Xper Flex Cardio Physiomonitoring System recallZ-0462-2019 · Invivo Corporation · initiated Sep 24, 2018

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Invivo Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

FDA root cause category: Software Design Change

What the company is doing

Customers were notified on about 09/20/2018 via an "URGENT - Medical Device Correction" letter. An initial letter was sent to distributors with instructions to provide the Field Safety Notice to all customers the devices were further distributed to. The "URGENT - Medical Device Correction" letter provided guidance on how to identify the Service Number and serial number on the Flex Cardio device, which is on the bottom right corner of the back of the device. Actions to be taken by the customer/user until Philips contacts the customer to arrange for a software update to address this issue, is if the IBP numeric values become frozen and do not update in real-time, opening the Waveform Setup Dialog and then immediately pressing the OK button will restore the proper display of the IBP numeric values. Alternative methods of performing the functions listed in the Problem Description section that do not cause the problem to occur include the following: 1. Change an IBP Site Label using the Pressure Setup Dialog or a macro; 2. Activate or deactivate the mean pressure waveform display option (On/Off) by using the Waveform Setup Dialog, the Pressure Setup Dialog or a macro; 3. Activate or deactivate the waveform hide option (On/Off) using the Waveform Setup Dialog, the Pressure Setup Dialog, the keyboard or a macro; 4. Activate or deactivate the pressure waveform filter option (On/Off) using a macro. A Philips representative will contact each customer regarding their affected Flex Cardio devices. A software update will be required on both the affected Flex Cardio device and on the host workstations driver. These actions will be implemented free of charge by Philips.

Recall details

Recall number
Z-0462-2019
Status
Completed
Initiated by the company
Sep 24, 2018
Recalling firm
Invivo Corporation
Quantity
76 units
Product code
DQK
FDA submissions
K101571
Distribution
Distributed to the Netherlands

Product description

Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

Model Number 453564669081;  System Serial Number (UDI): US1814000001 ((01)00884838091474(21)US1814000001), US1814000002 ((01)00884838091474(21)US1814000002), US1814000003 ((01)00884838091474(21)US1814000003), US1814000004 ((01)00884838091474(21)US1814000004), US1814000006 ((01)00884838091474(21)US1814000006), US1814000007 ((01)00884838091474(21)US1814000007), US1814000008 ((01)00884838091474(21)US1814000008), US1814000011 ((01)00884838091474(21)US1814000011), US1814000012 ((01)00884838091474(21)US1814000012), US1814000013 ((01)00884838091474(21)US1814000013), US1814000014 ((01)00884838091474(21)US1814000014), US1814000016 ((01)00884838091474(21)US1814000016), US1814000017 ((01)00884838091474(21)US1814000017), US1814000018 ((01)00884838091474(21)US1814000018), US1814000019 ((01)00884838091474(21)US1814000019), US1814000020 ((01)00884838091474(21)US1814000020), US1814000022 ((01)00884838091474(21)US1814000022), US1814000023 ((01)00884838091474(21)US1814000023), US1814000024 ((01)00884838091474(21)US1814000024), US1814000027 ((01)00884838091474(21)US1814000027), US1814000028 ((01)00884838091474(21)US1814000028), US1814000029 ((01)00884838091474(21)US1814000029), US1814000030 ((01)00884838091474(21)US1814000030), US1814000031 ((01)00884838091474(21)US1814000031), US1814000032 ((01)00884838091474(21)US1814000032), US1814000033 ((01)00884838091474(21)US1814000033), US1814000034 ((01)00884838091474(21)US1814000034), US1814000035 ((01)00884838091474(21)US1814000035), US1814000036 ((01)00884838091474(21)US1814000036), US1814000037 ((01)00884838091474(21)US1814000037), US1814000038 ((01)00884838091474(21)US1814000038), US1814000039 ((01)00884838091474(21)US1814000039), US1814000040 ((01)00884838091474(21)US1814000040), US1814000041 ((01)00884838091474(21)US1814000041), US1814000043 ((01)00884838091474(21)US1814000043), US1814000044 ((01)00884838091474(21)US1814000044), US1814000045 ((01)00884838091474(21)US1814000045), US1814000046 ((01)00884838091474(21)US1814000046), US1814000047 ((01)00884838091474(21)US1814000047), US1814000048 ((01)00884838091474(21)US1814000048), US1814000049 ((01)00884838091474(21)US1814000049), US1814000050 ((01)00884838091474(21)US1814000050), US1814000051 ((01)00884838091474(21)US1814000051), US1814000052 ((01)00884838091474(21)US1814000052), US1814000053 ((01)00884838091474(21)US1814000053), US1814000054 ((01)00884838091474(21)US1814000054), US1814000055 ((01)00884838091474(21)US1814000055), US1814000056 ((01)00884838091474(21)US1814000056), US1814000058 ((01)00884838091474(21)US1814000058), US1814000059 ((01)00884838091474(21)US1814000059), US1814000060 ((01)00884838091474(21)US1814000060), US1814000061 ((01)00884838091474(21)US1814000061), US1814000062 ((01)00884838091474(21)US1814000062), US1814000063 ((01)00884838091474(21)US1814000063), US1814000064 ((01)00884838091474(21)US1814000064), US1814000065 ((01)00884838091474(21)US1814000065), US1814000066 ((01)00884838091474(21)US1814000066), US1814000067 ((01)00884838091474(21)US1814000067), US1814000068 ((01)00884838091474(21)US1814000068), US1814000069 ((01)00884838091474(21)US1814000069), US1814000070 ((01)00884838091474(21)US1814000070), US1814000072 ((01)00884838091474(21)US1814000072), US1814000074 ((01)00884838091474(21)US1814000074), US1814000078 ((01)00884838091474(21)US1814000078), US1814000081 ((01)00884838091474(21)US1814000081), US1814000082 ((01)00884838091474(21)US1814000082), US1814000083 ((01)00884838091474(21)US1814000083), US1814000084 ((01)00884838091474(21)US1814000084), US1814000085 ((01)00884838091474(21)US1814000085), US1814000086 ((01)00884838091474(21)US1814000086), US1814000087 ((01)00884838091474(21)US1814000087), US1814000088 ((01)00884838091474(21)US1814000088), US1814000089 ((01)00884838091474(21)US1814000089), US1814000092 ((01)00884838091474(21)US1814000092),  US1814000094 ((01)00884838091474(21)US1814000094), US1814000096 ((01)00884838091474(21)US1814000096).