RingLoc Hip System recallZ-1087-2019 · Zimmer Biomet, Inc. · initiated Mar 1, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two lengths of the 6.5mm RingLoc Hip System Self-Tapping Bone Screw may have the potential for the 30mm screws to be packaged as 20mm screws and vice versa.

FDA root cause category: Packaging process control

What the company is doing

Zimmer Biomet conducted recall by letter dated 3/1/19 to the consumer/user level, including any intermediate wholesale or retail consignees. All distributors via email. Hospital risk managers and surgeons, as well as distributors with product notified Fed'x. Accounts asked to review notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory.3. Immediately return all affected product from your distributorship and from affected hospitals within your territory.a. Complete Attachment 1 Inventory Return Certification Form. Questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.

Recall details

Recall number
Z-1087-2019
Status
Terminated
Initiated by the company
Mar 1, 2019
Recalling firm
Zimmer Biomet, Inc.
Quantity
809 units total
Product code
JWH
FDA submissions
K912245
Distribution
Worldwide Distribution - US Nationwide Foreign: CHILE CHINA COSTA RICA GUJRAT,INDIA JAPAN MEXICO CDMX SINGAPORE

Product description

Biomet RingLoc Hip System Self-Tapping Bone Screw 6.5mmx30mm-Low Profile Titanium, Accessory to total hip implant Item Number: 103533 Product Usage: Accessory to total hip implant

Code information (models, lots, serial numbers)

Lot Numbers/UDI Number 859960 (01) 00880304008700 (17) 280809 (10) 859960 859980 (01) 00880304008700 (17) 280805 (10) 859980 860120 (01) 00880304008700 (17) 280806 (10) 860120 860310 (01) 00880304008700 (17) 280823 (10) 860310 860580 (01) 00880304008700 (17) 280807 (10) 860580

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

8 devices cleared under K912245 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 82332).