IOLMaster 700Recalls & reported problems

Carl Zeiss Meditec AG · The IOLMaster 700 is intended for biometric measurements and visualization of ocular structures. The measurements and visualization assist in the determination of the appropriate power and type of intraocular lens. The IOLMaster 700 measures: • Lens thickness • Corneal curvature and thickness • Axial length • Anterior chamber depth • Pupil diameter • White-to-white distance (WTW) For visualization, the IOLMaster 700 employs optical coherence tomography (OCT) to obtain two-dimensional images of ocular structures of the anterior and posterior segments of the eye. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

The Carl Zeiss Meditec IOLMaster 700 is an eye measuring machine used before cataract surgery to help choose the right artificial lens. Most of the 50 reports describe measurements coming out wrong or off, sometimes leading to a lens implant that left patients needing a second operation, along with problems assembling or maintaining the unit. 43 reports involved injuries and no deaths were reported. One recall, started in June 2018 and closed, involved a calibration setting entered incorrectly during manufacturing that produced incorrect eye images.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
43
Recalls naming it
1

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 04049471092080), received between Jun 19, 2018 and Mar 13, 2026 — classified as 43 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (22 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 13 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 7 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2018Terminated

    Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.

    Z-0108-2019 · Root cause: Employee error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Use of Device Problem22
  • Incorrect, Inadequate or Imprecise Result or Readings11
  • Adverse Event Without Identified Device or Use Problem9
  • Misassembly During Maintenance/Repair9
  • Incorrect Measurement4
  • Insufficient Device Problem Information3
  • High Readings2
  • Calibration Problem1

Patient problems

  • Visual Impairment40
  • No Clinical Signs, Symptoms or Conditions4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Insufficient Information2
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 13, 2026Insufficient Device Problem Information

    THE CUSTOMER REPORTED THAT THEIR IOLMASTER 700 IS MEASURING THE CYLINDER (CYL) VALUE 1.5 DIOPTERS HIGHER THAN EXPECTED. A LENS REIMPLANTATION WAS PERFORMED FOR ONE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 24591210 ↗
  • InjuryJan 15, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IOLMASTER 700 DEVICE MADE INCORRECT MEASUREMENTS FOR A PATIENT. THE PATIENT HAD 5 MEASUREMENT SESSIONS ON THE DEVICE. THE USER IMPLANTED AN IOL (INTRAOCULAR LENS) BASED ON MEASUREMENTS FROM (B)(6) 2025 WHICH RESULTED IN A HIGHLY HYPEROPIC ASTIGMATIC OUTCOME. THE USER IS PLANNING AN IOL EXCHANGE FOR THE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 24087315 ↗
  • InjuryDec 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair

    IT WAS REPORTED THAT A PATIENT IN IRELAND HAS HAD A RESULT OF -2 DIOPTERS SURGICAL OUTCOME FROM THE EXPECTED AND PLANNED SURGERY GOAL. IT IS VERY LIKELY THAT AN ADDITIONAL RE-OP (2ND SURGERY) WILL BE EXECUTED IN ORDER TO ACHIEVE THE PLANNED GOAL FOR THE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 23843952 ↗
  • InjuryDec 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair

    IT WAS REPORTED THAT A PATIENT IN IRELAND HAS HAD A RESULT OF -2 DIOPTERS SURGICAL OUTCOME FROM THE EXPECTED AND PLANNED SURGERY GOAL. IT IS VERY LIKELY THAT AN ADDITIONAL RE-OP (2ND SURGERY) WILL BE EXECUTED IN ORDER TO ACHIEVE THE PLANNED GOAL FOR THE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 23843711 ↗
  • InjuryDec 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair

    IT WAS REPORTED THAT A PATIENT IN IRELAND HAS HAD A RESULT OF -2 DIOPTERS SURGICAL OUTCOME FROM THE EXPECTED AND PLANNED SURGERY GOAL. IT IS VERY LIKELY THAT AN ADDITIONAL RE-OP (2ND SURGERY) WILL BE EXECUTED IN ORDER TO ACHIEVE THE PLANNED GOAL FOR THE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 23843458 ↗
  • InjuryDec 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair

    IT WAS REPORTED THAT A PATIENT IN IRELAND HAS HAD A RESULT OF -2 DIOPTERS SURGICAL OUTCOME FROM THE EXPECTED AND PLANNED SURGERY GOAL. IT IS VERY LIKELY THAT AN ADDITIONAL RE-OP (2ND SURGERY) WILL BE EXECUTED IN ORDER TO ACHIEVE THE PLANNED GOAL FOR THE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 23842830 ↗
  • InjuryDec 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair

    IT WAS REPORTED THAT A PATIENT IN IRELAND HAS HAD A RESULT OF -2 DIOPTERS SURGICAL OUTCOME FROM THE EXPECTED AND PLANNED SURGERY GOAL. IT IS VERY LIKELY THAT AN ADDITIONAL RE-OP (2ND SURGERY) WILL BE EXECUTED IN ORDER TO ACHIEVE THE PLANNED GOAL FOR THE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 23840714 ↗
  • InjuryDec 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair

    IT WAS REPORTED THAT A PATIENT IN IRELAND HAS HAD A RESULT OF -2 DIOPTERS SURGICAL OUTCOME FROM THE EXPECTED AND PLANNED SURGERY GOAL. IT IS VERY LIKELY THAT AN ADDITIONAL RE-OP (2ND SURGERY) WILL BE EXECUTED IN ORDER TO ACHIEVE THE PLANNED GOAL FOR THE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 23840374 ↗
  • InjuryDec 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair

    IT WAS REPORTED THAT A PATIENT IN IRELAND HAS HAD A RESULT OF -2 DIOPTERS SURGICAL OUTCOME FROM THE EXPECTED AND PLANNED SURGERY GOAL. IT IS VERY LIKELY THAT AN ADDITIONAL RE-OP (2ND SURGERY) WILL BE EXECUTED IN ORDER TO ACHIEVE THE PLANNED GOAL FOR THE PATIENT.

    Patient problems: Visual Impairment

    MAUDE report 23840145 ↗
  • MalfunctionNov 10, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER FROM BRAZIL COMPLAINED ABOUT IOL (INTRAOCULAR LENS) CALCULATIONS ON AN IOLMASTER 700 HAVING A DISCREPANCY OF APPROXIMATELY 2 DIOPTERS COMPARED TO ANOTHER BRAND'S DEVICE. A ZEISS SERVICE TECHNICIAN WENT ONSITE AND DURING INVESTIGATION OF THE DEVICE, INCORRECT AXIAL LENGTH MEASUREMENTS WERE FOUND. THERE IS NO INFORMATION ABOUT ANY NEGATIVE IMPACT ON A PATIENT (LIKE LENS RE-IMPLANTATION). IT IS ALSO NOT KNOWN WHETHER THE MEASUREMENT DATA FROM THE IOLMASTER 700 WAS USED FOR ANY IOL CALCULATION AND SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 23518874 ↗
  • MalfunctionNov 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Low Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Low Readings; Misassembly During Maintenance/Repair

    A ZEISS FIELD SERVICE ENGINEER (FSE) WAS PERFORMING A CMOS BATTERY REPLACEMENT ON AN IOL MASTER 700 AT A CUSTOMER SITE BECAUSE THE DEVICE WAS REPORTED TO NOT BE POWERING ON ANYMORE. DURING THE CHECKOUT, HE NOTICED A SIGNIFICANT DISCREPANCY IN THE AXIAL LENGTH MEASUREMENTS, WITH THE TEST EYE SHOWING 28.296 COMPARED TO THE REFERENCE VALUE OF 28.039, EXCEEDING THE ACCEPTABLE THRESHOLD OF 0.02. THE DEVICE HAD BECOME UNCALIBRATED. THE DEVICE WAS TAKEN OUT OF SERVICE UNTIL IT WAS REPAIRED AND BROUGHT BACK INTO ZEISS SPECIFICATIONS. THERE WAS NO NEGATIVE IMPACT ON A PATIENT (LIKE LENS RE-IMPLANTATION). CUSTOMER CONFIRMED THERE WERE NO RE-OPERATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23471140 ↗
  • MalfunctionSep 9, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Misassembly During Maintenance/Repair

    DURING A SYSTEM CHECKOUT BEING PERFORMED ON AN IOLMASTER 700 DEMONSTRATION DEVICE IN SOUTH AFRICA, A ZEISS FIELD SERVICE ENGINEER (FSE) FOUND THAT THE OCT LENGTH WAS OUT OF TOLERANCE. SECOND LEVEL SUPPORT WAS CONTACTED AND THE DEVICE WAS TAKEN OUT OF SERVICE. IT IS NOT KNOWN WHETHER THE DEVICE WAS BEING USED FOR MEASURING PATIENTS. THERE IS NO INFORMATION THAT INCORRECT MEASUREMENT VALUES WERE USED FOR THE CALCULATION AND IMPLANTATION OF INTRAOCULAR LENSES. THERE IS ALSO NO INFORMATION ABOUT ANY NEGATIVE IMPACT TO A PATIENT (LIKE LENS RE-IMPLANTATION).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23004430 ↗
  • InjuryJun 4, 2025Use of Device Problem; Use of Device Problem

    CUSTOMER REPORTED POST OP SURPRISE FOR THEIR PATIENT THAT HAD THE ALCON VIVITY LENS. THEREFORE, A RE-OP WAS PERFORMED.

    Patient problems: Visual Impairment

    MAUDE report 22144015 ↗
  • InjuryJun 4, 2025Use of Device Problem; Use of Device Problem

    CUSTOMER REPORTED POST OP SURPRISE FOR THEIR PATIENT THAT HAD THE ALCON VIVITY LENS. THEREFORE, A RE-OP LASIK WAS PERFORMED.

    Patient problems: Visual Impairment

    MAUDE report 22144014 ↗
  • InjuryMar 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTH CARE PROFESSIONAL REPORTED MEASUREMENT DISCREPANCIES AND REFRACTIVE SURPRISE ON A PATIENT. THE DOCTOR WILL BE DOING IOL EXCHANGE TO CORRECT PATIENT'S VISION.

    Patient problems: Visual Impairment

    MAUDE report 21654404 ↗
  • InjuryMar 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTH CARE PROFESSIONAL REPORTED MEASUREMENT DISCREPANCIES AND REFRACTIVE SURPRISE ON A PATIENT. THE DOCTOR IMPLANTED A SECOND IOL (PIGGYBACK) TO CORRECT PATIENT'S VISION.

    Patient problems: Visual Impairment

    MAUDE report 21654390 ↗
  • InjuryMar 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTH CARE PROFESSIONAL REPORTED MEASUREMENT DISCREPANCIES AND REFRACTIVE SURPRISE ON A PATIENT. THE DOCTOR WILL BE DOING IOL EXCHANGE TO CORRECT PATIENT'S VISION.

    Patient problems: Visual Impairment

    MAUDE report 21654214 ↗
  • InjuryMar 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTH CARE PROFESSIONAL REPORTED MEASUREMENT DISCREPANCIES AND REFRACTIVE SURPRISE ON A PATIENT. THE DOCTOR WILL BE DOING IOL EXCHANGE TO CORRECT PATIENT'S VISION.

    Patient problems: Visual Impairment

    MAUDE report 21654198 ↗
  • MalfunctionNov 8, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Calibration Problem; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Calibration Problem

    CUSTOMER IN THAILAND REPORTED THAT THE IOLMASTER 700 SEEMS TO HAVE UNSTABLE VALUES. THE VALUES DID NOT MATCH THE VALUES TAKEN FROM OTHER HOSPITALS IN THE SAME PATIENT CASE. A DEMO MACHINE WAS BROUGHT IN BY ZEISS SERVICE TO COMPARE AND FOUND THAT THE CALCULATION RESULTS WERE DIFFERENT ABOUT 1 DIOPTER. THE MEASURED AXIAL LENGTH VALUE IS LONGER BY AN AVERAGE OF ABOUT 0.23-0.27 MM COMPARED TO THE DEMO UNIT AND OTHER UNITS IN THE HOSPITAL. THIS WILL CAUSE THE CALCULATION OF THE ARTIFICIAL LENS VALUE TO BE INCORRECT BY ABOUT 0.5-1.0 D. PATIENT DATA WAS REQUESTED. THERE IS NO INFORMATION ABOUT NEGATIVE IMPACT TO A PATIENT (LENS RE-IMPLANTATION).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20646149 ↗
  • InjurySep 11, 2024Use of Device Problem; Use of Device Problem

    THE CUSTOMER REPORTED THAT THE POST-OP REFRACTION WAS FOUND TO BE OFF. MEDICAL INTERVENTION WAS NEEDED TO ADDRESS THE ISSUE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Visual Impairment

    MAUDE report 20195051 ↗
  • InjurySep 11, 2024Use of Device Problem; Use of Device Problem

    THE CUSTOMER REPORTED THAT THE POST-OP REFRACTION WAS FOUND TO BE OFF. MEDICAL INTERVENTION WAS NEEDED TO ADDRESS THE ISSUE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Visual Impairment

    MAUDE report 20194405 ↗
  • InjurySep 11, 2024Use of Device Problem; Use of Device Problem

    THE CUSTOMER REPORTED THAT THE POST-OP REFRACTION WAS FOUND TO BE OFF. MEDICAL INTERVENTION WAS NEEDED TO ADDRESS THE ISSUE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Visual Impairment

    MAUDE report 20194295 ↗
  • InjuryJul 12, 2024Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A DOCTOR PERFORMED SURGERY WITH CLEAR LENS AND THE DEVIATION IN SURGERY RESULT WAS +1.5 DIOPTER. THE DOCTOR HAD TO EXCHANGE THE LENS TO CORRECT PATIENT'S VISION. THE DOCTOR COMMUNICATED THAT THE TEST EYE CHECK PASSED SUCCESSFULLY.

    Patient problems: Visual Impairment

    MAUDE report 19735308 ↗
  • InjuryJul 12, 2024Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A DOCTOR PERFORMED SURGERY WITH CLEAR LENS AND THE DEVIATION IN SURGERY RESULT WAS +1.5 DIOPTER. THE DOCTOR HAD TO EXCHANGE THE LENS TO CORRECT PATIENT'S VISION. THE DOCTOR COMMUNICATED THAT THE TEST EYE CHECK PASSED SUCCESSFULLY.

    Patient problems: Visual Impairment

    MAUDE report 19735265 ↗
  • InjuryJun 7, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE PATIENT UNFORTUNATELY HAD A POOR OUTCOME. MEASUREMENTS ARE SIGNIFICANTLY DIFFERENT PRE AND POST OP. A LENS REPOSITIONING WAS PERFORMED.

    Patient problems: Visual Impairment

    MAUDE report 19484842 ↗

Other Biomicroscope, Slit-Lamp, Ac-Powered devices