Medical device recalls: April 2019
The FDA posted 23 medical device recall records in April 2019, from 11 recall events by 11 companies. 13 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- ICU Medical, Inc. · 1 event, 7 records, 7 Class I
- O-Two Medical Technologies, Inc. · 1 event, 3 records, 3 Class I
- Roche Diagnostics Corporation · 1 event, 3 records
- DePuy Spine, Inc. · 1 event, 2 records
- Edwards Lifesciences, LLC · 1 event, 2 records, 2 Class I
- Abbott · 1 event, 1 record
- Beaver Visitec · 1 event, 1 record
- Brainlab AG · 1 event, 1 record, 1 Class I
FDA root cause
- Vendor change control · 7
- Device Design · 4
- Process control · 3
- Under Investigation by firm · 2
- Software design · 2
- Process design · 2
- Packaging process control · 1
- Labeling design · 1
Class I recalls (13)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Edwards Lifesciences, Fogarty Dilation Atrioseptostomy Catheter, Sterile EO, Rx Only, REF: 830705F ↗Class IZ-1116-2019 · Edwards Lifesciences, LLC · posted Apr 25, 2019 · Terminated
Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of balloon fragmentation or detachment.
- Edwards Lifesciences, Miller Balloon Atrioseptostomy Catheter, Sterile EO, Rx Only, REF: 830515F ↗Class IZ-1117-2019 · Edwards Lifesciences, LLC · posted Apr 25, 2019 · Terminated
Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of balloon fragmentation or detachment.
- Z-1082-2019 · Brainlab AG · posted Apr 15, 2019 · Terminated
In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ),…
- Z-1092-2019 · ICU Medical, Inc. · posted Apr 15, 2019 · Terminated
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
- Z-1093-2019 · ICU Medical, Inc. · posted Apr 15, 2019 · Terminated
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
- Z-1094-2019 · ICU Medical, Inc. · posted Apr 15, 2019 · Terminated
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
- Z-1095-2019 · ICU Medical, Inc. · posted Apr 15, 2019 · Terminated
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
- icumedical ChemoLock, 20mm,Class IZ-1096-2019 · ICU Medical, Inc. · posted Apr 15, 2019 · Terminated
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
- Z-1097-2019 · ICU Medical, Inc. · posted Apr 15, 2019 · Terminated
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
- Z-1098-2019 · ICU Medical, Inc. · posted Apr 15, 2019 · Terminated
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
- Z-0976-2019 · O-Two Medical Technologies, Inc. · posted Apr 4, 2019 · Terminated
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the…
- Z-0977-2019 · O-Two Medical Technologies, Inc. · posted Apr 4, 2019 · Terminated
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the…
- Z-0978-2019 · O-Two Medical Technologies, Inc. · posted Apr 4, 2019 · Terminated
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the…
Class II recalls (10)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1225-2019 · DePuy Spine, Inc. · posted Apr 27, 2019 · Terminated
A complaint trend was observed for post-operative loss of height and migration for Concorde Lift G3+ implants.
- Z-1226-2019 · DePuy Spine, Inc. · posted Apr 27, 2019 · Terminated
A complaint trend was observed for post-operative loss of height and migration for Concorde Lift G3+ implants.
- Z-1220-2019 · Abbott · posted Apr 25, 2019 · Terminated
The device is unable to pair with the mobile app due to the device incorrectly determining the certificate has expired.
- Z-1203-2019 · Thoratec Corp. · posted Apr 19, 2019 · Terminated
Small black plastic particle was observed in the left ventricle during implant procedure.
- Bilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190, 05795320160, 05795338160, 05795346160Class IIZ-1145-2019 · Roche Diagnostics Corporation · posted Apr 13, 2019 · Terminated
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS…
- Z-1146-2019 · Roche Diagnostics Corporation · posted Apr 13, 2019 · Terminated
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS…
- Z-1147-2019 · Roche Diagnostics Corporation · posted Apr 13, 2019 · Terminated
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS…
- IMAGEnet 6 v1.53- ↗Class IIZ-1138-2019 · Topcon Medical Systems, Inc. · posted Apr 12, 2019 · Terminated
When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information…
- Z-1131-2019 · Johnson & Johnson Vision Care, Inc. · posted Apr 11, 2019 · Terminated
One lot of ACUVUE OASYS for ASTIGMATISM Brand Contact lenses has the potential for a limited number of individual contact lens packages to have an incomplete packaging seal.
- Z-1111-2019 · Beaver Visitec · posted Apr 9, 2019 · Terminated
Mislabeled. The package labeling indicates a size 2.4mm, 45 degree Double Bevel knife however, the marking on device shows a 2.75mm size device
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.