DxTerity TRARecalls & reported problems across all models
Medtronic, Inc. · 9 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls
- 1
We found 9 models of DxTerity TRA by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 9 adverse event reports in the FDA's MAUDE database, received between Apr 21, 2017 and Feb 16, 2021 — classified as 8 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
1 FDA recall record name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
DxTerity TRA recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 16, 2018Terminated
Medtronic has determined that the French size indicator on the inner pouch may incorrectly reflect a 6 French size, rather than the correct 5 French size. The catheters inside the pouch are the correct U5TRAN35 item, the outer carton has all the correct labeling, and all the other information and configuration details on both the carton and the pouch are correct.
Z-3253-2018 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported DxTerity TRA problems
Device problems
- Adverse Event Without Identified Device or Use Problem7
- Device Operates Differently Than Expected1
- Failure to Advance1
- Material Deformation1
Patient problems
- Intimal Dissection4
- Patient Problem/Medical Problem3
- Injury2
- No Clinical Signs, Symptoms or Conditions1
- Occlusion1
- Stroke/CVA1
- Vascular Dissection1
DxTerity TRA models
9 models, most reported first. Each model has its own page with its reports, recalls and device details.
- DxTerity™ TRA U5TRAN35CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
- DxTerity™ TRA U5ULTRA40SHCATHETER, INTRAVASCULAR, DIAGNOSTIC2 reports
- DxTerity™ TRA U5TRAP35SHCATHETER, INTRAVASCULAR, DIAGNOSTIC1 report
- DxTerity™ TRA U5TRAP40SHCATHETER, INTRAVASCULAR, DIAGNOSTIC1 report
- DxTerity™ TRA U6TRAP40SHCATHETER, INTRAVASCULAR, DIAGNOSTIC1 report
- DxTerity™ TRA U5TRAN35SHACATHETER, INTRAVASCULAR, DIAGNOSTIC1 report
- DxTerity™ TRA U5TRAN45SHACATHETER, INTRAVASCULAR, DIAGNOSTIC1 report
- DxTerity™ TRA U5TRACTSHACATHETER, INTRAVASCULAR, DIAGNOSTIC1 report
- DxTerity™ TRA U6ULTRA40SHACATHETER, INTRAVASCULAR, DIAGNOSTIC1 report
Recent DxTerity TRA adverse event reports
The 9 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionFeb 16, 2021Failure to Advance
ELDERLY MALE WITH HISTORY OF ATRIAL FIBRILLATION, HYPERTENSION AND HEART FAILURE. WITH NEW ONSET CHEST PAIN, TO HEART CATH LAB FOR DIAGNOSTIC HEART CATH. WHEN PREPPING THE CATH, THE J-WIRE WOULD NOT GO THROUGH THE CATHETER. PRODUCT WAS NOT USED ON PATIENT, NEW DEVICE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11331806 ↗InjuryJul 16, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS STATED THAT A DXTERITY DIAGNOSTIC CATHETER WAS ATTEMPTED TO BE USED. IT WAS REPORTED THAT A DAMPING EFFECT WAS OBSERVED AFTER THE CONTRAST INJECTION DESPITE BEEN IN THE AORTA.
Patient problems: Injury
MAUDE report 10281732 ↗InjuryJul 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A DXTERITY DIAGNOSIS CATHETER WAS USED DURING A PROCEDURE TO TREAT A NON TORTUOSITY, NONE CALCIFIED LESION IN THE MID ARTERY. THERE WAS NO DAMAGE NOTED TO THE PACKAGING. THE DEVICE WAS REMOVED FROM THE PACKAGING PER IFU WITH NO ISSUES. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES. RESISTANCE WAS NOT ENCOUNTERED AND EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT A DAMPING EFFECT WAS OBSERVED AFTER THE CONTRAST INJECTION DESPITE BEEN IN THE AORTA. THE PATIENT IS ALIVE WITH NO INJURY.
Patient problems: Injury; Occlusion
MAUDE report 10256712 ↗InjuryJul 29, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE A DXTERITY DIAGNOSTIC ANGIOGRAPHY CATHETER TO TREAT A MILDLY TORTUOUS, MILDLY CALCIFIED LESION EXHIBITING 100% CHRONIC TOTAL OCCLUSION IN THE RIGHT CORONARY ARTERY (RCA). THERE WAS NO DAMAGE NOTED TO THE PACKAGING. THE DEVICE WAS REMOVED FROM PACKAGING AS PER IFU WITH NO ISSUES. THE DEVICE WAS INSPECTED WITH NO ISSUES. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE AND EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT A DISSECTION OCCURRED DURING THE PROCEDURE. IT WAS INDICATED THAT THE CURVE OF THE DEVICE MAY HAVE CONTRIBUTED TO THIS EVENT.
Patient problems: Vascular Dissection
MAUDE report 8835192 ↗InjuryMay 24, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A DXTERITY DIAGNOSTIC CATHETER WAS USED DURING A NORMAL HEART CATH PROCEDURE. NO DAMAGE WAS NOTED TO THE PACKAGING. THE DEVICE WAS REMOVED FROM PACKAGING PER IFU AND INSPECTED WITH NO ISSUES NOTED. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES NOTED. NO ALLEGED PRODUCT MALFUNCTION IS REPORTED. THE VESSEL ASSOCIATED WITH THE EVENT HAD MILD CALCIFICATION AND MILD TORTUOUSITY. AFTER THE PROCEDURE THE PATIENT BECAME LETHARGIC. MRI SHOWED SHOWED STROKE FROM RIGHT CAROTID SIDE WHICH IS THE SIDE THAT THE CATHETER WAS INTRODUCED. PATIENT FULLY RECOVERED.
Patient problems: Stroke/CVA
MAUDE report 7540197 ↗InjurySep 8, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
THE PHYSICIAN WAS USING A DEXTERITY DIAGNOSTIC CATHETER DURING A PROCEDURE. NO DAMAGE WAS NOTED TO THE CATHETER PACKAGING. THE DEVICE WAS REMOVED FROM PACKAGING PER IFU WITH NO ISSUES NOTED. THE DEVICE WAS INSPECTED AND PREPPED PER IFU. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE, EXCESSIVE FORCE WAS NOT USED. THE CATHETER WAS USED IN THE PROXIMAL RCA, WHICH EXHIBITED SEVERE TORTUOSITY AND WAS NOT CALCIFIED. DURING USE OF THE DEXTERITY A SPIRAL DISSECTION OF THE RCA OCCURRED. THERE WAS NO LESION IN THE LOCATION OF THE DISSECTION. THE PATIENT WAS SENT FOR BYPASS SURGERY AS THE PHYSICIAN WAS UNABLE TO GET WIRE THROUGH THE DISSECTION TO TREAT THIS AE. IT IS REPORTED THAT PATIENT IS DOING FINE POST PROCEDURE.
Patient problems: Intimal Dissection; Patient Problem/Medical Problem
MAUDE report 6851351 ↗InjurySep 8, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PHYSICIAN WAS USING A DXTERITY DIAGNOSTIC CATHETER WHEN TREATING THE PROXIMAL LAD, MODERATE TORTUOSITY, NON-CALCIFIED, NO STENOSIS. NO DAMAGE WAS NOTED TO THE DEVICE PACKAGING. NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE PACKAGING. THE DEVICE WAS INSPECTED AND PREPPED PER IFU WITH NO ISSUES. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE, EXCESSIVE FORCE WAS NOT APPLIED. A DISSECTION OCCURRED NEAR THE LOCATION OF THE TIP OF THE CATHETER AND WAS TREATED WITH A RESOLUTE ONYX STENT. IT IS REPORTED THAT THE PHYSICIAN DOES NOT THINK THAT THE CATHETER CAUSED THE DISSECTION, BASED ON THE PATIENT'S CLINICAL SYMPTOMS, THIS DISSECTION MAY HAVE ALREADY BEEN THERE. NO OTHER DEVICE WAS IN THIS VESSEL. NO ADDITIONAL PATIENT CLINICAL SEQUELAE REPORTED.
Patient problems: Intimal Dissection; Patient Problem/Medical Problem
MAUDE report 6851125 ↗InjuryJun 8, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PHYSICIAN USED A DXTERITY GUIDE CATHETER WHEN TREATING THE RCA (OSTIUM/ PROXIMAL/ MID). NO DAMAGE WAS NOTED TO THE PACKAGING OF THE CATHETER. THE CATHETER WAS REMOVED FROM PACKAGING PER IFU AND INSPECTED WITH NO ISSUES NOTED. THE CATHETER WAS PREPPED WITH NO ISSUES NOTED. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE GC, EXCESSIVE FORCE WAS NOT USED. THE RCA HAD NO CALCIFICATION, NO TORTUOSITY. WHEN THE PHYSICIAN WENT TO ENGAGE THE CATHETER, A SPIRAL DISSECTION OCCURRED IN THE OSTIUM TO MID VESSEL , CAUSED BY THE DIAGNOSTIC CATHETER WHEN IT ENGAGED THE SLIGHTLY ANTERIOR TAKE OFF OF THE RCA. THE DISSECTION WAS TREATED WITH TWO DRUG-ELUTING STENTS. NO SPECIFIC MALFUNCTION REPORTED BUT THE PHYSICIAN SAID THE STIFFNESS OF THE TIP OF THE DXTERITY IS MORE THAN LIKELY THE CAUSE. THERE IS NO ALLEGED PRODUCT ISSUE. NO ADDITIONAL PATIENT CLINICAL SEQUELAE REPORTED.
Patient problems: Intimal Dissection; Patient Problem/Medical Problem
MAUDE report 6621801 ↗InjuryApr 21, 2017Material Deformation; Adverse Event Without Identified Device or Use Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem
THE PHYSICIAN INTENDED TO USE A DXTERITY DIAGNOSTIC CATHETER. THE DEVICE WAS REMOVED FROM PACKAGING PER IFU AND INSPECTED. THE PHYSICIAN NOTICED OVALIZATION OF THE TIP PRIOR TO INSERTION IN PATIENT. THE INTENDED LESION WAS IN THE PROXIMAL RCA, NON-TORTUOUS, SEVERE CALCIFICATION AND (B)(6) STENOSIS. THE VESSEL HAD CALCIFIED ECCENTRIC PLAQUE. A DISSECTION OCCURRED DURING THE PROCEDURE. THE PHYSICIAN WAS NOT SURE IF THE DXTERITY CATHETER CAUSED THE DISSECTION, OR IF IT WAS JUST ECCENTRIC PLAQUE. THE VESSEL WAS IVUSED TO LOOK AT THE POSSIBLE DISSECTION BUT THE DETERMINATION WAS UNCLEAR. THE LESION WAS FFR AT (B)(6) NEGATIVE. WHEN THE POSSIBLE DISSECTION AFTER DIAGNOSTIC CATH WAS NOTED ON ANGIO REVIEW, AND WITH THE UNCERTAIN IVUS RESULT, THE DECISION WAS MADE TO STENT THE DISSECTION. NO ADDITIONAL PATIENT CLINICAL SEQUELAE REPORTED.
Patient problems: Intimal Dissection
MAUDE report 6511096 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page