Spectrum Iq Infusion System With Dose Iq Safety Software… recallZ-0107-2019 · Baxter Healthcare Corporation · initiated Sep 18, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.

FDA root cause category: Software design

What the company is doing

The firm initiated their field correction by email on 09/14/2018. The notice alerted the consignee of the problem and stated that Baxter would be issuing a software update to correct the issue.

Recall details

Recall number
Z-0107-2019
Status
Terminated
Initiated by the company
Sep 18, 2018
Recalling firm
Baxter Healthcare Corporation
Quantity
354 units
Product code
PHC
FDA submissions
K173084
Distribution
NJ

Product description

SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009

Code information (models, lots, serial numbers)

UDI 00085412610900 All Serial Numbers distributed prior to 07/09/2018

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.