icumedical ChemoLock Vial Spike, 20mm, REF CL-80S recallZ-0805-2019 · ICU Medical, Inc. · initiated Jan 3, 2019
Class I Terminated All recalls posted in February 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for one lot of ChemoLock Vial Spike to contain burr particulate
FDA root cause category: Vendor change control
What the company is doing
On 01/03/2019, Stericycle, Inc. sent an " URGENT: MEDICAL DEVICE RECALL" notification of the recall to ICU Medical direct consignees of record via traceable mailing. Consignees who further distributed the product were asked to notify their customers. The notification described the product, problem and actions to be taken. The consignees were instructed to do the following: 1)Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7114 (M-F, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Please visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, ICU Medical will credit you for any product returned. 4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product/ Table 1 sections of this notification and ask them to contact Stericycle at 1-888-871-7114 (M-F, 8am-Spm ET) to obtain a response form. For further inquiries, including product replacement options, please contact ICU Medical using the information provided below. Global Complaint Management at 1-844-654-7780 or email to: ([email protected]) To report adverse events or product complaints or
Recall details
- Recall number
- Z-0805-2019
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Jan 3, 2019
- Posted by the FDA
- Feb 22, 2019
- Recalling firm
- ICU Medical, Inc.
- Quantity
- 2050 units
- Product code
- ONB
- FDA submissions
- K131549
- Distribution
- US Distribution to states of: OH, FL, IA, CA, PA, IL, VA; and internationally to: AUSTRIA, CHINA (TAIWAN), IRELAND, ITALY and SINGAPORE.
Product description
icumedical ChemoLock Vial Spike, 20mm, REF CL-80S The ChemoLock Closed System Drug Transfer Device prevents the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system.
Code information (models, lots, serial numbers)
Lot Number 3757712
Other devices under the same FDA submission
91 devices cleared under K131549 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ICU Medical CL3011Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls26 reports
- ICU Medical CL3511Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls6 reports
- ICU Medical CL4136Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls1 report
- ICU Medical CL-80Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls0 reports
- ICU Medical CL3900Icu Medical, Inc. · Set, administration, intravascular3 recalls0 reports
- ICU Medical CL3960Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls26 reports
- ICU Medical CL4130Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls6 reports
- ICU Medical CL4131Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls6 reports
- ICU Medical CL-80SIcu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls5 reports
- ICU Medical CL-82Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls3 reports
- ICU Medical CL-80S-5Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls2 reports
- ICU Medical CL-72Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page