NexGen recallZ-0630-2019 · Zimmer Biomet, Inc. · initiated Oct 31, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Screw is missing from the package
FDA root cause category: Process control
What the company is doing
Zimmer Biomet issued "URGENT MEDICAL DEVICE RECALL" Letter on 10/31/18 via email and FedEx to the consumer/user level, including any intermediate wholesale or retail consignees, distributors and Risk managers. Distributors responsibilities include locating and removing the product in their territory, as well as identifying hospitals that have previously used the product. Hospital risk managers responsibilities include the following: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].
Recall details
- Recall number
- Z-0630-2019
- Status
- Terminated
- Initiated by the company
- Oct 31, 2018
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 19 units
- Product code
- JWH
- FDA submissions
- K152494
- Distribution
- Worldwide Distribution: US (nationwide) to states of: OH, FL, NJ, NY, OH, PA, TX, WV and to countries of: CANADA, AUSTRALIA, JAPAN, GERMANY, and NETHERLANDS.
Product description
NexGen Complete Knee Solution Femoral Augment Block Distal only, 15mm, Size E with Screw Item Number: 00599003523 NexGen Distal Femoral Augment Block is used when femoral components can neither be downsized nor can be distally augmented by increasing the thickness of the cement mantle.
Code information (models, lots, serial numbers)
Lot Number: 62925336 UDI Number: (01)00889024224131(17)250228(10)62925336
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NexGen® 5990-35-23Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
Product lines
Other devices under the same FDA submission
3 devices cleared under K152494 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NexGen® 5990-34-24Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- NexGen® 5988-04-10Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- NexGen® 5990-32-21Zimmer, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page