Xper Flex Cardio Physiomonitoring SystemRecalls & reported problems across all models

Philips Ds North America LLC · 19 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one Xper Flex Cardio Physiomonitoring System model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
197
Deaths reported
4
Injuries reported
15
Recalls
3

We found 19 models of Xper Flex Cardio Physiomonitoring System by Philips Ds North America LLC in the FDA's device registry (GUDID). Together they are named in 197 adverse event reports in the FDA's MAUDE database, received between Nov 13, 2018 and Sep 23, 2026 — classified as 4 deaths, 15 injuries, 178 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (31 reports).

3 FDA recall records name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Xper Flex Cardio Physiomonitoring System recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 29, 2021Open, Classified

    Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform, erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor

    Z-1517-2021 · Root cause: Software design · names 10 of these models · Recall details · FDA record ↗

  2. Sep 24, 2018Completed

    The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

    Z-0462-2019 · Root cause: Software Design Change · Recall details · FDA record ↗

  3. Mar 14, 2018Terminated

    The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.

    Z-1867-2018 · Root cause: Software design · names 6 of these models · Recall details · FDA record ↗

Most reported Xper Flex Cardio Physiomonitoring System problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings31
  • Application Program Freezes or Fails to Launch23
  • Incorrect Measurement15
  • Computer Software Problem11
  • Defibrillation/Stimulation Problem11
  • Output Problem9
  • Therapeutic or Diagnostic Output Failure8
  • Application Program Problem7

Patient problems

  • No Clinical Signs, Symptoms or Conditions113
  • No Known Impact Or Consequence To Patient42
  • Insufficient Information17
  • No Consequences Or Impact To Patient16
  • Cardiac Arrest4
  • No Patient Involvement3
  • Death2
  • Arrhythmia1

Xper Flex Cardio Physiomonitoring System models

19 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Xper Flex Cardio Physiomonitoring System adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Philips Ds North America LLC products

All Philips Ds North America LLC devices →