Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745… recallZ-0108-2019 · Carl Zeiss Meditec AG · initiated Jun 25, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Carl Zeiss Meditec AG or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
FDA root cause category: Employee error
What the company is doing
On 08/06/2018 the firm sent a Field Safety Corrective Action letter to customers via FedEx Priority Overnight service informing them that based on routine post market surveillance activities, we have information that this specific IOLMaster 700 device is incorrectly calibrated. The calibration value for axial length measurement was not entered correctly in the device settings during the manufacturing process. The transmission discrepancy leads to an error during the axial length measurement of approx. 0.25mm. A deviation in the axial length of this magnitude results in an error of Single unit. approx. 0.8 Diopter with regard to the intraocular lens. This in turn results in a refractive error of about 0.5 Diopter. The local firm's service organization will remove this device from the field. The device manufacturer will recalibrate the device to ensure it meets specifications. Hazard involved: The likelihood of a faulty entered calibration value for axial length measurement in the production process of IOLMaster 700 is considered rare. The probability of injury, i.e. reduced visual acuity due to the implantation of a wrong IOL power, is very likely for the affected device, because the axial length of the eye cannot be measured correctly. The device failure may only be recognized through post-surgical examination. Affected product: Quantity: 1 Product Name: IOLMaster 700 Product Number: 000000-1932-169 Device Serial Number: 1185393 Manufacturing Date: 2017-09-18 Actions to be taken: 1. Discontinue use of the ZEISS IOLMaster 700 device (Serial Number 1185393). Physicians are asked to review refractive results during usage of this device and to inform the firm about any hyperopic refractive surprises noticed. Any measurement and calculation data from this device should not be taken into consideration for further treatments. 2. Complete and return the attached Medical Device Recall Return Response Form to [email protected] no later than Jun
Recall details
- Recall number
- Z-0108-2019
- Status
- Terminated
- Initiated by the company
- Jun 25, 2018
- Recalling firm
- Carl Zeiss Meditec AG
- Quantity
- 1
- Product code
- HJO
- FDA submissions
- K170171
- Distribution
- US: CA (Mountain View and San Jose) A single system. No distributed OUS.
Product description
Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 1185393 2017-09-01
Code information (models, lots, serial numbers)
Model: 700 Catalog Number: 000000-1932-169 Serial Number: 1185393 UDI: 04049471092080
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IOLMaster 700Carl Zeiss Meditec AG · Biomicroscope, Slit-Lamp, Ac-Powered1 recall50 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page