Xper Flex Cardio Physiomonitoring System recallZ-0462-2019 · Invivo Corporation · initiated Sep 24, 2018
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Invivo Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard
FDA root cause category: Software Design Change
What the company is doing
Customers were notified on about 09/20/2018 via an "URGENT - Medical Device Correction" letter. An initial letter was sent to distributors with instructions to provide the Field Safety Notice to all customers the devices were further distributed to. The "URGENT - Medical Device Correction" letter provided guidance on how to identify the Service Number and serial number on the Flex Cardio device, which is on the bottom right corner of the back of the device. Actions to be taken by the customer/user until Philips contacts the customer to arrange for a software update to address this issue, is if the IBP numeric values become frozen and do not update in real-time, opening the Waveform Setup Dialog and then immediately pressing the OK button will restore the proper display of the IBP numeric values. Alternative methods of performing the functions listed in the Problem Description section that do not cause the problem to occur include the following: 1. Change an IBP Site Label using the Pressure Setup Dialog or a macro; 2. Activate or deactivate the mean pressure waveform display option (On/Off) by using the Waveform Setup Dialog, the Pressure Setup Dialog or a macro; 3. Activate or deactivate the waveform hide option (On/Off) using the Waveform Setup Dialog, the Pressure Setup Dialog, the keyboard or a macro; 4. Activate or deactivate the pressure waveform filter option (On/Off) using a macro. A Philips representative will contact each customer regarding their affected Flex Cardio devices. A software update will be required on both the affected Flex Cardio device and on the host workstations driver. These actions will be implemented free of charge by Philips.
Recall details
- Recall number
- Z-0462-2019
- Status
- Completed
- Initiated by the company
- Sep 24, 2018
- Recalling firm
- Invivo Corporation
- Quantity
- 76 units
- Product code
- DQK
- FDA submissions
- K101571
- Distribution
- Distributed to the Netherlands
Product description
Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Model Number 453564669081; System Serial Number (UDI): US1814000001 ((01)00884838091474(21)US1814000001), US1814000002 ((01)00884838091474(21)US1814000002), US1814000003 ((01)00884838091474(21)US1814000003), US1814000004 ((01)00884838091474(21)US1814000004), US1814000006 ((01)00884838091474(21)US1814000006), US1814000007 ((01)00884838091474(21)US1814000007), US1814000008 ((01)00884838091474(21)US1814000008), US1814000011 ((01)00884838091474(21)US1814000011), US1814000012 ((01)00884838091474(21)US1814000012), US1814000013 ((01)00884838091474(21)US1814000013), US1814000014 ((01)00884838091474(21)US1814000014), US1814000016 ((01)00884838091474(21)US1814000016), US1814000017 ((01)00884838091474(21)US1814000017), US1814000018 ((01)00884838091474(21)US1814000018), US1814000019 ((01)00884838091474(21)US1814000019), US1814000020 ((01)00884838091474(21)US1814000020), US1814000022 ((01)00884838091474(21)US1814000022), US1814000023 ((01)00884838091474(21)US1814000023), US1814000024 ((01)00884838091474(21)US1814000024), US1814000027 ((01)00884838091474(21)US1814000027), US1814000028 ((01)00884838091474(21)US1814000028), US1814000029 ((01)00884838091474(21)US1814000029), US1814000030 ((01)00884838091474(21)US1814000030), US1814000031 ((01)00884838091474(21)US1814000031), US1814000032 ((01)00884838091474(21)US1814000032), US1814000033 ((01)00884838091474(21)US1814000033), US1814000034 ((01)00884838091474(21)US1814000034), US1814000035 ((01)00884838091474(21)US1814000035), US1814000036 ((01)00884838091474(21)US1814000036), US1814000037 ((01)00884838091474(21)US1814000037), US1814000038 ((01)00884838091474(21)US1814000038), US1814000039 ((01)00884838091474(21)US1814000039), US1814000040 ((01)00884838091474(21)US1814000040), US1814000041 ((01)00884838091474(21)US1814000041), US1814000043 ((01)00884838091474(21)US1814000043), US1814000044 ((01)00884838091474(21)US1814000044), US1814000045 ((01)00884838091474(21)US1814000045), US1814000046 ((01)00884838091474(21)US1814000046), US1814000047 ((01)00884838091474(21)US1814000047), US1814000048 ((01)00884838091474(21)US1814000048), US1814000049 ((01)00884838091474(21)US1814000049), US1814000050 ((01)00884838091474(21)US1814000050), US1814000051 ((01)00884838091474(21)US1814000051), US1814000052 ((01)00884838091474(21)US1814000052), US1814000053 ((01)00884838091474(21)US1814000053), US1814000054 ((01)00884838091474(21)US1814000054), US1814000055 ((01)00884838091474(21)US1814000055), US1814000056 ((01)00884838091474(21)US1814000056), US1814000058 ((01)00884838091474(21)US1814000058), US1814000059 ((01)00884838091474(21)US1814000059), US1814000060 ((01)00884838091474(21)US1814000060), US1814000061 ((01)00884838091474(21)US1814000061), US1814000062 ((01)00884838091474(21)US1814000062), US1814000063 ((01)00884838091474(21)US1814000063), US1814000064 ((01)00884838091474(21)US1814000064), US1814000065 ((01)00884838091474(21)US1814000065), US1814000066 ((01)00884838091474(21)US1814000066), US1814000067 ((01)00884838091474(21)US1814000067), US1814000068 ((01)00884838091474(21)US1814000068), US1814000069 ((01)00884838091474(21)US1814000069), US1814000070 ((01)00884838091474(21)US1814000070), US1814000072 ((01)00884838091474(21)US1814000072), US1814000074 ((01)00884838091474(21)US1814000074), US1814000078 ((01)00884838091474(21)US1814000078), US1814000081 ((01)00884838091474(21)US1814000081), US1814000082 ((01)00884838091474(21)US1814000082), US1814000083 ((01)00884838091474(21)US1814000083), US1814000084 ((01)00884838091474(21)US1814000084), US1814000085 ((01)00884838091474(21)US1814000085), US1814000086 ((01)00884838091474(21)US1814000086), US1814000087 ((01)00884838091474(21)US1814000087), US1814000088 ((01)00884838091474(21)US1814000088), US1814000089 ((01)00884838091474(21)US1814000089), US1814000092 ((01)00884838091474(21)US1814000092), US1814000094 ((01)00884838091474(21)US1814000094), US1814000096 ((01)00884838091474(21)US1814000096).
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Xper Flex Cardio Physiomonitoring System Xper FC2010 Rev D EXCHANGEPhilips Ds North America LLC · Monitor, physiological, patient (without arrhythmia detection or alarms)1 recall2 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗