Skip to content

CERTAIN BELLATEK recallZ-2142-2019 · Biomet 3i, LLC · initiated Feb 26, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet 3i, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.

FDA root cause category: Process control

What the company is doing

The firm disseminated the notices by email on 02/26/2019. The letter requested that the dental laboratories quarantine and return any unused product and to further notify the affected clinician referral.

Recall details

Recall number
Z-2142-2019
Status
Terminated
Initiated by the company
Feb 26, 2019
Recalling firm
Biomet 3i, LLC
Quantity
6 units
Product code
NHA
FDA submissions
K052648, K102209
Distribution
Nationwide and Puerto Rico, Canada, and Australia

Product description

CERTAIN(R) BELLATEK(R) TITANIUM ABUTMENT 3.4MM, Item Number IIAT3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible

Code information (models, lots, serial numbers)

Lot Numbers/UDI:  8441907-1/(01)00844868031284(10)8441907-1 8441907-2/(01)00844868031284(10)8441907-2 8441907-3/(01)00844868031284(10)8441907-3 8442367-1/(01)00844868031284(10)8442367-1 8439134A-3/(01)00844868031284(10)8439134A-3 8439134A-4/(01)00844868031284(10)8439134A-4

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 82263).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page