4), SERIAL NUMBER (B)(6). THE CUSTOMER SAID THAT THEY HAD INOCULATED A PATIENT'S SAMPLE (PERITONEAL LIQUID) ONTO BACT/ALERT FN PLUS BOTTLE AND THAT AFTER 9 DAYS OF INCUBATION THE BOTTLE WAS UNLOADED AS NEGATIVE. THE CUSTOMER REPORTED THAT THEY HAD INSPECTED THE BOTTLE AND FOUND THAT THE BOTTOM OF THE BOTTLE'S SENSOR WAS YELLOW, INDICATING THE BOTTLE TO BE POSITIVE THEREFORE THEY PERFORMED A GRAM STAIN AND HAD FOUND GRAM POSITIVE COCCI. THE CUSTOMER PERFORMED IDENTIFICATION TEST THAT CONFIRMED THE MICROOGANISM TO BE S. EPIDERMIDIS. NO FURTHER DETAILS WERE PROVIDED THE CUSTOMER STATES THAT THE CONCERNED PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17397175 ↗MalfunctionJun 28, 2023Communication or Transmission Problem; Communication or Transmission Problem
INTENDED USE: BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS AND FROM BLOOD PLATELET SAMPLES. ISSUE DESCRIPTION: A CUSTOMER IN SWEDEN NOTIFIED BIOMÉRIEUX OF OBTAINING NO RESULT IN ASSOCIATION WITH THEIR VIRTUO® A UNIT (REF. 411660, SERIAL (B)(6)). THIS COMPLAINT IS RELATED TO THE COMPLAINT (B)(4) (SAME CUSTOMER, SAME ISSUE, VIRTUO® A UNIT (REF. 411660, SERIAL (B)(6))). IT WAS REPORTED THAT CUSTOMER CAN¿T FIND TWO BOTTLES IN THEIR VIRTUO. BOTTLES WERE LOADED ON JUNE 3RD. THE INFORMATION FOR THESE BOTTLES HAS AGED OUT OF THE INSTRUMENT LOG FILES. BIOMÉRIEUX GLOBAL CUSTOMER SERVICE SEARCHED FOR THE BOTTLES ON MYLA AND FOUND A RESULT RECORD FOR BOTTLE ID NR2LSBDD AND COULD ONLY FIND A LOAD EVENT RECORD FOR BOTTLE ID AR53MDJ8. THE RESULTS WERE AS FOLLOWS: BOTTLE ID AR53MDJ8: STATUS UNKNOWN BOTTLE ID NR2LSBDD: NEGATIVE IN THIS CONTEXT, DELAYED RESULT IS SUSPECTED BUT NOT CONFIRMED AT THE TIME OF THE GLOBAL ASSESSMENT. NO PATIENT HEALTH IMPACT HAS BEEN REPORTED DUE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17218370 ↗MalfunctionJun 28, 2023Communication or Transmission Problem; Communication or Transmission Problem
INTENDED USE: BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS AND FROM BLOOD PLATELET SAMPLES. ISSUE DESCRIPTION: A CUSTOMER IN SWEDEN NOTIFIED BIOMÉRIEUX OF OBTAINING NO RESULT IN ASSOCIATION WITH THEIR VIRTUO® A UNIT (REF. (B)(4), SERIAL (B)(6)). IT WAS REPORTED THAT CUSTOMER CAN´T FIND BOTTLES IN THEIR VIRTUO, LAST ACTION FROM BOTH MYLA AND AUDIT TRAIL IS BOTTLE RELOADED 6/6. WHEN SEARCHING IN VIRTUO THERE IS NO RESULT FOR THE BOTTLE IDS MISSING. CUSTOMER IS CONCERNED ABOUT HOW THEY CAN FIND INFO OF THE BOTTLE, IF IT STAYED IN THE CELL 5 DAYS SO IT IS CONSIDERED NEGATIVE, OR IF IT WAS REMOVED BEFORE THAT. IN THIS CONTEXT, THE RESULT WAS DELAYED AS IT WAS MORE THAN 5 DAYS SINCE THE BOTTLES WERE LOADED. NO PATIENT HEALTH IMPACT HAS BEEN REPORTED DUE TO DELAYED RESULTS. BIOMÉRIEUX GLOBAL CUSTOMER SERVICE REQUESTED ADDITIONAL INFORMATION TO UNDERSTAND WHY THESE BOTTLES CANNOT BE FOUND IN THE VIRTUO APPLICATION AND WHY THE INFORMATION WAS NOT SENT TO MYLA. AN I
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17218358 ↗MalfunctionDec 9, 2022Failure to Power Up; Failure to Power Up
INTENDED USE: BACT/ALERT® VIRTUO¿ MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS. ISSUE DESCRIPTION: A BLOOD BANK CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX OF A MALFUNCTION WITH THE VIRTUO, A UNIT (REF (B)(4), SERIAL NUMBER (B)(4)) THAT OCCURRED AFTER A PC REPLACEMENT AND WHICH LEAD TO DELAYED RESULTS GREATER THAN ONE DAY. THE CUSTOMER REPORTED THAT :¿PC WILL NOT BOOT. CYCLING POWER TO THE INSTRUMENT WAS UNSUCCESSFUL¿. A LOCAL FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER SITE AND REPLACED THE PC. AFTER PC REPLACEMENT, THE EXISTING BOTTLES UNDER TEST UNLOADED. THE CUSTOMER RELOADED THE BOTTLES. AS A PRECAUTION, THE CUSTOMER WAS REQUIRED TO SUBCULTURE THE BOTTLES REMAINING IN THE INCUBATOR. AT THE TIME OF ASSESSMENT, THERE IS NO INDICATION OR REPORT FROM THE CUSTOMER THAT THIS EVENT LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. AN INVESTIGATION WILL BE INITIATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15948423 ↗MalfunctionAug 16, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
INTENDED USE:BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS AND FROM BLOOD PLATELET SAMPLES.ISSUE DESCRIPTION:A CUSTOMER IN ITALY NOTIFIED BIOMÉRIEUX OF A LACK OF RESULT FOR A MISSING BOTTLE IN ASSOCIATION WITH THEIR VIRTUO® A UNIT (REF. 411660, SERIAL (B)(4)) WHICH LED TO DELAYED RESULTS >24 HOURS.IT WAS REPORTED THAT A CUSTOMER HAD NO INFORMATION ON THE OUTCOME OF A PATIENT CULTURE BOTTLE THAT SHOULD HAVE COMPLETED TESTING. THE CUSTOMER STATED THE BOTTLE WAS LOADED ON THE (B)(6) IN THEIR VIRTUO INSTRUMENT AT 16:20 AND DOWNLOADED WITHOUT ANY RESULT AT 18:25. ACCORDING TO THE TECHNICIANS ON SHIFT, THE INSTRUMENT WOULD NOT ALLOW BOTTLES TO BE LOADED AND UNLOADED FROM THE INSTRUMENT; ASSISTANCE WAS CALLED AT APPROXIMATELY 17:45. IT WAS REPORTED THERE WAS GUIDANCE TO KEEP THE BOTTLES IN THE DARK AT ROOM TEMPERATURE UNTIL THE INSTRUMENT WAS UNLOCKED. THE NEXT MORNING, THE SITUATION SEEMED TO HAVE REMAINED THE SAME. A BIOMÉRIEUX FIELD SERVICE ENGINEER (FSE) CONNE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15233983 ↗MalfunctionJul 12, 2022False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN CANADA NOTIFIED BIOMÉRIEUX OF OBTAINING FALSE POSITIVE BOTTLE RESULTS FOR FIVE (5) PATIENTS WHEN USING THE BACT/ALERT VIRTUO A UNIT (REF.: 411660, SERIAL NUMBER: (B)(4)). BIOMÉRIEUX CUSTOMER SERVICE ANALYZED THE REPORTS AND IDENTIFIED THE EVENT AS FALSE POSITIVE RESULTS. THE RESULTS WERE AS FOLLOWS FOR EACH PATIENT (1 BOTTLE FA AND 1 BOTTLE FN PER ACCESSION NUMBER): PATIENT 1: ACCESSION NUMBER (B)(6). BOTTLE AR2P9K3F: RESULT IS NEGATIVE. BOTTLE NR0N5KWG: RESULT IS NEGATIVE, FLAGGED AS POSITIVE AFTER 2.7 - 2.8 HOURS. THE CUSTOMER APPROPRIATELY HANDLED THE BOTTLE BY RELOADING THE FALSE POSITIVE AS PER THE USER MANUAL. THERE WAS NO DELAY IN RESULTS, BASED ON LOG EVALUATION. PATIENT 2: ACCESSION NUMBER (B)(6). BOTTLE AR2PYYPM: RESULT IS POSITIVE. BOTTLE NR0N5KS2: RESULT IS POSITIVE. THE CUSTOMER APPROPRIATELY HANDLED THE BOTTLE BY RELOADING THE FALSE POSITIVE AS PER THE USER MANUAL. THERE WAS NO DELAY IN RESULTS, BASED ON LOG EVALUATION. THE CUSTOMER REPORTED THIS AS A TRUE POSITIVE. PATIENT 3: ACCESSION NUMBER (B)(6). BOTTLE AR2H60VV: RESULT IS NEGATIVE TO DATE. BOTTLE NR0N5K1Z: RESULT IS NEGATIVE TO DATE. CUSTOMER INDICATED THAT FOR THIS PATIENT, WHEN THE BOTTLE FLAGGED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14984420 ↗MalfunctionJul 12, 2022False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF OBTAINING FALSE POSITIVE BOTTLE RESULTS FOR FIVE (5) PATIENTS WHEN USING THE BACT/ALERT VIRTUO A UNIT (REF.: (B)(4), SERIAL NUMBER: (B)(4)). BIOMÉRIEUX CUSTOMER SERVICE ANALYZED THE REPORTS AND IDENTIFIED THE EVENT AS FALSE POSITIVE RESULTS. THE RESULTS WERE AS FOLLOWS FOR EACH PATIENT (1 BOTTLE FA AND 1 BOTTLE FN PER ACCESSION NUMBER): PATIENT 1: ACCESSION NUMBER 484984. BOTTLE AR2P9K3F: RESULT IS NEGATIVE. BOTTLE NR0N5KWG: RESULT IS NEGATIVE, FLAGGED AS POSITIVE AFTER 2.7 - 2.8 HOURS. THE CUSTOMER APPROPRIATELY HANDLED THE BOTTLE BY RELOADING THE FALSE POSITIVE AS PER THE USER MANUAL. THERE WAS NO DELAY IN RESULTS, BASED ON LOG EVALUATION. PATIENT 2: ACCESSION NUMBER 500323. BOTTLE AR2PYYPM: RESULT IS POSITIVE. BOTTLE NR0N5KS2: RESULT IS POSITIVE. THE CUSTOMER APPROPRIATELY HANDLED THE BOTTLE BY RELOADING THE FALSE POSITIVE AS PER THE USER MANUAL. THERE WAS NO DELAY IN RESULTS, BASED ON LOG EVALUATION. THE CUSTOMER REPORTED THIS AS A TRUE POSITIVE. PATIENT 3: ACCESSION NUMBER 493000. BOTTLE AR2H60VV: RESULT IS NEGATIVE TO DATE. BOTTLE NR0N5K1Z: RESULT IS NEGATIVE TO DATE. CUSTOMER INDICATED THAT FOR THIS PATIENT, WHEN THE BOTTLE FLAGGED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14984404 ↗MalfunctionFeb 22, 2022Failure to Obtain Sample; Failure to Obtain Sample
ISSUE DESCRIPTION: A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF EXPERIENCING AN ISSUE IN WHICH A POSITIVE BACT/ALERT® BOTTLE (REFERENCE NOT PROVIDED) WAS REPORTED TO BE LOST IN THEIR VIRTUO® A UNIT (REF. 411660, SERIAL (B)(4)). DELAYED RESULT (>8 HOURS) WAS OBTAINED FOR ONE ¿MISSING BOTTLE¿. THE CUSTOMER STATED THAT THEY HAD AN ISSUE REGARDING A "MISSING BOTTLE". BIOMÉRIEUX¿S FSE (FIELD SYSTEM ENGINEER) VISITED THE CUSTOMER SITE AND IDENTIFIED THAT THEY HAD A CONNECTION ISSUE DUE TO THE FIREWALL AND ALSO A ROBOT JAM DURING CALIBRATION. THE FIREWALL AND CONNECTION ISSUES WERE RESOLVED, BUT BOTTLES HAD GOTTEN STUCK IN RACKS SO THE RACKS HAD TO BE DISABLED. THE FSE RESOLVED THE ISSUE WITH THE RACKS SO THE INSTRUMENT COULD BE FULLY USED AGAIN WITH ALL RACKS ENABLED. THE ISSUE WITH THE "MISSING BOTTLE" CAUSED A DELAY IN REPORTING RESULT OF A POSITIVE BOTTLE. IT WAS REPORTED THAT THERE WAS BLOOD CULTURE RESULT DELAYED MORE THAN 8 HOURS TO THE PHYSICIAN. THE CUSTOMER ALSO REPORTED THAT THE USERS OF THE INSTRUMENTS MAY HAVE BEEN EXPOSED TO BLOOD DUE TO BROKEN BOTTLES. IT WAS REPORTED THAT FOUR (4) BOTTLES WERE ANALYZED. THE THREE (3) THAT WERE NOT BROKEN GAVE A NEGATIVE RESULT. IN THE END T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13575784 ↗MalfunctionDec 3, 2021False Negative Result; False Negative Result
INTENDED USE: BACT/ALERT® VIRTUO¿ MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS. DESCRIPTION: A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF EXPERIENCING AN ISSUE IN WHICH A BLOOD CULTURE BOTTLE WAS LEFT IN THE CUSTOMER'S VIRTUO, A UNIT (REF 411660, SERIAL (B)(4)) FOR NINE (9) DAYS WITH NO RESULT FOR THE BOTTLE (POTENTIAL FALSE NEGATIVE). THE CUSTOMER REPORTED THAT THE BOTTLE GRAPH LOOKS LIKE THE BOTTLE SHOULD BE POSITIVE FOR GROWTH AND THE THREE (3) OTHER BOTTLES ASSOCIATED WITH THIS PATIENT WERE ALL POSITIVE. DUE TO THE POSITIVE RESULTS OBTAINED FOR THE OTHER THREE (3) BOTTLES, THERE WAS NO DELAY IN REPORTING RESULTS TO THE TREATING PHYSICIAN. THERE IS NO INDICATION OR REPORT FROM THE CUSTOMER THAT THIS EVENT LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. AN INVESTIGATION HAS BEEN INITIATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12929108 ↗MalfunctionSep 7, 2021Failure to Obtain Sample; Failure to Obtain Sample
DESCRIPTION OF THE PROBLEM: A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF EXPERIENCING AN ISSUE IN WHICH THE CUSTOMER FOUND DROPPED BACT/ALERT® BOTTLES (BOTTLE TYPE, REFERENCE NOT REPORTED) IN ASSOCIATION WITH THEIR VIRTUO® A UNIT (REF 411660, SERIAL (B)(4)). NO RESULT WAS OBTAINED FOR ONE DROPPED BOTTLE. THE CUSTOMER STATED THAT THEY FOUND THREE BLOOD CULTURE BOTTLES ON THE FLOOR OF THE VIRTUO INSTRUMENT. THE CUSTOMER STATED THESE BOTTLES HADN'T BEEN FLAGGED AS BEING DROPPED. GLOBAL CUSTOMER SERVICE (GCS) REVIEWED THE INSTRUMENT LOGS. FINAL GRAM STAIN/SMEAR RESULTS FOR EACH AFFECTED BOTTLE, AS WELL AS FINAL RESULTS OF MATE BOTTLES, WERE AS FOLLOWS: ALL GRAM STAINS NEGATIVE. MATE BOTTLE RESULTS: * 210445641AN = NO GROWTH AT 48 HRS, * 210445641AE = NO GROWTH AT 48 HRS, * 210431854AE = NO GROWTH AT 5 DAYS INCUBATION. BOTTLE 210431854 UNABLE TO PROVIDE ANAEROBIC BLOOD CULTURE RESULTS; THE ISSUE WAS IDENTIFIED TWO (2) WEEKS AFTER LOADING, LEADING TO DELAYED RESULT. A HISTORY OF ROBOTIC ARM FAILURE WAS EVIDENCED PRIOR TO THESE ERRORS, WHICH HAD BEEN PREVIOUSLY RESOLVED BY FIELD SERVICE ENGINEER (FSE). NO ERROR CODES WERE IDENTIFIED ON THE DAY OF LOADING OF AFFECTED BOTTLES. THERE IS NO IN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12436512 ↗MalfunctionAug 27, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
INTENDED USE: BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM IS AN AUTOMATED MICROBIAL TEST SYSTEM CAPABLE OF INCUBATING, AGITATING, AND CONTINUOUSLY MONITORING FOR THE DETECTION OF AEROBIC, FACULTATIVE, AND ANAEROBIC MICROORGANISM GROWTH FROM BLOOD AND OTHER NORMALLY STERILE BODY FLUIDS AND FROM BLOOD PLATELET SAMPLES. IN BLOOD BANK FACILITIES, THE BACT/ALERT® VIRTUO® MICROBIAL DETECTION SYSTEM WILL PROVIDE IN-PROCESS TESTING FOR THE PRESENCE OR ABSENCE OF MICROORGANISMS. DESCRIPTION OF THE ISSUE: A CUSTOMER IN (B)(6) NOTIFIED BIOMERIEUX OF EXPERIENCING AN INCREASE OF FALSE POSITIVE FA PLUS AND FN PLUS BOTTLES USING THE BACT/ALERT® VIRTUO® A UNIT (REFERENCE 411660, SERIAL NUMBER (B)(4), SOFTWARE VERSION (B)(4)). AT THE TIME OF THE CUSTOMER NOTIFICATION, IT WAS INDICATED THAT DURING THE MONTH OF (B)(6) 2021 APPROXIMATELY 300 CULTURE BOTTLES FROM VARIOUS BACT/ALERT BOTTLE LOTS WERE CONCERNED BY FALSE POSITIVE RESULTS. THE CUSTOMER WAS NOT WILLING TO COLLECT/PROVIDE DETAILS REGARDING THE NUMBER OF PATIENTS OR CLINICAL DECISIONS ASSOCIATED WITH THE IMPACTED CULTURE BOTTLES. THE CUSTOMER PERFORMED GRAM STAINING AND SUBCULTURE FOR BOTTLES FLAGGING POSITIVE. THEN, THE BOTTLES WERE RELOAD
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12379199 ↗
Other Biomerieux , Inc. products
All Biomerieux , Inc. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page