Why it was recalled
a portion of a single lot was not manufactured in accordance with the internal formulation resulting in inhibited growth of Gram negative organisms.
FDA root cause category: Mixed-up of materials/components
What the company is doing
Distributors were notified via email on October 22, 2018, requesting their customer distribution list. Customer letters were sent via UPS beginning on October 29, 2018. **Distributors are Advised to do the following: 1.Discard any remaining inventory of the above referenced product. 2.Complete the attached Distributor Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. 3.Within three days, provide an Excel file or a list with name, address and telephone number of end-users who were shipped this product from your distribution branches so we may inform them of this issue. Fax or email to: BD Regulatory Compliance Fax: 410-316-4258 Email: [email protected] 4.A copy of the customer communication is attached for your records. **Customers area Advised to Take the Following Action(s): 1. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 2. Discard any remaining inventory of lot 8221508. BD will process replacement of any discarded inventory upon receipt of the Customer Response Form. 3. Share a copy of the Recall Letter with all users of the product within your facility to ensure awareness. Customers with questions may call 800-638-8663, Monday through Friday 7 am - 7 pm, EST.
Recall details
- Recall number
- Z-0528-2019
- Status
- Terminated
- Initiated by the company
- Oct 22, 2018
- Posted by the FDA
- Nov 30, 2018
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 54,300 plates
- Product code
- JSG
- Distribution
- US Nationwide distribution in the states of AL, AR, AZ, CA, CT, FL, GA, HI, MA, MD, MI, MN, NC, NH, NJ, NY, OH, OK, PA, RI, TX, VA, VT, WA, WV
Product description
BD BBL Trypticase Soy Agar with 5% Sheep Blood, Catalog No. 221261, (non-selective, non-differential)
Code information (models, lots, serial numbers)
UDI - (01)10382902212611(17)181130 (10)8221508 Lot No.: 8221508 Expiration Date: 2018-11-30
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD BBL™ Trypticase™ Soy Agar with 5% Sheep Blood (TSA II) 221261Becton, Dickinson And Company · CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL1 recall51 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page