RingLoc Hip System recallZ-1086-2019 · Zimmer Biomet, Inc. · initiated Mar 1, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Two lengths of the 6.5mm RingLoc Hip System Self-Tapping Bone Screw may have the potential for the 30mm screws to be packaged as 20mm screws and vice versa.
FDA root cause category: Packaging process control
What the company is doing
Zimmer Biomet conducted recall by letter dated 3/1/19 to the consumer/user level, including any intermediate wholesale or retail consignees. All distributors via email. Hospital risk managers and surgeons, as well as distributors with product notified Fed'x. Accounts asked to review notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory.3. Immediately return all affected product from your distributorship and from affected hospitals within your territory.a. Complete Attachment 1 Inventory Return Certification Form. Questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.
Recall details
- Recall number
- Z-1086-2019
- Status
- Terminated
- Initiated by the company
- Mar 1, 2019
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 809 total
- Product code
- JWH
- FDA submissions
- K912245
- Distribution
- Worldwide Distribution - US Nationwide Foreign: CHILE CHINA COSTA RICA GUJRAT,INDIA JAPAN MEXICO CDMX SINGAPORE
Product description
Biomet RingLoc Hip System Self-Tapping Bone Screw 6.5mmx20mm-Low Profile Titanium. Accessory to total hip implant Item Number: 103531 Product Usage: Accessory to total hip implant
Code information (models, lots, serial numbers)
Lot Numbers/UDI Number 523430 (01) 00880304001961 (17) 280804 (10) 523430 523450 (01) 00880304001961 (17) 280727 (10) 523450 860840 (01) 00880304001961 (17) 280806 (10) 860840 860850 (01) 00880304001961 (17) 280809 (10) 860850 860860 (01) 00880304001961 (17) 280806 (10) 860860
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RingLoc® Hip System 103531Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)1 recall0 reports
Product lines
Other devices under the same FDA submission
8 devices cleared under K912245 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- RingLoc® Hip System 103533Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall4 reports
- Biomet® Knee System 141205Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER14 reports
- RingLoc® Hip System 103538Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)2 reports
- RingLoc® Hip System 103532Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- RingLoc® Hip System 103536Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- RingLoc® Hip System 103535Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- RingLoc® Hip System 103534Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 death1 report
- RingLoc® Hip System 103537Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 82332).
- Z-1087-2019 · TerminatedBiomet RingLoc Hip System Self-Tapping Bone Screw 6.5mmx30mm-Low Profile Titanium, Accessory to total hip implant Item Number: 103533 Product Usage: Accessory to total hip implant
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page