Skip to content

CERTAIN® BELLATEK® IIAT3Recalls & reported problems

ZimVie US Corp LLC · CERTAIN® BELLATEK® TITANIUM ABUTMENT 3.4MM · Class 2 · Implant

All models of CERTAIN BELLATEK (12)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Feb 26, 2019Class IITerminated

    During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.

    Z-2142-2019 · Root cause: Process control · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code NHA (abutment, implant, dental, endosseous) under 21 CFR 872.3630 as a Class II device. 21 CFR 872.3630 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: To be used in conjunction with an endosseous dental implant fixture to aid in prosthetic rehabilitation.

Common questions about the CERTAIN® BELLATEK® IIAT3

Has the CERTAIN® BELLATEK® IIAT3 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2142-2019 (Class II), began on Feb 26, 2019: During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the CERTAIN® BELLATEK® IIAT3 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the CERTAIN® BELLATEK® IIAT3, and is it still sold?

The FDA's device database lists ZimVie US Corp LLC as the company responsible for it (brand CERTAIN® BELLATEK®, model IIAT3). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS devices

All ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page