Atellica IM CKMB Master Curve Material, Cat No. 10995534 recallZ-0117-2019 · Siemens Healthcare Diagnostics, Inc · initiated Jul 24, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A potential bias was identified with certain kits compared to the internal standards.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On July 24, 2018 Siemens Healthineers issued Urgent Medical Device Recall notices to U.S. customers and Urgent Field Safety Notices to OUS customers. The following are Actions to be Taken by the Customer: Customers using ADVIA Centaur CKMB Calibrator Lots ending in 64: - Review the notice with your Medical Director. - Review your inventory of these products and immediately order replacement ADVIA Centaur CKMB Calibrator kit lots ending in 68 by completing the Field Correction Effectiveness Check Form attached to the letter. - You may continue use of ADVIA Centaur CKMB Calibrator kit lots until replacement product is received in your laboratory. - Upon receipt of replacement product in your laboratory, discontinue use of and discard the ADVIA Centaur CKMB Calibrator kit lots listed in Table 1 of the letter. - ADVIA Centaur CKMB Calibrator kit lots ending in 63 may continue to be used until the assigned expiration date. - Refer to the Additional Information section for expected patient sample bias when transitioning to ADVIA Centaur CKMB Calibrator kit lots ending in 68. Customers using ADVIA Centaur Master Curve Material: - Discontinue use of and discard the ADVIA Centaur MCM lots in Table 1. - Review your inventory of these products and order replacement ADVIA Centaur CKMB Master Curve Material by completing the Field Correction Effectiveness Check Form attached to the letter. - Utilize ADVIA Centaur CKMB MCM Lot 13384 and future lots. - Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. Customers using Atellica IM CKMB Calibrator Lots ending in 66: - Review your inventory of these products and immediately order replacement Atellica IM CKMB Calibrator kit lots ending in 69 by completing the Field Correction Effectiveness Check Form attached to the letter. - You may continue use of Atellica IM CKMB Calibrator kit lots in Table 1 until replacement product is received in your laboratory
Recall details
- Recall number
- Z-0117-2019
- Status
- Terminated
- Initiated by the company
- Jul 24, 2018
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 2 sets
- Product code
- JHS
- FDA submissions
- K962126
- Distribution
- US Nationwide Distribution
Product description
Atellica IM CKMB Master Curve Material, Cat No. 10995534 Product Usage: For in vitro diagnostic use for evaluating the ADVIA Centaur CKMB assay. This material is intended to be run singly as unknown samples after a two-point calibration has been performed on the system.
Code information (models, lots, serial numbers)
SMN 10995534 Lot # Expiry UDI # 11803 2019-02-10 (01)00630414598611(10)11803(17)20190210
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica IM CKMB MCM 10995534Siemens Healthcare Diagnostics Inc. · Single (specified) analyte controls (assayed and unassayed)1 recall0 reports
Other devices under the same FDA submission
5 devices cleared under K962126 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVIA Centaur® CKMB MCM 10309782Siemens Healthcare Diagnostics Inc. · RADIOASSAY, VITAMIN B121 recall0 reports
- ADVIA Centaur® Systems, ACS:180® Systems CKMB Calibrator 10311570Siemens Healthcare Diagnostics Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES1 recall0 reports
- Atellica IM CKMB Calibrator 10995532Siemens Healthcare Diagnostics Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES1 recall0 reports
- Atellica IM CKMB 10995530Siemens Healthcare Diagnostics Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES2 reports
- Atellica IM CKMB 10995531Siemens Healthcare Diagnostics Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80954).
- Z-0113-2019 · TerminatedADVIA Centaur CKMB Calibrator, Cat No. 09318028 Product Usage: For in vitro diagnostic use in calibrating ADVIA Centaur systems CKMB assays.
- Z-0114-2019 · TerminatedADVIA Centaur CKMB Calibrator 2-pack Japan, Cat No. 02562748 Product Usage: For in vitro diagnostic use in calibrating ADVIA Centaur systems CKMB assays.
- Z-0115-2019 · TerminatedADVIA Centaur CKMB Master Curve Material, Cat No. 07867768 Product Usage: For in vitro diagnostic use for evaluating the ADVIA Centaur CKMB assay. This material is intended to be…
- Z-0116-2019 · TerminatedAtellica IM CKMB Calibrator, Cat No. 10995532 Product Usage: For in vitro diagnostic use for evaluating the ADVIA Centaur CKMB assay. This material is intended to be run singly a…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page