Medical device recalls: November 2018
The FDA posted 29 medical device recall records in November 2018, from 17 recall events by 15 companies. 12 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 3 events, 7 records, 4 Class I
- Baxter Healthcare Corporation · 1 event, 5 records
- Aroa Biosurgery · 1 event, 3 records
- Beckman Coulter Inc. · 1 event, 2 records, 2 Class I
- Maquet Cardiovascular Us Sales, Llc · 1 event, 2 records, 2 Class I
- Becton Dickinson & Co. · 1 event, 1 record
- Collagen Matrix, Inc. · 1 event, 1 record
- Howmedica Osteonics Corp. · 1 event, 1 record
FDA root cause
- Process control · 7
- Nonconforming Material/Component · 6
- Software design · 4
- Employee error · 4
- Other · 3
- Under Investigation by firm · 2
- Mixed-up of materials/components · 2
- Device Design · 1
Class I recalls (12)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0463-2019 · GE Healthcare, LLC · posted Nov 30, 2018 · Terminated
The ventilator Inspiratory Safety Guard (ISG) may disconnect from the breathing circuit pathway. As a result, this could create a loss of ventilation which may lead to inadequate oxygenation for patients, increasing the possibility of hypoxia.
- Z-0436-2019 · Maquet Cardiovascular Us Sales, Llc · posted Nov 28, 2018 · Terminated
There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
- Z-0437-2019 · Maquet Cardiovascular Us Sales, Llc · posted Nov 28, 2018 · Terminated
There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
- Z-0470-2019 · Synaptive Medical Inc · posted Nov 26, 2018 · Terminated
This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the…
- Z-0471-2019 · Beckman Coulter Inc. · posted Nov 20, 2018 · Terminated
A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.
- Z-0472-2019 · Beckman Coulter Inc. · posted Nov 20, 2018 · Terminated
A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.
- GE Healthcare Millennium MyoSIGHT Nuclear Cardiology Imaging System H3000ZH Acquisition System ↗Class IZ-0369-2019 · GE Healthcare, LLC · posted Nov 14, 2018 · Terminated
A detector can detach and fall.
- Z-0370-2019 · GE Healthcare, LLC · posted Nov 14, 2018 · Terminated
A detector can detach and fall.
- Z-0371-2019 · GE Healthcare, LLC · posted Nov 14, 2018 · Terminated
A detector can detach and fall.
- Z-0391-2019 · Oscor, Inc. · posted Nov 8, 2018 · Terminated
During the use of some TB - Temporary Bipolar Pacing Leads, featuring the 2mm unshrouded connectors, the connector cap housing may slide and potentially expose the connection wire. In some instances, this may cause the wire to be more susceptible to loss…
- REF VV19F Reinforced Dual Lumen Cannula, 19Fr., Sterile EO, DISPOSABLE: DO NOT RE-USE CE 0473 ↗Class IZ-0269-2019 · OriGen Biomedical, Inc. · posted Nov 7, 2018 · Terminated
Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage
- Z-0360-2019 · Roche Diagnostics Corporation · posted Nov 2, 2018 · Terminated
Abnormally high INR test results with the affected CoaguChek test strips
Class II recalls (16)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0528-2019 · Becton Dickinson & Co. · posted Nov 30, 2018 · Terminated
a portion of a single lot was not manufactured in accordance with the internal formulation resulting in inhibited growth of Gram negative organisms.
- Z-0477-2019 · The Binding Site Group, Ltd. · posted Nov 17, 2018 · Terminated
After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been…
- Z-0464-2019 · Johnson & Johnson Surgical Vision Inc · posted Nov 14, 2018 · Terminated
Expired intraocular lenses were distributed.
- Z-0445-2019 · Luminex Corporation · posted Nov 10, 2018 · Terminated
Potential to lead to a false negative results.
- Z-0411-2019 · GE Healthcare, LLC · posted Nov 9, 2018 · Terminated
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not…
- Z-0412-2019 · GE Healthcare, LLC · posted Nov 9, 2018 · Terminated
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not…
- Z-0413-2019 · GE Healthcare, LLC · posted Nov 9, 2018 · Terminated
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not…
- Z-0363-2019 · Baxter Healthcare Corporation · posted Nov 2, 2018 · Terminated
Potential ability to leak once used for compounding.
- Z-0364-2019 · Baxter Healthcare Corporation · posted Nov 2, 2018 · Terminated
Potential ability to leak once used for compounding.
- Z-0365-2019 · Baxter Healthcare Corporation · posted Nov 2, 2018 · Terminated
Potential ability to leak once used for compounding.
- Z-0366-2019 · Baxter Healthcare Corporation · posted Nov 2, 2018 · Terminated
Potential ability to leak once used for compounding.
- Z-0367-2019 · Baxter Healthcare Corporation · posted Nov 2, 2018 · Terminated
Potential ability to leak once used for compounding.
- Z-0373-2019 · Howmedica Osteonics Corp. · posted Nov 2, 2018 · Terminated
Potential for outer pouch sterile barrier to be compromised
- Z-0377-2019 · Aroa Biosurgery · posted Nov 2, 2018 · Terminated
Potential for pouch seal failure
- Z-0378-2019 · Aroa Biosurgery · posted Nov 2, 2018 · Terminated
Potential for pouch seal failure
- Z-0379-2019 · Aroa Biosurgery · posted Nov 2, 2018 · Terminated
Potential for pouch seal failure
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0414-2019 · Collagen Matrix, Inc. · posted Nov 9, 2018 · Terminated
There is a possibility that the product was packaged in the wrong box. Specifically, the product may have been placed inside a BioMend box instead of a BioMend Extend box. The only difference between the two boxes is the product trade name. The label…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.