Medical device recalls: February 2019
The FDA posted 99 medical device recall records in February 2019, from 15 recall events by 15 companies. 24 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips Medical Systems Nederlands · 1 event, 64 records
- Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · 1 event, 13 records, 13 Class I
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · 1 event, 7 records, 7 Class I
- Zimmer Biomet Spine Inc. · 1 event, 3 records
- Physio-Control Inc · 1 event, 2 records, 2 Class I
- Alcon Research Laboratories Ltd. · 1 event, 1 record
- Arjohuntleigh Magog · 1 event, 1 record
- Bard Medical Division · 1 event, 1 record
FDA root cause
- Unknown/Undetermined by firm · 64
- Process control · 13
- Component design/selection · 8
- Nonconforming Material/Component · 4
- Device Design · 4
- Software design · 3
- Under Investigation by firm · 2
- Vendor change control · 1
Class I recalls (24)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0862-2019 · Physio-Control Inc · posted Feb 26, 2019 · Terminated
Certain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no response in keypad and…
- Z-0863-2019 · Physio-Control Inc · posted Feb 26, 2019 · Terminated
Certain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no response in keypad and…
- Z-0805-2019 · ICU Medical, Inc. · posted Feb 22, 2019 · Terminated
There is a potential for one lot of ChemoLock Vial Spike to contain burr particulate
- Z-0806-2019 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 14, 2019 · Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that…
- Z-0807-2019 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 14, 2019 · Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that…
- Z-0808-2019 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 14, 2019 · Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that…
- Z-0809-2019 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 14, 2019 · Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that…
- Z-0810-2019 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 14, 2019 · Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that…
- Z-0811-2019 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 14, 2019 · Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that…
- Z-0812-2019 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Feb 14, 2019 · Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that…
- Z-0710-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0711-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0712-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0713-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0714-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0715-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0716-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0717-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0718-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0719-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0720-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0721-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- Z-0722-2019 · Medex Cardio-Pulmonary Inc., d.b.a. Smiths Medical Company · posted Feb 4, 2019 · Terminated
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- CoaguChek XS PT Test ↗Class IZ-0787-2019 · Terrific Care LLC · posted Feb 1, 2019 · Terminated
Potential for abnormally high INR test results with the CoaguChek XS PT Test Strips for customers who purchased products directly from Terrific Care LLC dba MedEx Supply via website/telephone/facsimile.
Class II recalls (73)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0961-2019 · Alcon Research Laboratories Ltd. · posted Feb 27, 2019 · Terminated
Potential for the optical head to detach from the stand due to the lack of thread adhesive.
- Z-0939-2019 · Merge Healthcare, Inc. · posted Feb 25, 2019 · Terminated
Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault…
- Z-0866-2019 · Zimmer Biomet Spine Inc. · posted Feb 22, 2019 · Terminated
The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications
- Z-0867-2019 · Zimmer Biomet Spine Inc. · posted Feb 22, 2019 · Terminated
The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications
- Z-0868-2019 · Zimmer Biomet Spine Inc. · posted Feb 22, 2019 · Terminated
The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications
- Z-0869-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0870-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0871-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0872-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0873-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0874-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0875-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0876-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0877-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0878-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0879-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0880-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Intera 1.5T Explorer/Nova Dual, Model Number 781108. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0881-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0882-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Intera 3.0T Master/Quasar Dual, Model Number 781150. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0883-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0884-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0885-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0886-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Intera Achieva 1.5T Pulsar, Model Number 781171. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0887-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0888-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Intera Achieva 1.5T Nova Dual, Model Number 781173. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0889-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Intera 1.5T Achieva CV Nova, Model Number 781174. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0890-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Intera Achieva 1.5T IT Nova, Model Number 781175. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0891-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Intera Achieva 1.5T IT Nova Dual, Model Number 781176. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0892-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0893-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0894-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0895-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Achieva 1.5T High Performance, Model Number 781196. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0896-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0897-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0898-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0899-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- SmartPath to dStream for 1.5T, Model Number 781260. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0900-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0901-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0902-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0903-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0904-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0905-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0906-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0907-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Intera 1.5T High Performance, Model Number 781295. Nuclear Magnetic Resonance Imaging System ↗Class IIZ-0908-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0909-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0910-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0911-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0912-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0913-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0914-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0915-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0916-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0917-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0918-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0919-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0920-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0921-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0922-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0923-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0924-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0925-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0926-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0927-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0928-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0929-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0930-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0931-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0932-2019 · Philips Medical Systems Nederlands · posted Feb 22, 2019 · Terminated
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
- Z-0813-2019 · Stryker Medical Division of Stryker Corporation · posted Feb 7, 2019 · Terminated
It was identified that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub. A Health Hazard Evaluation was completed which identified…
- Z-0776-2019 · GE Healthcare, LLC · posted Feb 6, 2019 · Open, Classified
On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a location that is not what was…
- Z-0803-2019 · Bard Medical Division · posted Feb 6, 2019 · Terminated
potential for urine leakage around the vent on the vented sample port
- Z-0802-2019 · Arjohuntleigh Magog · posted Feb 5, 2019 · Terminated
Inability or difficulty in unbuckling/releasing the proning arm buckle might delay patient access in an emergency situation such as cardiac arrest. This, in turn, could contribute to a life-threatening event.
Class III recalls (2)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0822-2019 · Luminex Corporation · posted Feb 11, 2019 · Terminated
On lot of VERIGENE(R) BC-GP Utility Kits (20-011-018) were labeled with the VERIGENE(R) Amplification Tray label.
- Z-0792-2019 · Becton, Dickinson and Company, BD Biosciences · posted Feb 1, 2019 · Terminated
A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.