DxTerity TRA recallZ-3253-2018 · Medtronic Vascular · initiated Aug 16, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Vascular or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medtronic has determined that the French size indicator on the inner pouch may incorrectly reflect a 6 French size, rather than the correct 5 French size. The catheters inside the pouch are the correct U5TRAN35 item, the outer carton has all the correct labeling, and all the other information and configuration details on both the carton and the pouch are correct.

FDA root cause category: Error in labeling

What the company is doing

Medtronic initiated consignee communications on 16-Aug-2018 to Risk Managers and Health Care Professionals. US notifications were sent via 2-Day UPS delivery. Consignees are asked to quarantine and return any unused potentially affected product to Medtronic.

Recall details

Recall number
Z-3253-2018
Status
Terminated
Initiated by the company
Aug 16, 2018
Recalling firm
Medtronic Vascular
Quantity
250 units
Product code
DQO
FDA submissions
K161287
Distribution
CA, IA, NE, TN, TX, VA, WA, Jamaica, New Zealand, Thailand

Product description

Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35

Code information (models, lots, serial numbers)

UDI/GTIN: 20643169738717 Lot Number 60068186

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

14 devices cleared under K161287 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.