icumedical ChemoClave Vented Vial Spike, 20 mm, REF CH-80S.… recallZ-1093-2019 · ICU Medical, Inc. · initiated Feb 21, 2019
Class I Terminated All recalls posted in April 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
FDA root cause category: Vendor change control
What the company is doing
The firm, Icumedical, initiated the recall by "URGENT: MEDICAL DEVICE RECALL" letter on 02/21/2019 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form via fax to: 1-855-497-9856 or email to: [email protected], even if you do not have the affected product. 3) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7114 (M-F, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Please visit http://expertezlabel.com to request additional labels for returning affected product. 4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle at 1-888-871-7114 (M-F, 8am-5pm ET) to obtain a response form. UPDATED INFORMATION: On 03 January 2019, ICU Medical, Inc. issued an Urgent Medical Device Recall letter informing customers of a recall of one lot (3757712) of the ChemoLock" Vial Spike, 20mm (CL-80S). As a result of continued evaluation, ICU Medical is expanding the scope of the previously distributed Urgent Medical Device Recall to include certain lots of ChemoLock" and ChemoClave Vial Spikes that may also be affected by this issue. All product lots identified in this communication are impacted by this recall. If you have any questions, please contact Customer Care at 1- 877-946-7747 General inquires /Product replacement options.
Recall details
- Recall number
- Z-1093-2019
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Feb 21, 2019
- Posted by the FDA
- Apr 15, 2019
- Recalling firm
- ICU Medical, Inc.
- Quantity
- 60500 devices
- Product code
- FPA
- FDA submissions
- K081361
- Distribution
- Worldwide distribution: US nationwide and countries of: Canada, France, and Singapore.
Product description
icumedical ChemoClave Vented Vial Spike, 20 mm, REF CH-80S. Used in IV administration sets. The ChemoClave is a needle-free Closed System Transfer Device (CSTD) that mechanically prohibits the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system during drug preparation and administration, thereby minimizing exposure of individuals, healthcare personnel and the environment to hazardous drugs.
Code information (models, lots, serial numbers)
Lot Numbers: 3826647, 3826648, 3826844, 3827598, 3839689, 3839690, 3854000, 3854001, 3866761, 3873006, 3936065, 3952512
Other devices under the same FDA submission
20 devices cleared under K081361 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ICU Medical CH3025Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- ICU Medical CH3035Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- ICU Medical CH3322Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- ICU Medical Z7107Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- ICU Medical CL4001Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- ICU Medical CH3639Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- ICU Medical CH3731Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- ICU Medical CH3581Icu Medical, Inc. · Set, administration, intravascular16 reports
- ICU Medical CH4000Icu Medical, Inc. · Set, administration, intravascular11 reports
- ICU Medical CH3943Icu Medical, Inc. · Set, administration, intravascular6 reports
- ICU Medical CH-72Icu Medical, Inc. · Set, administration, intravascular5 reports
- ICU Medical CV-100Icu Medical, Inc. · Set, administration, intravascular2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 82291).
- Z-1092-2019 · Terminatedicumedical ChemoClave Vented Vial Spike, 20 mm, REF CH-70S . Used in IV administration sets. The ChemoClave is a needle-free Closed System Transfer Device (CSTD) that mechanica…
- Z-1096-2019 · Terminatedicumedical ChemoLock, 20mm, (a) REF CL-80S (b) REF CL-80S-10 (10 units). Used in IV administration sets. The ChemoLock Closed System Drug Transfer Device prevents the tran…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page